background image

GE Healthcare

eBike II / eBike II L / eBike II EL

Service Manual

2018112-069 Revision F

Summary of Contents for eBike BASIC

Page 1: ...GE Healthcare eBike II eBike II L eBike II EL Service Manual 2018112 069 Revision F ...

Page 2: ......

Page 3: ...911 204 EBIKE BASIC BP US 2017911 205 EBIKE COMFORT 2017911 206 EBIKE COMFORT US 2017911 207 EBIKE COMFORT BP 2017911 208 EBIKE COMFORT BP US 2017911 209 EBIKE L 240V 2017911 210 EBIKE L BP 240V 2017911 211 EBIKE L EXT TRMNL 240V 2017911 212 EBIKE L BP EXT TRMNL 240V 2017911 213 EBIKE L 120V US 2017911 214 EBIKE L BP 120V US 2017911 215 EBIKE L EXT TRMNL 120V US 2017911 216 EBIKE L BP EXT TRMNL 12...

Page 4: ...anged p 147 P N lift tool corrected p 185 load accuracy added 2018112 069 Rev D 2013 06 p 137 SPARE eBIKE II Basic Comfort BP Module added 2018112 069 Rev E 2014 02 p 22 23 Motor control eBike L BCU added p 26 27 Motor control eBike EL SCU added p 31 32 Check USB interface and Change COM port assignment modified p 101 Replacement control unit motors and remote control compatible modules eBike II L...

Page 5: ...ank pedal set new pedal set p 151 new power cord p 157 new blood pressure cuffs p 159 new adjustable cranks new USB driver CD p 160 new power cord p 163 modified picture p 165 new crank pedal set new pedal set p 166 picture changed p 180 new part added p 187 picture changed fuses removed p 189 new blood pressure cuffs p 191 4 new parts p 192 new USB driver CD p 195 new part added p 201 203 chapter...

Page 6: ...gram eBike II L 21 Wiring diagram eBike II L with NIBP 22 Wiring diagram eBike II L with BCU 23 Wiring diagram eBike II L with NIBP and BCU 24 Wiring diagram eBike II EL with NIBP 25 Wiring diagram eBike II EL control terminal detached with NIBP 26 Wiring diagram eBike II EL with NIBP and SCU 27 Wiring diagram eBike II EL control terminal detached with NIBP and SCU 28 Control terminal 29 ZME Main ...

Page 7: ...alibration 51 Ergometer Type 52 System Config 52 Error log 53 NIBP Test NIBP Calibration 53 Analog Comparison calibration 54 Alignment Input Analog Loadcontrol 55 Alignment Ouput Analog Load value 56 Alignment Output Analog Heartrate 56 Alignment Ouput Analog Systole 56 Alignment Ouput Analog Diastole 57 Alignment Ouput RPM 57 Alignment Output ECG 57 Alignment Calibration 58 Static calibration 58 ...

Page 8: ...3 Replace motor control unit with BCU 105 Replace saddle motor 108 Replace Lift Motor for Couch Adjustment 109 Replace saddle motor with BCU 110 Replace Lift Motor for Couch Adjustmen with BCU 111 Replace External Speed Indicator eBike L ext 112 eBike II EL 113 Preventive Maintenance 113 Final checkout procedure Functional test 113 Electrical Safety 115 Mechanical assembly 116 Controls and Indicat...

Page 9: ...57 eBike II basic comfort tools 162 eBike II basic comfort add on kits 163 eBike II L EL spare parts 164 eBike II L EL accessories 189 eBike II L EL tools 196 eBike II L EL add on kits 197 Illustrations for special spareparts eBike EL 198 Cleaning Maintenance Disposal 201 Technical Specifications eBike II basic comfort 204 Technical Specifications eBike II L 206 Technical Specifications eBike II E...

Page 10: ...n the service invoice Patient Hazard Equipment Damage Use original spareparts only For information about product changes refer to the manufacturer s original documentation only Observe the fuses original ratings characteristics and Ampere interrupting capacity Parts of the device that ensure its safe operation must be neither damaged nor obviously unsuitable This applies to insulation and insulati...

Page 11: ...nnection to protective ground The necessary electrical safety tests for these devices together with the limit values are listed in the chapter eBike II L Preventive maintanance and eBike II EL Preventive maintanance Inspection according to German Medical Device Operator Ordinance Technical Safety Inspections and Inspections of the Measuring Systems According to 6 and 11 of the German Medical Devic...

Page 12: ...5 5 mm 1 pc 2 5 mm straight 2 pc 7 mm 1 pc 3 mm straight and elbow 1 pc 13 mm 1 pc 4 mm straight and elbow 1 pc 17 mm 1 pc 5 mm straight 1 pc 19 mm 1 pc 6 mm straight and elbow Socket wrenches Pliers 1 pc 5 5 mm 1 pc side cutter 1 pc 13 mm 1 pc flat nose pliers 1 pc multigrip pliers Special tools 1 pc crank extractor P N 2005737 001 2 pc round pin max 3 mm minimum length 15 mm 1 pc ergometer eBike...

Page 13: ...2018112 069 Rev F eBike II eBike II L eBike II EL 13 ...

Page 14: ... stable and standardized protocol used in the automotive industrie The connection between different modules is made by use of standard shielded patch cables These cables carry the power lines 24 volt and the CAN bus signal lines Abbreviations ZME Main control unit LRE Load regulation unit BKE Basic communication unit BDE Bloodpressure unit PNE Pneumatic unit NIBP MAE Saddle motor unit DMS Strain g...

Page 15: ... PNE MAE BKE DMS ZME Switch connector microfone connector cuff Control terminal Brake unit eddycurrent brake NIBP Modul Saddle motor Interface power cord power supply motor saddleheight position sensor Block diagram bike ergometer eBike II basic comfort electronics ...

Page 16: ...D LRE BDE PNE BKE DMS ZME Brake unit eddycurrent brake NIBP Modul Interface connector microfone left side connector cuff left side connector microfone right side connector cuff right side Control terminal switch power cord power supply motor control saddle motor motor longit motor lateral 1200 only remote ctrl ...

Page 17: ...D LRE BDE PNE BKE DMS ZME Brake unit eddycurrent brake NIBP Modul Interface connector microfone left side connector cuff left side connector microfone right side connector cuff right side Control terminal switch power cord power supply motor control saddle motor motor longit motor lateral 1200 only remote ctrl ...

Page 18: ...CD LRE BDE PNE BKE DMS ZME EDA Brake unit eddycurrent brake NIBP Modul Interface connector microfone left side connector cuff left side connector microfone right side connector cuff right side Control terminal detached RPM display switch power cord power supply motor control saddle motor motor longit motor lateral 1200 only remote ctrl ...

Page 19: ... supply BKE module interface patch cable RED RS 232 Eddie current brake LRE PCB Saddle motor eBike comfort only MAE PCB X6 3 pole to sensor X5 2 pole to motor RJ 45 RJ 45 RJ 45 RS 232 RJ 45 RJ 45 RJ 45 X11 brown white patch cable BLUE patch cable BLACK 2018112 069 Rev F eBike II eBike II L eBike II EL 19 Wiring diagram eBike II basic comfort ...

Page 20: ...RE PCB Saddle motor eBike comfort only MAE PCB X6 3 pole to sensor X5 2 pole to motor patch cable BLUE RJ 45 RJ 45 RJ 45 RS 232 RJ 45 RJ 45 X11 brown white Bloodpressure module patch cable BLACK patch cable RED RS 232 RJ 45 RJ 45 RJ 45 BDE PCB X1 X16 microphone cable BKE module interface 20 eBike II eBike II L eBike II EL 2018112 069 Rev F Wiring diagram eBike II basic comfort with NIBP ...

Page 21: ...rake LRE PCB RJ 45 RJ 45 RJ 45 RS 232 RJ 45 X11 brown white patch cable BLACK BKE module interface patch cable RED RS 232 RJ 45 X16 Power switch Modular connector Power input E C G B F PE A D H 1 5 2 PE 3 4 Motor control Saddle motor Motor longitudinal Remote control 7 8 6 PE 2 1 2 1 DSUB connector 2018112 069 Rev F eBike II eBike II L eBike II EL 21 Wiring diagram eBike II L ...

Page 22: ... white Bloodpressure module patch cable BLACK BKE module interface patch cable RED RS 232 RJ 45 RJ 45 RJ 45 BDE PCB X1 X16 microphone cable cuff connector panel left cuff connector panel right Power switch Modular connector Power input E C G B F PE A D H 1 5 2 PE 3 4 Motor control Saddle motor Motor longitudinal Remote control 7 8 6 PE 2 1 2 1 DSUB connector 22 eBike II eBike II L eBike II EL 2018...

Page 23: ...or Power supply Eddie current brake LRE PCB RJ 45 RJ 45 RJ 45 RS 232 RJ 45 X11 brown white patch cable BLACK BKE module interface patch cable RED RS 232 RJ 45 X16 Power switch Modular connector Power input E C G B F PE A D H 1 5 2 PE 3 4 Motor control Saddle motor Motor longitudinal Remote control 7 8 6 PE 2 1 2 1 DSUB connector 2018112 069 Rev F eBike II eBike II L eBike II EL 23 ...

Page 24: ...te Bloodpressure module patch cable BLACK BKE module interface patch cable RED RS 232 RJ 45 RJ 45 RJ 45 BDE PCB X1 X16 microphone cable cuff connector panel left cuff connector panel right Power switch Modular connector Power input E C G B F PE A D H 1 5 2 PE 3 4 Motor control Saddle motor Motor longitudinal Remote control 7 8 6 PE 2 1 2 1 DSUB connector Wiring diagram eBike II L with NIBP and BCU...

Page 25: ...pressure module patch cable BLACK BKE module interface patch cable RED RS 232 RJ 45 RJ 45 RJ 45 BDE PCB X1 X16 microphone cable cuff connector panel left cuff connector panel right Power switch Modular connector Power input E C G B F PE A D H 1 5 2 PE 3 4 Motor control Saddle motor Motor longitudinal Remote control 7 8 6 PE 1 3 1 3 DSUB connector Motor lateral tilt 2 2 2018112 069 Rev F eBike II e...

Page 26: ...E module interface patch cable RED RS 232 RJ 45 RJ 45 RJ 45 BDE PCB X1 X16 microphone cable cuff connector panel left cuff connector panel right Power switch Modular connector Power input E C G B F PE A D H 1 5 2 PE 3 4 Motor control Saddle motor Motor longitudinal Remote control 7 8 6 PE 1 3 1 3 DSUB connector Motor lateral tilt 2 2 RPM display patch cable GREY patch cable BLACK 26 eBike II eBike...

