Operating instructions
3
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
43
Fig 3-7.
Voltage profile.
Note:
The set voltage is the voltage limit. Under conditions of limiting current, it
may take some time for the maximum voltage to be reached, and under
certain power conditions, it may not be reached at all.
A lower voltage step may be entered at the end of the Ettan IPGphor 3 protocol.
Band sharpness will decrease however.
Verifying the protocol
To verify the protocol:
1
scroll through each step: position the cursor in the step number field using
the
right
and
left arrow
keys, and
2
then advance through the steps using the
up
and
down
arrow keys.
Edit as required by placing the cursor with the
right
and
left
arrow keys and
adjusting the value with the
up
and
down
arrow keys.
Make sure that the voltage field in all unused steps displays a value of 0, or that
the value in the Hrs field is 00:00.
Note:
Pressing the
up
arrow key in
screen 4
wraps back to the
rehydration
screen.
Exiting edit mode
Press the
EDIT
key to save any changes and exit the step edit mode. The
protocol
identification
screen will appear.