A-8
DINAMAP
®
ProCare Monitor Service Manual
Revision C
2009381-001
Technical Specifications:
Specifications
‡The Masimo SET
®
SpO
2
parameter has been validated for low perfusion accuracy in bench top testing
against a Biotek Index 2 stimulator and Masimo’s simulator with signal strengths of greater than 0.02% and
a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or
minus, one standard deviation.Plus or minus one standard deviation encompasses 68% of the population.
Masimo
®
Sensor Accuracy
Sensor Model
SpO2 Range
70% - 100%
LNOP
LNOP-ADT
± 2 digits
LNOP-ADT Long
± 2 digits
LNOP-PDT
± 2 digits
LNOP-NEO
± 3 digits
LNOP-NEO PT
± 3 digits
LNOP-DCI (reusable)
± 2 digits
LNOP-DCSC (reusable)
± 2 digits
LNOP-DCIP (reusable)
± 2 digits
NRI25 (reusable)
± 2 digits
LNOP-DC195 (reusable)
± 2 digits
Resolution
Saturation (% SpO
2
)
1%
Pulse Rate (bpm)
1
Low Perfusion Performance
>0.02% Pulse Amplitude
Saturation (% SpO
2
) ±2 digits
and % Transmission >5%
Pulse Rate ±3 digits
Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The level of increase is
approximately equal to the amount of carboxyhemoglobin present. Dyes, or any
substance containing dyes, that change usual arterial pigmentation may cause
erroneous readings.
Sensor Light Source
Wavelength
Infrared: 905 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation
Infrared: 22.5 mW (max)
Red: 27.5 mW (max)
Factory Default Settings
SpO
2
(%)
HIGH: 100
SpO
2
(%)
LOW: 90
Sensitivity Mode
2 (for low perfusion-Default)
Averaging Time
12 seconds
FastSAT Mode
0 (for Off)
Masimo Patents
5,482,036; 5,490,505; 5,632,272; 5,685,299; 5,758,644; 5,769,785; 6,002,952;
6,036,642; 6,067,462; 6,206,830; 6,157,850, and international equivalents.
Summary of Contents for DINAMAP ProCare
Page 2: ......
Page 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Page 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Page 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Page 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Page 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Page 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 133: ......