Page 27: ...sure module patch cable BLACK BKE module interface patch cable RED RS 232 RJ 45 RJ 45 RJ 45 BDE PCB X1 X16 microphone cable cuff connector panel left cuff connector panel right Power switch Modular connector Power input E C G B F PE A D H 1 5 2 PE 3 4 Motor control Saddle motor Motor longitudinal Remote control 7 8 6 PE 1 3 1 3 DSUB connector Motor lateral tilt 2 2 2018112 069 Rev F eBike II eBike...

Page 28: ...dule interface patch cable RED RS 232 RJ 45 RJ 45 RJ 45 BDE PCB X1 X16 microphone cable cuff connector panel left cuff connector panel right Power switch Modular connector Power input E C G B F PE A D H 1 5 2 PE 3 4 Motor control Saddle motor Motor longitudinal Remote control 7 8 6 PE 1 3 1 3 DSUB connector Motor lateral tilt 2 2 RPM display patch cable GREY patch cable BLACK 28 eBike II eBike II ...

Page 29: ... Control terminal ZME Main electronic board The ZME board represents the mainboard of the eBike II electronics The LC display and the foil keyboard are connected to this board Remove the flat cable of the display and loose the 4 screws fixing the bearing metal plate The display unit can be removed ...

Page 30: ...ounted with the component side towards the aluminium housing of the drive unit The microprocessor of the LRE board also generates the interface signals for external communication e g with PC ECG or ECG recorders The BKE interface board generating the final RS 232 and USB signals is always connected directly to the LRE board LRE board 1 Connector for power supply cable from main switch 2 Connector ...

Page 31: ...32 USB Interface Driver installation virtual COMx Administrator rights are required to install the driver Start PC and Monitor Finish all running programs in background running programs as well Insert the USB driver CD PN 2017911 130 into the CD ROM drive Depending on the Windows operating system used start the appropriate driver from the CD Windows 2000 CP210x_VCP_Win2K_XP_S2K3 exe Windows XP Win...

Page 32: ...l Port Silicon Labs CP 210x USB to UART Bridge is assigned from COM1 up to COM4 no change is necessary If an USB Serial Port higher COM4 is assigned one of the COM ports 1 to 4 has to be deactivated to use this port number for eBike For CardioSoft CS V6 73 and above If an USB Serial Port Silicon Labs CP 210x USB to UART Bridge is assigned from COM1 up to COM32 no change is necessary If an USB Seri...

Page 33: ...o set the load of the eBike while the bloodpressure measurement is controlled with digital signals The eBike II is by default only equipped with a digital interface To connect an eBike II with a MAC 5000 ST 5500 a special adapter COM module PN 2008xxx 001 has to be installed to add an analogue interface to the eBike II A jumper adapter has to be placed onto the BKE board to connect all required si...

Page 34: ...o the 9 pole DSUB connector PORT 1 and screwed to the ergometer chassis The analog adapter has to be activated as follows In Service Menu go to System Configuration and set Analog Interface to I Then power OFF ON the eBike Go to Settings select ECG Type enter password 003 and select Analog Digital Insert the appropriate digital and analog connecting cables and fix the cables to the chassis using a...

Page 35: ...e measurement electronics is placed on the BDE board The valves and the power driver for the pump are placed on a separate board located in an sound and vibration absorbing foam carrier The pump is also placed in a separate sound absorbing foam carrier Valve and driver PCB Pump and compensating container Bloodpressure measuring electronicse BDE ...

Page 36: ...pressure unit cuff connector BDE electronic board pressure sensor 1 pressure sensor 2 PNE pneumatic unit deflating valve compensating container measuring valve pump sound absorber check valve sound absorber microfone connector BDE electronic board PNE pneumatic unit deflating valve measuring valve pump 6 2 patch cable patch cable ...

Page 37: ...ystem 2 nd pressure sensor continously monitores the actual cuff pressure and the pump time By use of the two RJ 45 jacks the BDE board is connected to the eBike II bus system PNE Bloodpressure Pneumatic unit Depending on the actual ergometer type the pneumatic components pump valves compensating container are arranged on different positions Electronic cabling and connection of the pressure tubes ...

Page 38: ...ed with a special cable to the MAE board DRE External rpm display PCB The eBike supine ergometers eBike II 1000 and 1200 can be equipped with an external control terminal In this case a special rpm display for patient information is needed This electronic board has its own microprocessor and communicates via the eBike bus CAT 5 patch cable The rpm display unit has to be connected as the last unit ...

Page 39: ...2018112 069 Rev F eBike II eBike II L eBike II EL 39 ...

Page 40: ...n bloodpressure Disconnect the microfon cable by pulling out the 3 pole jack see picture Disconnect the blue hose at the hose connector see picture Opening of the control terminal Loose the 4 Phillips head screws at the bottom of the housing Lift carefully the upper part of the housing with the foil keyboard and disconnect the cable from the ZME board The front panel is plugged into the enclosure ...

Page 41: ...d both ends of the small fixation spring towards the axis by use of a flat nose plier Assembly of the load unit 1 Cast aluminium housing 2 Precision bearings 3 Rotational speed sensor magnetic 4 LRE load regulation electronics 5 Sprocket with chain 6 Eddy current brake 1000 W 7 Strain gauge unit DMS 8 Pretensioning spring Hint Exchange of the chain is possible without disassembling the complete ho...

Page 42: ... retaining spring towards the end of the chain The chain lock consists of a retaining spring 1 an outer plate 2 and an outer plate with pins 3 Turn the sprocket to move the chain coupling to the top position Remove carefully the fixation clamp of the chain with a pliers Remove the outer plate with pins ...

Page 43: ...n the sprocket Disassemble the old chain Close the new chain Bring the two ends of the new chain together and connect with the chain lock Insert outer plate with pins into the chain ends attach outer plate and press chain lock together Attach retaining spring with the closed end of the retaining ring pointing in the direction of chain travel Slide retaining spring in the grooves in the pins ...

Page 44: ...v F LRE board exchange disconnect strain gauge unit cable remove 3 Phillips screws swing down the LRE board disconnect the RJ 45 cables disconnect the power supply connector remove the cable shoes to the eddy current brake by using a flat screw driver ...

Page 45: ...atch the sealing wax from the top suspension of the tensioning rope e g by use of a screw driver Remove the tensioning rope to the top by pulling it off the vertical slot The assembly of the replacement DMS unit is performed in reverse order WARNING The rubber coated side of the strain gauge should not be hurt To fix the tensioning rope in the upper fixture a standard glue can be used ...

Page 46: ...ing see detailed description at the appropriate ergometer type Disconnect the patch cables and the power supply cable from the LRE board The load unit is fixed with 6 Allen key screws at the ergometer chassis Loose the screws with a ratchet prevent the load unit from falling down weight approx 10 kg ...

Page 47: ...2018112 069 Rev F eBike II eBike II L eBike II EL 47 ...

Page 48: ... key label describing the function is shown on the display Turning the System On You turn the ergometer on by pressing the power switch the green indicator in the switch lights up The ergometer runs a self test As soon as the self test appears on the ergometer display press the two softkeys synchronously and hold To prevent settings from being changed by chance the service menu is protected by a p...

Page 49: ... beep ON date time current values ECG type digital rpm min 54 max 65 pulse display OFF Service menu Baudrate 8N2 4800 ECG Startpulse leadtime 60 sec pulswidth 100 msec NIBP Parameter leadtime 60 sec Resting NIBP ON PC Mode Automatic ON Ergometer Type ER900 Analog comparison 20W 0 2 V 300W 3 0 V System Config Saddle engine eBike basic L EL 0 eBike comfort 1 NIBP module devices w o NIBP 0 devices wi...

Page 50: ...ometers is 4 800 Baud Select the desired parameter by use of the right and left softkeys and confirm with ENTER ECG Startpulse To configurate the remote start pulse for external ECG recorders ergometer is controlled by internal programs Select the desired parameter by use of the right and left softkeys and confirm with ENTER Adjust the appropriate valuewith the arrow keys and store the value with ...

Page 51: ...s received from the interface Saddle Calibration eBike II comfort only This parameter is used to adjust the electrical saddle motor display scale 1 40 Select TOP Position and confirm with ENTER By use of the arrow keys move the saddle to the desired top position standard 36 cm above saddle guidance and confirm with SET Select Bottom Position and confirm with ENTER By use of the arrow keys move the...

Page 52: ...g to I if the module is integrated in the actual ergometer 60 Hz Power This setting depends on the mains frequency used and sets internal filters on the LRE board Select 0 for a 50 Hz mains frequency Select I for a 60 Hz mains frequency Configuration of identification data string Ergometer Type ER 900 ER 800 ENTER Hint Must be set to ER900 always Settting of ergometer configuartion System Config S...

Page 53: ...the arrow keys the following error messages can be displayed RETURN closes the error log display The software of the eBike II ergometers stores the signal strength of each measure and indicates failed measures NIBP Test NIBP Calibration This menu is visible if the ergometer is equipped with an automatic NIBP module and this module is activated Test and calibration of the NIBP module is described i...

Page 54: ...al to be aligned and confirm with ENTER Alignment of analog input output signals Analog Calibration Input 1 Load Output 1 Load Output 2 Heart Output 3 Syst Output 4 Diast RETURN Hint To allign the analog load signals the corresponding ECG recorder should be connected as a reference voltage source Disconnect a RS232 or USB cable if connected The allignment of the analog signals can be controlled af...

Page 55: ... the load value to the ECG recorder vaue e g 20 Watt and confirm with SET The setting is stored in the ergometer Set the ECG recorder output to a high load value e g 300 Watt The according voltage is displayed IN Volt Wait until the voltage has settled Select GAIN and confirm with ENTER By using the arrow keys set the load value to the ECG recorder vaue e g 300 Watt and confirm with SET The settin...

Page 56: ...possible to adjust the voltage and or the heartrate value The setting is performed as described above under Alignment Ouput Analog Load value Setting range GAIN voltage 500 to 2000 mV in 2 mV steps GAIN heartrate 50 to 250 bpm in 10 bpm steps Alignment Ouput Analog Systole The output voltage representing the systolic value is defined with this setting It is possible to adjust the voltage and or th...

Page 57: ...the actual rpm value is defined with this setting It is possible to adjust the voltage and or the rpm value The setting is performed as described above under Alignment Ouput Analog Load value Setting range GAIN voltage 200 to 1000 mV in 2 mV steps GAIN rpm 30 to 130 rpm in 5 rpm steps Alignment Output ECG For eBike II REHA Ergometer with integrated ECG amplifier only The zeroline and the amplitude...

Page 58: ...ibration Static calibration Lift up tool for eBike basic comfort Parts of ergometer Lift up tool Dismount pedal crank screws Carefully lift the protective cover off the crank using a small flat screw driver Plug the extractor P N 2005737 001 onto the screw of the crank ...

