Operating instructions
3
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
47
To toggle between VhT, VhS and Hrs:
1
move the cursor to the
Hrs/Vhrs
field of the display using the
right
or
left
arrow key, and
2
change the display to reflect VhT, VhS and Hrs using the
up
and
down
keys.
To change the display to show remaining Hrs, VhS or VhT:
1
move the cursor under the
Elapsed
or
~ Remain
field and use the
up
or
down
buttons.
The temperature, voltage and current limits for the current step may be adjusted
directly on this screen as well.
Pressing the
Edit
key will access the
step parameter
screen. Pressing
Edit
again
will return to the
Running
screen. Use the
up
arrow in the
S1
field to access the
Rehydration/IEF
screen.
Note:
If you adjust the voltage during a gradient step the step will be converted
to Step for the remainder of the step.
As isoelectric focusing proceeds, the bromophenol blue tracking dye will migrate
toward the anode and leave the IPG strip. This generally occurs well before
focusing is complete.
Note:
A colorless strip is no indication that IEF is complete.
If the tracking dye does not move, then no current is flowing across the strip.
Check that the strip holder electrodes are in contact with the electrode areas. See
the Troubleshooting section for additional suggestions.
When the protocol ends, audible beeps will sound and the following screen is
displayed:
The final instrument values are displayed. Press the
STOP
key, as indicated, to
reset the instrument.
R u n n i n g P r o t #
S 1 G r a d 5 0 0 V 1 : 0 0 H r s
S 2 G r a d 2 0 0 0 V 1 : 0 0 H r s
S 3 S t e p 8 0 0 0 V 2 : 0 0 H r s
R u n e n d e d a t
8 0 0 0 V 1 2 0 0 0 V h r s To t a l
5 0
µ
A 8 : 0 0 H r To t a l
P r e s s S t o p t o r e s e t
Summary of Contents for Ettan IPGphor 3
Page 1: ...GE Healthcare Ettan IPGphor 3 Instrument User Manual Ettan...
Page 2: ......
Page 4: ......
Page 54: ...4 Care and maintenance 54 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Page 62: ...6 Customer service information 62 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Page 63: ......