Chapter 3. Creating Data Recorder Files
74
DMS Go Operator’s Manual
Note:
Setting the reading type to “extended” stores additional information with the thickness measurement, including: the time/date of
reading, time/date of last calibration, measurement mode, rectification, probe type, velocity and gain. Once stored, this additional
information is only accessible using UltraMATE®.
Custom Point
File
Reading Type
Set file style to Standard (normally recommended) or Extended.
Number of Locations
Quantity of locations
Location Label Length Sets the maximum length of the label identifying each location.
Points Per Location
Defines the number of points assigned to each file location.
Point Label Length
Sets the maximum length of the label identifying each point.
File Name
16 character file name, set using the virtual keyboard. No two files in the DMS Go memory
may have the same file name.
Units
Selects either inches or millimeters as the unit of measurement
T-MIN
Defines the minimum thickness limit. Measurements below this limit cause an alarm to signal.
T-MAX
Defines the maximum thickness limit. Measurements above this limit cause an alarm to signal.
Table 3: Items to be Defined for Each File Type (cont.)