Page 59: ...e screw with an open ended wrench 17 mm Remove the screw and the lock washer DO NOT REMOVE THE CRANKS Mounting the front support Tilt the ergometer onto the left side Mount the front support with the screw M10 x 20 Bring the ergometer back in an upright position ...

Page 60: ...gging the two angled piped in the appropiate counter parts marked 1 1 and 2 2 until the lock and clamp indexing plunger clicks into place Fix the connections by clockwise screwing the plunger Turn over the main support frame and shift it under the ergometer Hint The support frame should lie symmetrically under the ergometer ...

Page 61: ...attention to identic distance between crank thread and tube on both sides Insert the bushings through the frame into the cranks the shoulder have to fit completely into the crank check for equal distance on both sides Insert the long screws with a washer through the bushings and screw them tight into the cranks Mount the lifting arm marked 3 3 ...

Page 62: ...nd tighten it Lifting up the ergometer Stand in front of the ergometer Stabilize the ergometer with one hand at the handle Press the lever arm downwards until it lies flat on the bottom and the ergometer stands on the front support Caution During lift up the ergometer will move forward approx 20 cm ...

Page 63: ...ount the tilt protection 5 5 Carefully lift the lever arm of the main support frame whilst holding the ergometer at the handle bar until the ergometer chassis stands stable on the floor and the front support Disassemble the lift tool in reverse order Ergometer in lift up position with tilt protection mounted Warning If the ergometer is lifted up the tilt protection has always to be mounted and sec...

Page 64: ...tion support under the left castor see arrow b in illustration Then lift the eBike L EL at the front right hand side and place the calibration support under the right castor Caution The castors must be located between the two border strips see arrow c in illustration Extend the calibration stay to approx 835 mm and lock Hang the hook of the calibration weight adaptor into the eye of the strain gau...

Page 65: ... adhesive tape see e in illustration Lift the base a little and wait for the calibration stay to stop swinging With the aid of the calibration stay align the bottom edge of the connection box exactly horizontal To do so open clamping lever g and adjust top section Perform calibration see following section Caution Place a weight e g the tool chest on the baseplate of the calibration stay Make sure ...

Page 66: ...e ergometer has to be switched off and on again As soon as the self test appears on the ergometer display press the two softkeys synchronously and hold To prevent settings from being changed by chance the service menu is protected by a password Select 999 with the arrow keys and confirm with SET The service menu appears Select Static load calibration with the arrow keys and confirm with ENTER Self...

Page 67: ...ATTENTION The storage of the value needs some seconds Next step is the calibration of the gain factor for the strain gauge Static Load calibration OFFSET Static Load Calib OFFSET DMS Volt 147 mV Typ 130 170 mV Max 50 250 mV SET Static Load calibration OFFSET Storage in progress Static Load Calib DMS Volt 147 mV Wait max 20 sec Static load calibration GAIN Static Load Calibration GAIN DMS Volt 1676...

Page 68: ...wn and should be in the given range SET stores the GAIN value ATTENTION The storage of the value needs some seconds The successfull calibration is confirmed Remove the calibration weight Leave the menu by pressing ESC top right key Static Load calibration finished Static Load Calibration DMS Volt 1676 mV Calibr succesfull 8 kg calibration weight Calibration weight mounted 8 kg ...

Page 69: ...rt the calibration process with START Set the calibrator unit to 50 rpm Set the ergometer to 50 Watt via PC interface Wait approx 20 seconds until the load display is stable By using the arrow keys input the actual value of the calibrator 54 3 Watt at the ergometer Store the value with SET Dynamic load calibration waiting for START Dynamic Load Cal Start calibration Rotation 0 DMS Volt 147 mV LRE ...

Page 70: ... value of the calibrator e g 412 5 Watt at the ergometer Store the value with SET The successfull completion of the dynamic load calibration is confirmed Stop the calibration unit Dynamic load calibration 50 Watt at 110 rpm Dynamic Load Cal Set at 110 rpm 50 Watt Rotation 50 DMS Volt 327 mV LRE Load 50 Watt Dyn Load 55 0 Watt SET Dynamic load calibration 400 Watt at 110 rpm Dynamic Load Cal Set at...

Page 71: ... components Calibrated pressure measuring device mmHg manometer Compensating container with 500 ml volume min Handpump Adapter for eBike cuff connector Connecting tubes with t adaptors Calibrated pressure measuring device mmHg Compensating container example Hand pump Adapter for eBike cuff connector ...

Page 72: ...ervice menu and confirm with ENTER or SELECT Using the hand pump apply a fixed pressure to the system The actual pressure value is displayed Compare the displayed ergometer value with the value of the manometer The maximum allowed deviation is 3mmHg compensating container manometer hand pump Calibration setup NIBP Test of the internal NIBP pressure system NIBP Test NIBP pressure 123 mmHg CLOSE VAL...

Page 73: ...ff connector with a short tube The actual pressure value is displayed Press SET or START The zero value is stored Calibration with 200 mmHg Connect the complete measurement system hand pump manometer and compensating container as described under NIBP test Using the hand pump apply a fixed pressure of 200 mmHg to the system control by the external manometer Wait some seconds until the pressure is s...

Page 74: ...ometer Port 1 oder USB The software update is started in the service menu To activate the service menu the ergometer has to be switched off and on again As soon as the self test appears on the ergometer display press the two softkeys synchronously and hold To prevent settings from being changed by chance the service menu is protected by a password Select 999 with the arrow keys and confirm with SE...

Page 75: ...ns of the different electronic modules Grp name of the module Ap SW actually installed version if 0 no connection to module X XCUT of software version always 0 S No serial number of PCB if electronically available Press the right key again to display the installed bootloader versions of the different electronic modules Grp name of the module Op SW required software version has to be equal to the A...

Page 76: ...ftware connection As soon as the connection is established the START button becomes active or an error message appears Initiate the update by pressing START the update progress is displayed on the ergometer display and on the PC After the update is performed the ergometer has to be switched off and on again Software Update in progress Software Update is running ZME FLASH writing 12 Software Update...

Page 77: ...2018112 069 Rev F eBike II eBike II L eBike II EL 77 ...

Page 78: ...ers every 2 to 6 months depending on applicationpurpose Use of an universal high performance lubricant such as OKS 470 is recommended eBike comfort Check electrical saddle adjustment adjust minimum and maximum saddle height then return saddle to mid level Check correct indication of saddle height display Check handlebar adjustment angle Check that the clamping lever for handlebar adjustment angle ...

Page 79: ...connectors Check blood pressure measurement on test subject at rest Check blood pressure readings on LCD Expected result BP values are displayed and in proper range Pass Fail Start manual exercise test change load e g start 50 W increase 25 W Check load level and load indication Expected result Load changes as expected Pass Fail If ergometer is remote controlled from ECG recorder or PC ECG system ...

Page 80: ...A 3 Ground open Forward Polarity N A 4 Ground open Reverse Polarity N A 5 Neutral open Reverse Polarity SFC __________ μA Pass Fail 500 μA 6 Reverse Polarity NC __________ μA Pass Fail 100 μA Patient Leakage Current to Ground 3 1 Forward Polarity NC __________ μA Pass Fail 100 μA 2 Neutral open Forward Polarity SFC __________ μA Pass Fail 500 μA 3 Ground open Forward Polarity N A 4 Ground open Rev...

Page 81: ...ndlebar height eBike comfort only 6 Castors 7 Speed display RPM for patient 8 Adjustment of the saddle height eBike 100 only 9 Digital saddle height indication eBike comfort only 10 Power switch green button 11 Sockets for power cord and connection cables underside of ergometer 12 Levelling devices to adjust the ergometer to uneven floors eBike basic controls connections and indicators eBike comfo...

Page 82: ...te Turn the screws a in and tighten moderately Screw in clamping lever b by turning it clockwise and tighten moderately Taking off the clamping plate Mounting the handlebar Note When mounting the handlebar it should first point down until the control terminal has been installed Then bring the handlebar to the correct position and secure by tightening the clamping lever and the screws At the eBike ...

Page 83: ...using a small flat screw driver Plug the extractor P N 2005737 001 onto the screw of the crank Lose the screw with an open ended wrench 17 mm Remove the screw and the lock washer Turn the extractor and screw it strongly into the winding of the crank screw Use an open ended wrench 17 mm to remove the crank by screwing the bolt onto the axis ...

Page 84: ...ve top cover Use a small socket wrench or a round pin 3 mm to unlock the two plastic locks through the small holes left and ride side of top cover Remove the top cover Dismount the wheels Carefully remove the cover off the wheels using a small flat screwdriver ...

Page 85: ...pull off the wheel Threaded bar front Loose one of the two cover nuts with a socket wrench and pull the threaded bar out the housing Hint When mounting the wheels again check that the notch inside the plastic wheel has to be positioned outwards otherwise the cover will not snap into the wheel ...

Page 86: ...Rev F Remove screws connections Loose the screw connections front and rear on one side with a socket wrench The housing is screwed to the bottom of the metal frame on both sides of the eddy current brake remove these screws with a long socket wrench ...

Page 87: ...th sides of the housing Philips screw driver Remove the protective cover a at the top front part of the housing pull the bellows topwards and loose the Allen key screw b Pull the bellows at the saddle bar topwards and loose the Allen key screw Press the dark plastic insert to the outside ...

Page 88: ... 88 eBike II eBike II L eBike II EL 2018112 069 Rev F Separate the two housing parts with a flat screwdriver below the front and rear bellows snap in locks The housing can be removed ...

Page 89: ...move the housing of the ergometer Loose the limiting screw for the saddle pillar with an Allen key Remove the lever Loose the Philips screw Disassemble the clamping plate and the threaded plate The saddle pillar can be pulled out Mounting of the new saddle pilar ist performed in reverse way ...

Page 90: ...e to allow the cylindrical pin to be removed from the saddle shaft Remove clamping plate from saddle tube Pull saddle tube off the saddle motor Unscrew the 4 screws of the saddle guide tube and remove the tube Loosen bottom fixation of the saddle motor screw and nut are accessible via a hole at the bottom Remove the saddle motor The new motor is assembled in reverse order Perform a saddle calibrat...

Page 91: ...nit exchange The power supply is mounted in the rear of the ergometer Remove the mounting bracket 2 screws Exchange the power supply unit and fix the mounting bracket again Exchange the connecting cables at the main switch accordingly Fix the cables with a cable strap ...

Page 92: ... the clamping levers every 6 months depending on application purpose Use of an universal high performance lubricant such as OKS 470 is recommended Expected result All threads properly greased cranks rests and pedals OK Pass Fail Power on Using the power cord connect the device to the power line and turn it on with the power switch Wait for self test to end Check LCD e g for error messages Check in...

