Revision C
DINAMAP
®
ProCare
Monitor Service Manual
A-7
2009381-001
Technical Specifications:
Specifications
Power Dissipation
Infrared: 22.5 mW (max)
Red: 30 mW (max)
Factory Default Settings
SpO
2
(%)
HIGH Alarm Limit: 100
SpO
2
(%)
LOW Alarm Limit: 90
Response Mode
2 (for Mode 2: Fast Response)
SatSeconds
™
0
NELLCOR
®
Patents
US Patent No. 4,802,486; 4,869,254; 4,928,692; 4,934,372; 5,078,136; 5,485,847;
5,743,263; 5,865,736; 6,035,223; 6,298,252; 6,463,310; 6,591,123; 6,675,031;
6,708,049; 6,801,797; Re. 35,122 and foreign equivalents.
Masimo SpO
2
Measurement Range
SpO
2
1 to 100%
Pulse Rate
25 to 240 beats/min
Perfusion Range
0.02 to 20%
Accuracy and Motion Tolerance
Saturation
Without Motion - Adult/Ped* 70 to 100% ±2 digits
Without Motion - Neonate*
70 to 100% ±3 digits
With Motion - Adult/Ped/Neo**†70 to 100% ±3 digits
Low Perfusion‡
70 to 100% ±2 digits
0 to 69% unspecified
Pulse Rate
Without Motion
25 to 240 beats/min ±3 digits
With Motion
normal physiologic range
25 to 240 beats/min ±5 digits
* The Masimo SET
®
SpO
2
parameter with LNOP-Adt sensors has been validated for no motion accuracy in human
blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70-100% SpO
2
against a
laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or
minus one standard deviation encompasses 68% of the population.
**The Masimo SET
®
SpO
2
parameter with LNOP-Adt sensors has been validated for motion accuracy in
human blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and
tapping motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a nonrepetitive motion between 1 to 5 Hz at
an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory
co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one
standard deviation encompasses 68% of the population.
†The Masimo SET
®
SpO
2
parameter with LNOP-Neo Pt sensors has been validated for neonatal motion
accuracy in human blood studies on neonates while moving the neonate’s foot at 2 to 4 cm against a
laboratory co-oximeter and ECG monitor. This variation equals plus or minus, one standard deviation.Plus or
minus one standard deviation encompasses 68% of the population.
Summary of Contents for DINAMAP ProCare
Page 2: ......
Page 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Page 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Page 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Page 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Page 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Page 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 133: ......