Ettan IPGphor 3 Isoelectric Focusing System function and description
1
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
17
1.3
Specifications
Working surface
User interface
Operating environment
Electrode areas
Gold-plated copper
Capacity
Up to 12 fixed-length strip holders or one
manifold
Platform temperature
15–30 °C ±2 °C
Control panel
7-key membrane keypad.
Liquid crystal display (LCD)
4 lines, 24 characters per line
Programmable parameters
Rehydration time, platform temperature,
maximum current limit per strip, voltage
limit in each step, voltage step or
gradient, and step duration.
Protocol capacity
10 protocols, with up to 9 steps each
Serial port
RS232, 1200 or 9600 baud, 8 data bits 1
stop bit, no parity, no flow control.
9600 baud is mandatory for connection
with Ettan IPGphor 3 Control Software.
Temperature
15–32 °C
Relative humidity
0–70%*
* High relative humidity may result in condensation forming on the platform sur-
face. Runs conducted at a platform temperature of 20 °C and relative humidity
less than 70% will generally not cause condensation. Platform temperatures
greater than 20 °C allow for higher relative humidity without condensation. Envi-
ronments for which condensation occurs on the platform surface may vary. If
condensation occurs, the unit may require conditioned air to reduce tempera-
ture and relative humidity.
Summary of Contents for Ettan IPGphor 3
Page 1: ...GE Healthcare Ettan IPGphor 3 Instrument User Manual Ettan...
Page 2: ......
Page 4: ......
Page 54: ...4 Care and maintenance 54 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Page 62: ...6 Customer service information 62 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Page 63: ......