Page 93: ...oth displays and no noticeable noise Pass Fail Units with NIBP module Check cuff tubing and tubing connections microfon connectors Check blood pressure measurement on test subject at rest Check blood pressure readings on LCD Expected result BP values are displayed and in proper range Pass Fail Start manual exercise test change load e g start 50 W increase 25 W Check load level and load indication ...

Page 94: ...en Forward Polarity SFC __________ μA Pass Fail 500 μA 4 Ground open Reverse Polarity SFC __________ μA Pass Fail 500 μA 5 Neutral open Reverse Polarity SFC __________ μA Pass Fail 500 μA 6 Reverse Polarity NC __________ μA Pass Fail 100 μA Patient Leakage Current to Ground 3 1 Forward Polarity NC __________ μA Pass Fail 100 μA 2 Neutral open Forward Polarity SFC __________ μA Pass Fail 500 μA 3 G...

Page 95: ...tomatic blood pressure measurement only 3 Handgrip mounting and dismounting aid 4 Speed indication for patient 5 Control terminal 6 Power switch green button 7 Connectors for power cord and connection cables 8 Leg rest option 9 Castors locking 10 Tissue roll 11 Remote control for couch and saddle adjustment 12 Connection for blood pressure cuff on either side of the couch on systems with automatic...

Page 96: ...using a small flat screw driver Plug the extractor P N 2005737 001 onto the screw of the crank Loose the screw with an open ended wrench 17 mm Remove the screw and the lock washer Turn the extractor and screw it strongly into the winding of the crank screw Use an open ended wrench 17 mm to remove the crank by screwing the bolt onto the axis ...

Page 97: ... eBike II EL 97 Remove saddle Using the remote control move the saddle into a low position approx 15 cm distance to housing Pull the bellows back and remove the snap ring Pull out the retaining bolt and remove the saddle Remove the bellows ...

Page 98: ...ev F Remove cover sheet Loose the screws off the cover sheets and remove the covers on both sides of the housing Remove the clamping nuts Remove the housing Loose the retaining screws on both sides 3 each Pull the housing towards the top of the supine ...

Page 99: ... II L eBike II EL 99 Disconnect the protective earth connector Open both strain reliefs for the connecting cable patch cable ZME LRE Connecting cable ZME LRE is released Pull the housing sidewards to pass the crank axis and remove it ...

Page 100: ... connection box Remove 6 fixing screws Support the connecting box from below Disconnect the connecting cable to the mains switch at the modular connector a Disconnect the connecting cable LRE BKE from the LRE board CAT 5 Patchkabel The connecting box can be removed ...

Page 101: ...Replacing modules Replace power supply Switch the device off Disconnect the power cable After removing the housing loose the fixing screws of the power supply Disconnect both connectors an remove the module The new power module is mounted in reverse order ...

Page 102: ...BIKE L MOTOR PITCH ANGLE Compatible modules eBike II L No Spare Part number Designation 6 2018111 210 EBIKE L EL SADDLE MOTOR BCU SCU 7 2018111 211 EBIKE L MOTOR CONTROLLER 240V BCU 8 2018111 212 EBIKE L MOTOR CONTROLLER 120V BCU 9 2018111 215 EBIKE L REMOTE CONTROL BCU 10 2018111 217 EBIKE L MOTOR PITCH ANGLE BCU Replacement control unit motors and remote control This Parts are used on eBike II L...

Page 103: ...ace motor control unit Disconnect device from mains Disconnect all cables prior the disassembling of the unit using a suitable fork wrench or the special tool Troubleshooting Before replacing the motor control unit make sure that the motor is actually defective For this check use a new hand switch the hand switch for eBike EL can also be used and test the motorfunctions If the control unit does no...

Page 104: ...turning the connector bayonet fixing and pulling out Loose the fixing screws of the motor control unit Assembly of the new control unit is performed in reverse order Hint Do not interchange the motor connectors Cable and jacks are marked Following the exchange of a motor or after any malfunc tion the control unit has to be initialised To initialise the control unit both motors have to be moved in ...

Page 105: ... control unit Replace motor control unit Troubleshooting Before replacing the motor control unit make sure that the motor is actually defective For this check use a new hand switch the hand switch for eBike EL can also be used and test the motorfunctions If the control unit does not work it is out of order and must be replaced If it works the hand switch is defective 2018112 069 Rev F eBike II eBi...

Page 106: ...onnect the remote control Disconnect the saddle motor A and the motor longitudinal tilting M Remove the 6 screws on the upholstery Place the upholstery away Remove the 2 screws from the motor control 106 eBike II eBike II L eBike II EL 2018112 069 Rev F ...

Page 107: ...initialise the control unit both motors have to be moved in defined end positions The motor for the supine angle has to be retracted completely supine horizontal The saddle motor have to be extended completely Loose the fixing screws of the motor control unit Assembly of the new control unit is performed in reverse order 2018112 069 Rev F eBike II eBike II L eBike II EL 107 ...

Page 108: ... suitable fork wrench then pullout the jack plug Open cable tie and pull cable towards the saddlemotor Troubleshooting Before replacing the saddle motor make sure that the motor is actually defective For this check switch the hexagon socket of the saddlemotor from channel 2 to channel 1 If the motor does not work it is out of order and must be replaced If the motor works the motor control unit or ...

Page 109: ...wise as far as possible e g using a suitable fork wrench then pull out the jackplug Loosen cable tie of the connection cable Remove pins b c with self locking screw Remove lift motor Note Reverse the above steps to install the new lift motor Secure the connection cable to the motor controlunit with cable ties Troubleshooting Before replacing the lift motor for couch adjustment make sure that the m...

Page 110: ...ooting Before replacing the saddle motor make sure that the motor is actually defective For this check switch the hexagon socket of the saddlemotor from channel 2 to channel 1 If the motor does not work it is out of order and must be replaced If the motor works the motor control unit or the hand switch is defective Note To facilitate pulling in the new saddle motor cable attach a draw rope to the ...

Page 111: ...self locking screw Remove lift motor Note Reverse the above steps to install the new lift motor Secure the connection cable to the motor controlunit with cable ties Troubleshooting Before replacing the lift motor for couch adjustment make sure that the motor is actually defective For this check switch the hexagon socket of the liftmotor from channel 2 to channel 1 If the motor does not work it is ...

Page 112: ...xternal Speed Indicator eBike L ext Open connection box Disconnect connection cable for external speed indicator from LRE or from BDE devices with NIBP board Unscrew the screws securing the external speedindicator and remove the speed indicator by pulling it upward ...

Page 113: ... of the clamping levers every 6 months depending on application purpose Use of an universal high performance lubricant such as OKS 470 is recommended Expected result All threads properly greased cranks rests and pedals OK Pass Fail Power on Using the power cord connect the device to the power line and turn it on with the power switch Wait for self test to end Check LCD e g for error messages Check...

Page 114: ...both displays and no noticeable noise Pass Fail Units with NIBP module Check cuff tubing and tubing connections microfon connectors Check blood pressure measurement on test subject at rest Check blood pressure readings on LCD Expected result BP values are displayed and in proper range Pass Fail Start manual exercise test change load e g start 50 W increase 25 W Check load level and load indication...

Page 115: ...n Forward Polarity SFC __________ μA Pass Fail 500 μA 4 Ground open Reverse Polarity SFC __________ μA Pass Fail 500 μA 5 Neutral open Reverse Polarity SFC __________ μA Pass Fail 500 μA 6 Reverse Polarity NC __________ μA Pass Fail 100 μA Patient Leakage Current to Ground 3 1 Forward Polarity NC __________ μA Pass Fail 100 μA 2 Neutral open Forward Polarity SFC __________ μA Pass Fail 500 μA 3 Gr...

Page 116: ...ed indication for patient 5 Control terminal 6 Power switch green button 7 Connectors for power cord and connection cables 8 Leg rest option 9 Castors locking 10 Tissue roll 11 Upper handgrip adjustable 12 Remote control for couch and saddle adjustment 13 Armpit support adjustable 14 Drop section of upholstery 15 Connection for blood pressure cuff on either side of the couch on systems with automa...

Page 117: ...ank using a small flat screw driver Plug the extractor P N 2005737 001 onto the screw of the crank Loose the screw with an open ended wrench 17 mm Remove the screw and the lock washer Turn the extractor and screw it strongly into the winding of the crank screw Use an open ended wrench 17 mm to remove the crank by screwing the bolt onto the axis ...

Page 118: ...018112 069 Rev F Remove saddle Using the remote control move the saddle into a low position approx 15 cm distance to housing Pull the bellows back and remove the snap ring Pull out the retaining bolt and remove the saddle Remove the bellows ...

Page 119: ... 119 Remove cover sheet Loose the screws off the cover sheets and remove the covers on both sides of the housing Remove the clamping nuts Remove the housing Loose the retaining screws on both sides 3 each Pull the housing towards the top of the supine ...

Page 120: ... EL 2018112 069 Rev F Disconnect the protective earth connector Open both strain reliefs for the connecting cable patch cable ZME LRE Connecting cable ZME LRE is released Pull the housing sidewards to pass the crank axis and remove it ...

Page 121: ... connection box Remove 6 fixing screws Support the connecting box from below Disconnect the connecting cable to the mains switch at the modular connector a Disconnect the connecting cable LRE BKE from the LRE board CAT 5 Patchkabel The connecting box can be removed ...

Page 122: ... Replacing modules Replace power supply Switch the device off Disconnect the power cable After removing the housing loose the fixing screws of the power supply Disconnect both connectors an remove the module The new power module is mounted in reverse order ...

Page 123: ...L SCU 11 2018111 218 EBIKE EL MOTOR PITCH ANGLE SCU 12 2018111 219 EBIKE EL MOTOR TRANSVERSE ANGLE SCU This Parts are used on eBike II L up to serial NO 2014006999 This Parts are used on eBike II L from serial NO 2014007000 and higher Compatible modules eBike II EL No Spare Part number Designation 1 2018111 067 EBIKE L EL SADDLE MOTOR 2 2018111 094 EBIKE EL MOTOR CONTROLLER 240V 3 2018111 095 EBIK...

Page 124: ...gram motor control unit Exchanging the motor control unit Disconnect device from mains Disconnect all cables prior the disassembling of the unit using a suitable fork wrench or the special tool Troubleshooting Before replacing the motor control unit make sure that the motor is actually defective For this check use a new hand switch for eBike EL and test the motorfunctions If the control unit does ...

Page 125: ...tor bayonet fixing and pulling out Loose the fixing screws of the motor control unit Assembly of the new control unit is performed in reverse order Hint Do not interchange the motor connectors Cable and jacks are marked Following the exchange of a motor or after any malfunc tion the control unit has to be initialised To initialise the control unit both motors have to be moved in defined end positi...

