Revision C
DINAMAP
®
ProCare Monitor Service Manual
4-15
2009381-001
Maintenance:
Parameter Level Functional Testing
Alaris Turbo Temp Probe Date Code
1. Turn Monitor off. Make sure the temp probe is properly stored in the probe
well.
2. Disconnect the temp probe cable from the monitor.
3. Connect Temp simulator; set to 80.2°F.
4. Turn the Monitor on.
5. Put the Monitor into temp monitor mode:
a. partially remove the probe from the well (stop when you hear the
Monitor beep).
b. Quickly re-insert the probe and remove again (you should hear two
beeps of a different tone).
c. A temperature value should appear quickly, if not, repeat Step 5b.
6. Record and verify the reading in the temp display is 80.2°F ±0.2°F.
7. Set the simulator to 98.6°F.
8. Record and verify the reading in the temp display is 98.6°F ±0.2°F.
9. Set the simulator to 107.8°F.
10. Record and verify the reading in the temp display is 107.8°F ±0.2°F.
11. Calibration is complete. If the monitor does no pass the calibration
verification then contact Technical Support.
Summary of Contents for DINAMAP ProCare
Page 2: ......
Page 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Page 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Page 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Page 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Page 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Page 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 133: ......