2018112-069 Rev. F
eBike II, eBike II L, eBike II EL
- 33 -
o
pen
the
houSing
of
the
bKe (
bootoM
Side
)
• For CardioSoft/CS V6.73 and above:
Click onto „COM Port Number“ and select an unused
COM port between COM1 and COM32.
• Confirm with OK.
• Close all open windows.
CoM - I
nterfaCe
extensIon
(MaC 5000 st/5500)
The MAC 5000 ST / MAC 5500 use analogue signals
to set the load of the eBike, while the bloodpressure
measurement is controlled with digital signals..
The eBike II is by default only equipped with a digital
interface.
To connect an eBike II with a MAC 5000 ST / 5500 a
special adapter (COM module PN 2008xxx-001) has to be
installed to add an analogue interface to the eBike II.
A jumper adapter has to be placed onto the BKE board to
connect all required signals to the DSUB connector:
• Open the housing of the BKE on the bottom side of
the ergometer.
coM
Module
(pn 2008
xxx
)
with
JuMper
and
Screw
Summary of Contents for eBike BASIC
Page 1: ...GE Healthcare eBike II eBike II L eBike II EL Service Manual 2018112 069 Revision F ...
Page 2: ......
Page 13: ...2018112 069 Rev F eBike II eBike II L eBike II EL 13 ...
Page 39: ...2018112 069 Rev F eBike II eBike II L eBike II EL 39 ...
Page 47: ...2018112 069 Rev F eBike II eBike II L eBike II EL 47 ...
Page 77: ...2018112 069 Rev F eBike II eBike II L eBike II EL 77 ...
Page 133: ...2018112 069 Rev F eBike II eBike II L eBike II EL 133 ...
Page 141: ...2018112 069 Rev F eBike II eBike II L eBike II EL 141 ...
Page 212: ... 212 eBike II eBike II L eBike II EL 2018112 069 Rev F ...
Page 214: ... 214 eBike II eBike II L eBike II EL 2018112 069 Rev E ...
Page 233: ......