Page 126: ...n see wiring diagrams 3 3 Exchange motor control unit with SCU Block diagram motor control unit Exchanging the motor control unit Troubleshooting Before replacing the motor control unit make sure that the motor is actually defective For this check use a new hand switch for eBike EL and test the motorfunctions If the control unit does not work it is out of order and must be replaced If it works the...

Page 127: ...nit is performed in reverse order Hint Do not interchange the motor connectors Cable and jacks are marked Following the exchange of a motor or after any malfunc tion the control unit has to be initialised To initialise the control unit both motors have to be moved in defined end positions The motor for the supine angle has to be retracted completely supine horizontal The motor for the supine angle...

Page 128: ...itable fork wrench then pullout the jack plug Open cable tie and pull cable towards the saddlemotor Troubleshooting Before replacing the saddle motor make sure that the motor is actually defective For this check switch the hexagon socket of the saddlemotor from channel 3 to channel 1 or 2 If the motor does not work it is out of order and must be replaced If the motor works the motor control unit o...

Page 129: ...orkwrench then pull out the jack plug Loosen cable tie of the respective connectioncables Remove pins 1a 1b lift motor for longitudinaladjustment and or pins 2a 2b lift motor forleft lateral tilt with self locking screw Remove lift motor Note Reverse the above steps to install the new lift motor Secure the connection cable to the motor controlunit with cable ties Troubleshooting Before replacing l...

Page 130: ...tie and pull cable towards the saddlemotor Troubleshooting Before replacing the saddle motor make sure that the motor is actually defective For this check switch the hexagon socket of the saddlemotor from channel 3 to channel 1 or 2 If the motor does not work it is out of order and must be replaced If the motor works the motor control unit or the hand switch is defective Note To facilitate pulling...

Page 131: ...1b lift motor for longitudinaladjustment and or pins 2a 2b lift motor forleft lateral tilt with self locking screw Remove lift motor Note Reverse the above steps to install the new lift motor Secure the connection cable to the motor controlunit with cable ties Troubleshooting Before replacing lift motor 1 or 2 for couch adjust ment make sure that the motor is actually defective For this check swit...

Page 132: ...xternal Speed Indicator eBike EL ext Open connection box Disconnect connection cable for external speed indicator from LRE or from BDE devices with NIBP board Unscrew the screws securing the external speedindicator and remove the speed indicator by pulling it upward ...

Page 133: ...2018112 069 Rev F eBike II eBike II L eBike II EL 133 ...

Page 134: ...Error bits set Bit 3 LCD communication error Bit 5 5V out of range Error code bits ZME Bit Error Reaction Display Info Troubleshooting 0 Illegal program jump Safety shut down restart eBike OFF ON if error persists replace ZME 1 Wrong SW configuration Warning perform software update 2 not used 3 LCD communication error Safety shut down check LCD connection replace LCD replace ZME PCB 4 24V out of r...

Page 135: ...rror persists replace strain gauge 4 24V out of range Internal error check power supply check connections power supply to LRE cable ZME LRE or ZME BDE BDE LRE 5 5V out of range Safety shut down restart eBike OFF ON if error persists replace LRE 6 Brake not released Safety shut down check connections LRE brake LRE strain gauge perform static calibration 7 Safety state mode Safety shut down restart ...

Page 136: ...lace pneumatic unit 4 Errror Calibration Module is switched off Internal error perform NIBP test perform NIBP calibration replace BDE PCB 5 Error offset Module is switched off Internal error perform NIBP test perform NIBP calibration replace BDE PCB 6 not used 7 not used 8 Safety state mode Module is switched off Internal error restart eBike OFF ON if error persists replace BDE 9 Setup not ready M...

Page 137: ...e motor MAE if error persists replace seat motor 2 Wrong cabling Module is switched off Internal error check cable sensor MAE check cable motor MAE replace cables 3 not used 4 24V out of range Warning check power supply check connections power supply to LRE cable LRE MAE replace MAE 5 not used 6 not used 7 Safety state mode Module is switched off Internal error restart eBike OFF ON if error persis...

Page 138: ...place COM 5 not used 6 not used 7 Safety state mode Module is switched off Internal error restart eBike OFF ON check mounting of COM DSUB connector check jumper on BKE if error persists replace COM 8 Setup not ready Module is switched off Internal error check mounting of COM DSUB connector check jumper on BKE restart eBike OFF ON if error persists replace COM 9 CAN zykl Timeout Module is switched ...

Page 139: ...tt oder 5 40 25 3 Watt 40 50 3 Watt 40 100 5Watt 70 200 10 Watt 90 500 25 Watt 100 600 30 Watt Prüfstand Type SN geprüft bis Prüf ID Nr Separates Protokoll Jahr der nächsten notwendigen messtechnischen Kontrolle 20_ _ Datum ______________________ Unterschrift __________________________ GE Medical Systems Information Technologies GmbH Munzinger Str 5 D 79111 Freiburg Germany Only valid for Germany ...

Page 140: ...gerät Patientensimulator für Blutdruck Sonstige Sonstige Sonstige Jahr der nächsten turnusmäßigen messtechnischen Kontrolle 20_ _ Datum ______________________ Unterschrift __________________________ Durchgeführte Prüfung Prüfergebnis ermittelte Werte Bemerkungen 1 Sichtprüfung i O n i O 2 Funktionsprüfung BDM Werte systolisch _________ mmHg Proband diastolisch _________ mmHg Pulsrate _________ min...

Page 141: ...2018112 069 Rev F eBike II eBike II L eBike II EL 141 ...

Page 142: ... 80x35 silver 1x adhesive label roll ø25 silver SPARE EBIKE B COVER SET R L US Service Kit casing eBike basic US kit 2018111 017 1x side panel eBike basic left with silk screened GE logo 1x side panel eBike basic right with silk screened GE logo 2x connection piece diam 10 length 100 mm 10x washer U4 3x9x0 8 4x screw M4x12 cyl 2x screw M4x30 cyl 4x screw M4x10 combi slotted 2x screw M4x25 cyl 2x h...

Page 143: ...sive label roll 80x35 silver 1x adhesive label roll ø25 silver SPARE EBIKE C COVER SET R L US Service Kit casing eBike comfort US kit 2018111 018 1x side panel eBike comfort left with silk screened GE logo 1x side panel eBike comfort right with silk screened GE logo 2x connection piece diam 10 length 100 mm 10x washer U4 3x9x0 8 4x screw M4x12 cyl 2x screw M4x30 cyl 2x screw M4x25 cyl 1x threaded ...

Page 144: ...COVER TOP BLUE US cover top GE blue for eBike comfort US 2018111 016 SPARE EBIKE C COVER TOP BLUE cover top GE blue for eBike comfort 2018111 205 SPARE EBIKE B COVER TOP GREY US cover top grey for eBike basic US 2018111 015 SPARE EBIKE B COVER TOP GREY cover top grey for eBike basic ...

Page 145: ...galvanized 1 x screw M6x20 1x setscrew M5x16 1x clamping lever M10x 40 for clamping of saddle handlebar 1 x compensating sheet 0 5 mm 1x high performance lubricant white approx 5 g SPARE EBIKE II B SADDLE TUBE Service Kit Assembled saddle tube w o motor for eBike II basic 2018111 014 1x saddle host with motor 55x55x75 1x straight pin diam 10x36 2x screw M6x16 SPARE EBIKE C SADDLE HOST Service Kit ...

Page 146: ...SPLY Window for saddle height display 2018111 005 1x base for saddle height display 2x screw M3x12 SPARE EBIKE BASE FOR SADDLE HEIGHT DSPLY Service Kit Base for saddle height display 2018111 145 1x saddle motor eBike II pre assembled 1x shoulder bolt M8x25 1x washer U 8 4x16x1 6 1x washer U10 5x20x2 1x nut M8 self locking SPARE EBIKE II C SADDLE MOTOR Service Kit saddle motor eBike II comfort ...

Page 147: ...LEVELLERS Service Kit Levelling device kit 2018111 019 2x plastic wheel diam 80 mm with slot 2x spring washer w o cap diam 12x28 mm galvanized white 2x cap for plastic wheels x lock washer diam 9 mm SPARE EBIKE WHEEL SET Service Kit Plastic wheels kit 2018111 020 3x cable strain relief size 1 black 4 5 5 7 mm 1x cable strain relief size 3 black 6 5 7 7 mm 4x screw M5x18 Li Kr SPARE EBIKE STRAIN RE...

Page 148: ...e Kit Control terminal adapter for eBike 2018111 026 SPARE EBIKE HANDLEBAR 500MM Handlebar with clamping opening width 500 mm 2018111 148 1x threaded plate 34x35x8 mm 1x clamping sheet 30x26x5 mm 1x clamping lever M10x 40 1x screw M6x20 1x screw M5x16 1x high performance lubricant white approx 5 g SPARE EBIKE II B CLMPG LEVER PLATE SET Service Kit Plate with threaded bushing and clamping lever inc...

Page 149: ...x cap for crank thread SPARE EBIKE SCREW SET FOR CRANKS Service Kit screw set for pedal kit 2018111 302 1x pedal E 10 left and right with rubber strap SPARE EBIKE II III PEDAL SET R L LARGE Pedal set 2018111 301 1x crank left square 170 mm 1x crank right square 170 mm 1x pedal E 10 left and right with rubber strap 2x combi slotted screw M8x1x18 2x lock washer 2x cap for crank thread SPARE EBIKE II...

Page 150: ... F3 2 2x screw M3x8 SPARE EBIKE RETURN SPRING Service Kit return spring with micro rope 2018111 032 1x strain gauge pre assembled for drive unit eBike 1x clamping plate 2x star washer F3 2 2x screw M3x10 1x micro rope short diam 0 81 mm 7x7 length 88 7 mm for eBike SPARE EBIKE STRAIN GAUGE Service Kit Strain gauge kit 2018111 150 1x drive unit eBike II replacement assembled 6x screw M8x20 6x star ...

Page 151: ...led 4x screw M3x8 4x plastic washer 3 2x7x0 5 SPARE EBIKE II B PCB ZME PCB ZME kit for eBike II basic 2017911 140 POWER CORD C EURO EBIKE II III Power cord black length 2 5 m German connector 2018111 226 1x power switch pre assembled for eBike II 1x cable connecting the power supply to the PCB LRE 1x cable tie SPARE EBIKE II POWER SWITCH Service Kit power switch with paneling for eBike II kit ...

Page 152: ... 2018111 160 1x PCB BDE SPARE EBIKE II C PCB BDE PCB BDE kit for eBike II with bloodpressure 2018111 161 1x blood pressure pump 1x input filter unit 2x sound absorber 1x check valve SPARE EBIKE II BP PUMP MODULE Service Kit blood pressure module pump unit for eBike II 2018111 209 1x BP module assembled incl PCB SPARE EBIKE II BASIC COMFORT BP MODULE Service Kit blood pressure module eBike II Basic...

