Maintenance:
Parameter Level Functional Testing
Calibration Procedures and Tests
NOTE:
All devices are tested and calibrated during manufacturing and are
certified for operation at installation.
To adequately test the safety and integrity of the ProCare Monitor, the
following test equipment is recommended:
12VDC power supply
IEC 601-1 approved leakage tester
Digital manometer (with range to 350 mmHg)
Stopwatch/timer (capable of measuring seconds)
Adult BP cuff, Neonate BP cuff, hose, inflation bulb, and associated tubing
Calibration kit (p/n 320246, available through GE Medical Systems
SpO
2
cable (for appropriate SpO
2
type, if SpO
2
is installed)
TE 1811 Temperature Probe Simulator (if TEMP is installed) available from
Alaris™
Printer paper (if PRINTER is installed)
3” diameter rigid cylinder (mandrel)
CAUTION
Calibration equipment should always be kept dry and free of
particulate matter. Moisture or foreign substances
introduced to the pneumatic system will likely cause
damage to the Monitor and/or the accessories.
Annual Procedures
Perform the following test procedures every 12 months, or whenever the
accuracy of any reading is in doubt.
Parameter Test Procedures
Complete the Test Record at the end of this section as tests are performed.
NOTES
This test is written so that a knowledgeable technician who is familiar with
the ProCare Monitor and the test equipment and will be able to follow the
test procedure.
To enter Service Mode press and hold
Cycle
button while pressing the
On/
Off
button.
Setup
1. Connect manometer to unit as shown.
2. ‘T’ an inflation bulb into the pneumatic setup.
Summary of Contents for DINAMAP ProCare
Page 2: ......
Page 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Page 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Page 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Page 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Page 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Page 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 133: ......