• Check speed indication on LCD and additional speed indication for test subject.
Check noise level of idling drive unit (e.g. for grinding noises or noise from the bearing)
Expected result: Speed displayed on both displays and no noticeable noise
Pass/Fail
• Units with NIBP module:
- Check cuff tubing and tubing connections, microfon connectors
- Check blood pressure measurement on test subject at rest
- Check blood pressure readings on LCD
Expected result:
BP values are displayed and in proper range
Pass/Fail
• Start manual exercise test, change load (e.g. start 50 W, increase 25 W)
- Check load level and load indication
Expected result:
Load changes as expected
Pass/Fail
• If ergometer is remote controlled from ECG recorder or PC-ECG system:
- Check connecting cable
- Check load control via ECG recorder or PC ECG system
- Check remote start if applicable
Expected result:
Ergometer can be controlled by external unit
Pass/Fail
Summary of Contents for eBike BASIC
Page 1: ...GE Healthcare eBike II eBike II L eBike II EL Service Manual 2018112 069 Revision F ...
Page 2: ......
Page 13: ...2018112 069 Rev F eBike II eBike II L eBike II EL 13 ...
Page 39: ...2018112 069 Rev F eBike II eBike II L eBike II EL 39 ...
Page 47: ...2018112 069 Rev F eBike II eBike II L eBike II EL 47 ...
Page 77: ...2018112 069 Rev F eBike II eBike II L eBike II EL 77 ...
Page 133: ...2018112 069 Rev F eBike II eBike II L eBike II EL 133 ...
Page 141: ...2018112 069 Rev F eBike II eBike II L eBike II EL 141 ...
Page 212: ... 212 eBike II eBike II L eBike II EL 2018112 069 Rev F ...
Page 214: ... 214 eBike II eBike II L eBike II EL 2018112 069 Rev E ...
Page 233: ......