Page 153: ...XCHANGE 2018111 164 1x control terminal eBike II PC replacement standard w o BP GE 2x screw M4x10 combi slotted SPARE EBIKE II TRMNL PC EXCHANGE 2018111 163 1x control terminal eBike II PC new with BP GE 2x screw M4x10 combi slotted SPARE EBIKE II TRMNL PC W BP NEW 2018111 162 1x control terminal eBike II PC new standard w o BP GE 2x screw M4x10 combi slotted SPARE EBIKE II TRMNL PC NEW ...

Page 154: ...ff 1x cable panel to microphone cable 1x tube coupling control terminal eBike II 1x tube 3x2 mm 285 mm length SPARE EBIKE II TRMNL PANEL W BP 2018111 172 SPARE EBIKE II TRMNL PC HOUSING BOTTOM 2018111 166 1x top shell for control terminal PC eBike II 1x membrane keypad PC with small display GE version 4x screw M3x12 SPARE EBIKE II TRMNL PC HOUSING TOP Service Kit control terminal housing PC eBike ...

Page 155: ...hilips 2x screw M6x16 2x screw M6x25 6x screw M8x20 cyl 2x screw M8x35 cyl 1x shoulder bolt 10 M8x25 1x knurled screw M3x8 5x washer F 3 2x6x0 4 12x washer U 4 3x9x0 8 10x washer Z 4 3x8x0 5 14x washer Z 8 4x15x0 8 1x washer U 8 4x16x1 6 1x washer U 10 5x20x2 1x hex nut M3 1x hex nut M4 2x hex nut M4 self locking 1x nut M8 self locking 4x nut M10 2x setscrew M5x16 1x setscrew M10x20 1x straight pi...

Page 156: ...ike II eBike II L eBike II EL 2018112 069 Rev F 2018112 069 MNL SVCE EBIKE II Servicing Instructions System eBike II English GE Healthcare eBike II eBike II L eBike II EL Service Manual 2018112 069 Revision F ...

Page 157: ...E CUFF ADULT LARGE TUBE 130CM BP cuff adults long version 130 cm tubing 2017911 010 SPLY EBIKE CUFF ADULT STD TUBE 130CM BP cuff adults 130 cm tubing eBike II basic comfort accessories 2017911 014 SPLY EBIKE CUFFADULT SMALL TUBE 130CM BP cuff adults small version 130 cm tubing ...

Page 158: ...ADDLE INFANT children s saddle black 2017911 015 SPLY EBIKE SADDLE MOUNT INFANT saddle mount for children s saddle eBike basic comfort 2017911 018 SPLY EBIKE SADDLE MOUNT HOR ADJUSTABLE saddle fixture horizontally adjustable 2017911 017 SPLY EBIKE SADDLE SPORT racing saddle black ...

Page 159: ...HTS RESERVED ERGOLINE LINDENSTR 5 72475 BITZ GERMANY REF 2017911 130 SPLY EBIKE II III USB DRIVER CD 2018111 227 2 pc combi screw M8 x 1 x 18 SW 14 mm 2 pc ribbed disk 2 pc terminal cap for pedal cranks 1 pc Allan key black width 5 mm for adjustment of crank length SPARE EBIKE CRANK SET ADJUSTABLE Adjustable cranks for eBike 2017911 132 SPLY EBIKE II COM MODULE 2017911 131 SPLY EBIKE II USB CABLE ...

Page 160: ... II eBike II L eBike II EL 2018112 069 Rev F 2017911 141 SPLY EBIKE II PWR CORD G2 UK 2017911 140 POWER CORD C EURO EBIKE II III 2017911 143 SPLY EBIKE II PWR CORD AC CHINA 2017911 142 SPLY EBIKE II PWR CORD A US ...

Page 161: ...2018112 069 Rev F eBike II eBike II L eBike II EL 161 2017911 145 SPLY EBIKE II PWR CORD I2 AUSTRALIA 2017911 144 SPLY EBIKE II PWR CORD M2 SOUTH AFRICA ...

Page 162: ...2018111 143 ERGOMETER EBIKE LIFT UP SERVICE TOOL Service tool Lift support for eBike 2005737 001 TOOL PEDAL PULLER VARIOBIKE 500 Pedal puller 2018111 135 SPARE EBIKE L EL CALIBRATION WEIGHT 8KG 8 kg calibration weight incl Hook eBike II basic comfort tools ...

Page 163: ...2018112 069 Rev F eBike II eBike II L eBike II EL 163 2017911 133 ERGOMETER EBIKE II BP ADD ON KIT Bloodpressure upgrade for use with eBike II eBike II basic comfort add on kits ...

Page 164: ...3 5 x 16 SPARE EBIKE L EL HOLDER FOR BELLOW SDLE Service Kit Holder for bellows saddle 2018111 065 1x bellows saddle grey for eBike SPARE EBIKE L EL BELLOWS FOR SADDLE 2018111 064 1x saddle for eBike L EL 1x socket pin1x locking washer SPARE EBIKE L EL SADDLE STANDARD Service Kit Saddle for eBike L EL eBike II L EL spare parts ...

Page 165: ...eft and right with rubber strap 2x combi slotted screw M8x1x18 2x lock washer 2x end cap for crank SPARE EBIKE II III CRANK PEDAL SET R L Service Kit Pedal set left and right cranks and pedals 2018111 302 1x pedal E 10 left with rubber strap 1x pedal home trainer right with Velco strap SPARE EBIKE II III PEDAL SET R L LARGE Service Kit Pedal set 2018111 067 1x saddle motor eBike L EL 1x stud screw...

Page 166: ... 1x drive unit eBike II assembled 6x screw M8x20 6x star washer 8 4x15x0 8 SPARE EBIKE II L EL DRIVE UNIT Service Kit drive unit new for eBike II 2018111 175 2x Strap 450 mm for pedal shoe with plastic clip 2x grub screw 8x setscrew M3x5 SPARE EBIKE II L STRAP PEDAL SHOE R L Service Kit strap for pedal shoe 2018111 029 2x combi slotted screw M8x1x18 2x lock washer 2x cap for crank thread SPARE EBI...

Page 167: ...ce Kit chain for drive unit 2018111 033 1x return spring with micro rope diam 0 6 mm 1x19 length 125 mm 1x clamping plate 2x star washer F3 2 2x screw M3x8 SPARE EBIKE RETURN SPRING Service Kit return spring with micro rope 2018111 032 1x strain gauge pre assembled for drive unit eBike 1x clamping plate 2x star washer F3 2 2x screw M3x10 1x micro rope short diam 0 81 mm 7x7 length 88 7 mm for eBik...

Page 168: ... II BP PUMP MODULE Service Kit blood pressure module pump unit for eBike II 2018111 160 1x PCB BDE SPARE EBIKE II PCB BDE PCB BDE kit for eBike II with bloodpressure 2018111 158 1x PCB BKE SPARE EBIKE II PCB BKE PCB BKE kit for eBike II 2018111 176 1x PCB LRE 3x screw M3x6 3x washer 3 2x6 SPARE EBIKE II L EL PCB LRE PCB LRE kit for eBike II L and EL only ...

Page 169: ...RE EBIKE II TRMNL PC EXCHANGE 2018111 163 1x control terminal PCplus new with BP module GE 2x screw M4x10 combi slotted SPARE EBIKE II TRMNL PC W BP NEW 2018111 162 1x control terminal PCplus new standard w o BP GE 2x screw M4x10 combi slotted SPARE EBIKE II TRMNL PC NEW 2018111 151 1x chain for drive unit SPARE EBIKE CHAIN Service Kit chain for drive unit ...

Page 170: ...5 mm length SPARE EBIKE II TRMNL PANEL W BP 2018111 172 SPARE EBIKE II TRMNL PC HOUSING BOTTOM 2018111 166 1x top shell for control terminal PC plus eBike II 1x membrane keypad PCplus with large display GE version 4x screw M3x12 SPARE EBIKE II TRMNL PC HOUSING TOP Service Kit control terminal housing PCplus eBike II top shell 2018111 165 1x control terminal PC plus replacement with BP module GE 2x...

Page 171: ...l terminal eBike II L EL 4x washer U4 3 4x screw M4x12 1x retaining piece control terminal eBike II L EL 2x spacer control terminal POM 1x screw M6x60 2x washer U6 4 1x nut M6 self locking SPARE EBIKE II L EL HOLDER FOR TERMINAL Service Kit Terminal bracket 2018111 081 1x bellows control terminal 2x retaining piece bellows control terminal 2x washer U4 3 2x screw 3 5x16 SPARE EBIKE L EL BELLOW FOR...

Page 172: ...tie holder 2x cable tie 2x retaining ledge for cover L EL 8x screw M3x6 6x screw M3x12mm SPARE EBIKE II L EL COVER WITH HANDHOLD Service Kit Casing eBike II L EL 2018111 179 1x foot external eBikeII L EL 4x leg grey 2x screw M4x10 SPARE EBIKE II L EL STND PANEL EXT TRMNL Service Kit Foot for external control terminal eBike II L EL 2018111 178 2 5m Flexa tube AD 19PG 13 5mm assembled with 2x screw ...

Page 173: ... 8 5 16 SPARE EBIKE L EL CASTOR SET Service Kit Castor locking 2018111 182 1x connection box for eBike II L EL preassembled 6x screw M3x12 2x screw M4x10 2x nut M4 3x washer 4 3x9 3x lock washer 4 3x8 1x screw M3X6 1x plastic washer 3 2x7 SPARE EBIKE II L EL CONNECTION BOX Service Kit Connection box for eBike II L EL 2018111 181 2x retaining ledge for cover L EL 8x screw M3x6 6x screw M3x12mm SPAR...

Page 174: ...1 1x paper dispenser for eBike L 2x washer U4 3 2x screw M4x12 1x tissue roll white SPARE EBIKE L HOLDER FOR PAPER ROLL 60 Service Kit Paper roll holder 60 cm for eBike 2018111 091 1x guide rail4x stand off diam 14 length 12 4x screw M6x30 SPARE EBIKE L EL GUIDE BAR SET Service Kit Guide rail 2018111 115 1x transportation lock eBike EL with sticker TOP 2x screw M10x30 SPARE EBIKE EL TRANSPORTATION...

Page 175: ...x screw M6x12 SPARE EBIKE EL FOOTBOARD LARGE RIGHT Service Kit Foot board large for eBike EL 2018111 111 1x foot board small 370x180x20 4x screw 5x17 with eBike L 1 each is used on the right and left with eBike EL 1 is used on the left 4x washer U6 4 SPARE EBIKE L FOOTBOARD SMALL RIGHT Service Kit Foot board small for eBike L EL 2018111 110 2x spacer sheets 4x threaded disk 30x6 4x washer U8 4 4x ...

Page 176: ... Service Kit Motor controller magnetic 120 V for eBike EL three channel 2018111 094 1x motor controller magnetic 230V 50Hz three channel 4x washer 4 34x screw M4x16 SPARE EBIKE EL MOTOR CONTROLLER 240V Service Kit Motor controller magnetic 230 V for eBike EL three channel 2018111 093 1x motor controller magnetic 120V 60Hz dual channel 4x washer 4 3 4x nut M4 SPARE EBIKE L MOTOR CONTROLLER 120V Ser...

Page 177: ...otor max 10 for eBike EL 2x stud screw SW19 2x washer U24 2x washer U13 4x ring Hostaform D 30 2x nut M12 self locking 3x cable tie 95 mm SPARE EBIKE EL MOTOR PITCH ANGLE Service Kit Motor longitudinal tilt for eBike EL 2018111 098 1x motor longitudinal tilt motor max 30 for eBike L 1x grommet black 2x stud screw SW19 4x washer U13 4x ring Hostaform D 30 2x nut M12 self locking SPARE EBIKE L MOTOR...

Page 178: ...TROLLER 240V SCU Service Kit Motor controller magnetic 230 V for eBike EL three channel 2018111 212 1x motor controller magnetic 120V 60Hz dual channel 2x screw M5 x 10 SPARE EBIKE L MOTOR CONTROLLER 120V BCU SCU Service Kit Motor controller magnetic 120 V for eBike L dual channel 178 eBike II eBike II L eBike II EL 2018112 069 Rev F 2018111 211 1x motor controller magnetic 230V 50Hz dual channel ...

Page 179: ... 2x nut M12 self locking 3x cable tie 95 mm SPARE EBIKE EL MOTOR PITCH ANGLE SCU Service Kit Motor longitudinal tilt for eBike EL 2018111 217 1x motor longitudinal tilt motor max 30 for eBike L 1x grommet black 2x stud screw SW19 4x washer U13 4x ring Hostaform D 30 2x nut M12 self locking SPARE EBIKE L MOTOR PITCH ANGLE BCU Service Kit Motor longitudinal tilt for eBike L Hand switch for eBike EL ...

Page 180: ...e Upgrade Kit Motor Controler for eBike EL GE Healthcare Instruction eBike EL Motor Controller Firmware Installation 2018111 341 A 17 March 2017 Introduction The eBike EL Motor Controller Firmware Update is intended for GE Healthcare Field Engineers The purpose of this manual is to update the firmware in the eBike EL Motor Controller from revision 11 to 12 Affected eBike Motor controller firmware ...

Page 181: ... knob eBike II EL 2018111 185 SPARE EBIKE II L EL INTERNAL TUBE SET Service Kit tubing kit complete all types of tubing 2018111 184 SPARE EBIKE II L EL INTRNL MIC CABLE Service Kit Microphone cable internal incl connections 2018111 183 1x connection angle cuff 1x tube coupling for connector panel 2x washer Z4 3 2x screw M4x12 SPARE EBIKE II L EL CONN RACK FOR CUFF Service Kit Connection angle for ...

Page 182: ... L EL STAR HANDLE M8 Service Kit Star knob eBike L EL M8 2018111 159 1x handgrip SPARE EBIKE L EL HANDGRIP Service Kit Handgrip 2018111 186 1x patch cable 0 25m 1x patch cable 0 5m 1x patch cable 1 0m 1x patch cable 1 5m 1x patch cable 5 0m 2x RJ 45 connector 1x patch cable LRE to external speed display SPARE EBIKE II L EL INTERNAL CABLE SET Service Kit patch cable kit complete all types of patch ...

Page 183: ...lamping board 30 x 8 length 34 mm 2x screw M4 x 12 for paper roll holder 2x washer U 4 3 x 12x 1 for paper roll holder 1x high performance lubricant white approx 5 g SPARE EBIKE EL HOLDER FOR HEAD SUPPORT Service Kit Holder for head rest chromeplated 2018111 117 1x leg rest left complete with upholstery grey mounted SPARE EBIKE EL LEG REST LEFT Service Kit Leg rest left for eBike L EL 2018111 118 ...

Page 184: ...tar knob M10x15 2x reinforcing angles 34x25x2 2x clamping board 30 x 8 length 34 mm 1x retaining screw M8x20 1x high performance lubricant white approx 5 g SPARE EBIKE II EL GUIDE TUBE ARMPIT SUPPORT Service Kit Guide tube armpit support eBike II EL only 2018111 122 1x armpit cushion grey eBike EL only 1x screw M8x25 SPARE EBIKE EL ARMPIT CUSHION Service Kit Armpit cushion grey eBike EL only 20181...

Page 185: ...th 34 mm 1x high performance lubricant white approx 5 g1x star knob M8x20 SPARE EBIKE II EL FIX TUBE HIP SUPPORT Service Kit support tubing hip support eBike II EL only 2018111 197 1x guide tube hip support 2x clamping plate 30x8 length 34 2x reinforcing angle 34x25x2 2x star knob M10x15 1x retaining screw M8x20 SPARE EBIKE II EL GUIDE TUBE HIP SUPPORT Service Kit Guide tube hip cushion eBike II E...

Page 186: ...L only 1x retaining screw M8x20 SPARE EBIKE EL DROP SECTION COMPLETE Service Kit Drop section complete with holder eBike EL only 2018111 130 1x upholstery couch cushion grey for eBike EL 5x screw M5x60 SPARE EBIKE EL COUCH CUSHION Service Kit Upholstery couch surface grey for eBike EL 2018111 129 1x upholstery couch cushion grey for eBike L 6x screw M5x60 SPARE EBIKE L COUCH CUSHION Service Kit Up...

Page 187: ...M3x6 SPARE EBIKE II L EL MAINS CONNECTOR SET Service Kit Mains connection kit 120V 230 V 2018111 188 SPARE EBIKE II L EL POWER SUPPLY Service Kit power supply module for eBike II L EL 1x power supply RPS 50 60 Hz 1x interface cable connecting LRE to power supply RPS 1x cable from power supply to modular connector 2x spacer 12 7 mm 2x screw 3 5x9 5 2x spacer M 3 12 mm 4x washer 3 2x6 mm 2x washer 3...

Page 188: ...assis 4x nut M4 1x connector 6 3 mm SPARE EBIKE II L EL EARTH CABLE SET Service Kit Earth cable kit for eBike II L EL 2018111 190 1x switch ON OFF round illuminated 5 cables different colours mounting material SPARE EBIKEL II EL POWER SWITCH Service Kit power switch complete for eBike II L EL 2018111 075 SPARE EBIKE CABLE MAINS EURO 5M Power cord grey length 5 m German connector GE Healthcare eBik...

Page 189: ...EBIKE CUFF ADULT LARGE TUBE 130CM BP cuff adults long version 130 cm tubing 2017911 010 SPLY EBIKE CUFF ADULT STD TUBE 130CM BP cuff adults 130 cm tubing eBike II L EL accessories 2017911 014 SPLY EBIKE CUFFADULT SMALL TUBE 130CM BP cuff adults small version 130 cm tubing ...

Page 190: ...b M8x20 SPLY EBIKE L EL ARM REST W HOLDER Arm rest with holder ebike L EL 2017911 040 1x leg rest left incl all accessories 1x leg rest right incl all accessories 1x retaining clamp for leg rests with 2 screws M4x12 2x spacer sheets 4x threaded disk 30x6 4x washer U8 4 4x screw M8x50 required for eBike L only 4x screw M8x70 required for eBike EL only SPLY EBIKE L EL LEG REST SET R L Leg rest set r...

Page 191: ...ted screw M8x1x18 2x lock washer 2x end cap SPLY EBIKE L EL PEDAL SHOE JD 30A Set Pedal shoe JD 30A set 2018111 223 8x screw M5x16 8x hexagon nut M5 SPARE EBIKE II MOUNTING STRAP FOR PEDAL SHOE JD 30A Mounting strap for pedal shoe JD 30A 2x strap including screws 2018111 224 8x screw M5x16 8x hexagon nut M5 SPARE EBIKE II MOUNTING STRAP WITH PLUG FOR PEDAL SHOE JD 30A Mounting strap for pedal shoe...

Page 192: ...GIES eBike II eBike III USB DRIVER 2017911 130 ERG705463 REV S4 2016 03 01 ASSEMBLED IN GERMANY 2016 ALL RIGHTS RESERVED ERGOLINE LINDENSTR 5 72475 BITZ GERMANY REF 2017911 130 SPLY EBIKE II III USB DRIVER CD 2017911 042 1x infusion stand for rail eBike L EL 1x Novo Grip clamping lever 1x star knob SPLY EBIKE II L EL INFUSION HOLDER Infusion holder incl fastening device ...

Page 193: ...v F eBike II eBike II L eBike II EL 193 2017911 150 SPLY EBIKE L EL PWR CORD F EURO 2017911 152 SPLY EBIKE L EL PWR CORD B US 2017911 151 SPLY EBIKE L EL PWR CORD G3 UK 2017911 153 SPLY EBIKE L EL PWR CORD I3C CHINA ...

Page 194: ... L eBike II EL 2018112 069 Rev F 2017911 157 SPLY EBIKE L EL PWR CORD H ISRAEL 2017911 156 SPLY EBIKE L EL PWR CORD D INDIA 2017911 155 SPLY EBIKE L EL PWR CORD I3 AUSTRALIA 2017911 154 SPLY EBIKE L EL PWR CORD M3 STH AFRICA ...

Page 195: ...PLY EBIKE L EL PWR CORD J SWITZERLAND 2017911 160 SPLY EBIKE L EL PWR CORD L ITALY 2017911 159 SPLY EBIKE L EL PWR CORD K DENMARK 2018111 225 2x T 2 0 AH 250 V 2x T 3 0 AH 250 V 2x T 4 0 AH 250 V 2x T 5 0 AH 250 V 2x T 8 0 AH 250 V SPARE EBIKE II L EL FUSE SET Fuse set ...

Page 196: ...Service Kit Tool for disassembly of motor cables 2018111 135 1 x weight 8 kg 1 x hook SPARE EBIKE L EL CALIBRATION WEIGHT 8KG Service Kit Calibration Weight 2018111 134 1 x calibration stay 2 x calibration support SPARE EBIKE L EL SUPPORT FOR CALIBRATION Service Kit Device Support for Calibration eBike II L EL tools ...

Page 197: ...x nut M4 6x screw M3x12mm 1x patch cable 0 25m 2x mounting anglel BDE eBike II L EL 2x tube connector Rectus 4x screw M4x12 2x tube 3x2 550mm 5x aluminium clips 1x BP cuff standard 1x Addendum for BP Module retrofit kit ERGOMETER EBIKE II L EL BP ADD ON KIT Service Kit Retrofit Kit Blood Pressure Module for eBike II L EL eBike II L EL add on kits ...

Page 198: ... 198 eBike II eBike II L eBike II EL 2018112 069 Rev F 2018111 125 2018111 201 2018111 198 2018111 197 2018111 201 2018111 103 2018111 126 Illustrations for special spareparts eBike EL ...

Page 199: ...2018112 069 Rev F eBike II eBike II L eBike II EL 199 2018111 122 2018111 159 2018111 105 2018111 200 2018111 195 2018111 201 2018111 201 2018111 196 2018111 103 ...

Page 200: ... 200 eBike II eBike II L eBike II EL 2018112 069 Rev F 2018111 103 2018111 119 2018111 118 2018111 133 ...

Page 201: ...ontain alcohol they may damage and or discolor the upholstery Disinfection Only the following disinfectants are approved for disinfec tion Schülke Mayr GmbH Antifect AF FF FD 10 Terralin 0 5 Quartamon Med B Braun Melsungen AG Hexaquart plus 0 5 5 0 Hexaquart S 1 5 5 0 Meliseptol Melsept SF 0 5 5 0 ECOLAB Incidin Foam Warning Shock Hazard Disconnect the device from the power line before cleaning Eq...

Page 202: ...o cleaning agents containing alcohol Clean the microphone with a cloth moistened with alcohol or soap water Allow the microphone to dry before reinserting it in its pocket Disinfection For disinfection spray a disinfectant sparingly on the cuff the tubing and the microphone After the contact time indicated by the manufacturer wipe all components dry Only the following disinfectants are approved fo...

Page 203: ...ules of the art by a Service Engineer authorized by ergoline Similarly the automatic sphygmomanometer in the control terminal must be checked and calibrated by an authorized specialist every two years to fulfill legal requirements The date of the next inspection is indicated on the inspection sticker attached next to the type plate on the ergometer Disposal Do not dispose the product described in ...

Page 204: ...30 to 130 RPM Deviation of Measured Load according to DIN VDE 0750 0238 25 60 watt 3 watt 60 400 watt 5 according to manufacturer 6 60 watt 3 watt 60 999 watt 5 Load Increments user programmable Internal Protocols 5 fixed incremental exercise test protocols e g WHO 10 user programmable protocols manual load control Permitted Patient Weight 160 kg 352 lbs max Saddle Height Adjustment continuous for...

Page 205: ...width 460 mm 18 in handlebar approx 575 mm 23 in height 900 mm to 1350 mm 35 53 in weight eBike basic approx 61 kg 135 lbs eBike comfort approx 69 kg 153 lbs Safety Standards DIN EN 60601 1 DIN EN 60601 1 2 DIN VDE 0750 238 Protection Class Degree of Protection II B to DIN EN 60601 1 MDD Classification class IIa to 93 42 EEC RF Emission class B to DIN EN 55011 5 0 DIN EN 60601 1 2 Environment oper...

Page 206: ...ation internal backup battery IEC CR 2032 3 V 230 mAh Braking Principle computer controlled eddy current brake with torque measurement speed independent to DIN VDE 0750 0238 Load Range 6 999 watt speed independent Speed Range 30 to 130 RPM see characteristic Deviation of measured load according to DIN VDE 0750 0238 25 60 watt 3 watt 60 400 watt 5 according to manufacturer 6 60 watt 3 watt 60 999 w...

Page 207: ... Weight minimum 800 x 2350 mm 31 x 93 in W x L 1720 mm 68 in H tilted 45 headrest extended maximum 800 x 2520 mm 31 x 100 in W x L tilted 0 horizontal headrest extended weight approx 110 kg 243 lbs Safety Standards DIN EN 60601 1 DIN EN 60601 1 2 DIN VDE 0750 238 Protection Class Degree of Protection I B to DIN EN 60601 1 MDD Classification class IIa to 93 42 EEC RF Emission class B to DIN EN 5501...

Page 208: ...on internal backup battery IEC CR 2032 3 V 230 mAh Braking Principle computer controlled eddy current brake with torque measurement speed independent to DIN VDE 0750 0238 Load Range 6 999 watt speed independent Speed Range 30 to 130 RPM see characteristic Deviation of measured load according to DIN VDE 0750 0238 25 60 watt 3 watt 60 400 watt 5 according to manufacturer 6 60 watt 3 watt 60 999 watt...

Page 209: ...ions Weight minimum 800 x 2100 mm 31 x 83 in W x L headrest retracted 1720 mm 68 in H tilted 45 maximum 1200 x 2600 mm 47 x 102 in W x L tilted 0 headrest extended weight approx 140 kg 309 lbs Safety Standards DIN EN 60601 1 DIN EN 60601 1 2 DIN VDE 0750 238 Protection Class Degree of Protection I B to DIN EN 60601 1 MDD Classification class IIa to 93 42 EEC RF Emission class B to DIN EN 55011 5 0...

Page 210: ...rement Error pressure readout error 3 mmHg readout resolution 1 mmHg Inflation Pressure 300 mmHg max during inflation the inflation pressure automatically adapts to patient s BP Inflation Rate between approx 6 seconds to 140 mmHg and approx 18 seconds to 300 mmHg Max Cuff Pressure 300 mmHg Cuff Deflation Method pulse dependent deflation rate approx 3 mmHg beat or approx 3 mmHg s Calibration calibr...

Page 211: ...raking torque control range black speed independent range to DIN VDE 0750 0238 black grey speed independent range of the eBike II ergometer Family of characteristics of load duration due to IEC 60601 1 Under permanent load conditions the shown load and pause times white have to be observed ...

Page 212: ... 212 eBike II eBike II L eBike II EL 2018112 069 Rev F ...

Page 213: ...2018112 069 Rev E eBike II eBike II L eBike II EL 213 Uncrating and basic assembly instruction eBike II L EL eBike II L EL Uncrating and basic assembly instructions Revision 2013 08 ...

Page 214: ... 214 eBike II eBike II L eBike II EL 2018112 069 Rev E ...

Page 215: ...ansport screws 216 Castors mounting 217 Outrigger mounting eBike EL 218 Footboard mounting 219 Paper roll holder 219 Setting up the eBike EL 220 Removing the transport lock eBike EL only 221 Power and ECG connections 221 Leg Rest Installation option 223 eBike II L EL uncrating instructions 209 Contents ...

Page 216: ... 216 eBike II eBike II L eBike II EL 2018112 069 Rev E 210 eBike II L EL uncrating instructions ...

Page 217: ...L EL uncrating instructions 211 General remarks 2 persons are needed to uncrate the eBike EL The following tools are recommended to uncrate and assemble the eBike EL cardboard cutter box end wrench open end wrench or socket wrench 17 mm metric Philips screwdriver Ph 2 ...

Page 218: ... 218 eBike II eBike II L eBike II EL 2018112 069 Rev E 212 212 eBike II L EL uncrating instructions ...

Page 219: ... uncrating instructions 213 remove PaCkinG Place the palett in a position with enough working space around it Cut the 2 straps with a cardboard cutter and remove them Cut the red seal and remove the red plastic inset Turn the white lock counter clockwise and remove it ...

Page 220: ... 220 eBike II eBike II L eBike II EL 2018112 069 Rev E 214 eBike II L EL uncrating instructions Cut the tapes at all 4 corners see pictures Lift the cardboard box and remove it ...

Page 221: ...e II L eBike II EL 221 eBike II L EL uncrating instructions 215 Take the accessories box out contains power cord and user manuals Cut the tapes of the packed remote control unpack it and place it at the head support of the eBike L EL ...

Page 222: ...ions Cut and remove the tapes fixing the 2nd accessories box This box contains 4 double castors with mounting screws and Allen key outrigger eBike EL only paper rol holder remove transPort sCrews 4 screws are fixing the eBike L EL on the palett Unscrew all 4 screws with a wrench 17 mm metric ...

Page 223: ...e eBike L EL 90 on the palett The corners of the base frame should overlap on both sides Screw all 4 double castors into the base frame using the included Allen key and tighten them carefully Lift the eBike L EL at the bottom end shift it onto the castors and position it on the floor Remove the palett with the cardboard ...

Page 224: ...s outriGGer mountinG eBike el Unscrew the 2 screws 17 mm with washer from the base frame Fix the outrigger with these screws wrench 17 mm and tighten them Take care to mount the outrigger in parallel to the base frame For transport to the final location screw the screw foot into the top position ...

Page 225: ...tBoard mountinG Unscrew 4 scres from the footboard and fix the board on the base frame PaPer roll holder Separate the paper roll from the holder Slide the right holder to the middle to free the screw hole Unscrew 2 screws from the head support and fix the paper roll holder with them onto the support ...

Page 226: ... E 220 eBike II L EL uncrating instructions Unpack the paper roll and insert it in the holder settinG uP the eBike el Move the eBike EL to the final location and apply the castor brakes Adjust the screw foot at the outrigger until it is flat on the floor ...

Page 227: ...is fixed to the base frame with 2 screws 17 mm Unscrew carefully both screws with a wrench and remove the transport lock Keep the transport lock and fixing screws because the transport lock must be reinstalled before transporting the ergometer Power and eCG ConneCtions Refer to the eBike EL user manual for installation of the power cable and ECG connection ...

Page 228: ... 228 eBike II eBike II L eBike II EL 2018112 069 Rev E 222 eBike II L EL uncrating instructions ...

Page 229: ...on Necessary tools Philips screwdriver Allen key 6 mm metric pipe wrench Contents of the set 1 2 2 leg rests 3 2 spacing plates 4 4 special thread washer 5 4 washer Ø 8 4 x 16 6A 4 screw M8x70 with hexagon socket eBike EL 6B 4 screw M8x50 with hexagon socket eBike L 7 1 retaining clamp for leg rests with 2 screws 8 1 installation instruction ...

Page 230: ...ews M8x50 for eBike L M8x70 for eBike EL with washers from top to bottom into the bore holes Make sure that the screws do not sqeeze the microphone cable and silicon tubing already in place Place a spacing plate onto the screws the short end pointing to the inside towards the padding and secure it with the special thread washers Holding the thread washers carefully e g with a pipe wrench tighten t...

Page 231: ...8112 069 Rev E eBike II eBike II L eBike II EL 231 eBike II L EL uncrating instructions 225 hexagon socket screw washer spacing plate special thread washer Install the two leg rests in the plug in devices ...

Page 232: ...E 226 eBike II L EL uncrating instructions installation oF retaininG ClamP The retaining clamp for the two leg rests is mounted onto the base frame Remove the plastic caps from the holes Screw the retaining clamp with 2 flat head screws M4x12 to the frame ...