Technical Description:
Technical Specifications – CASE System
Dynamic range
320 mV, ±10 mV signal superimposed on ±150 mV DC offset
Resolution
4.88 µV/LSB @ 500 Hz
Noise
< 15 µV peak-to-peak noise over 0.01 to 150 Hz (-3 dB) bandwidth
Frequency response
- 3dB, display and writer
High pass filter
0.01 Hz (or 0.05 Hz, special use) with DC offset control
Low pass filter
20, 40, 100 or 150 Hz (selectable)
Line filter
50.0 or 60.0 Hz notch filter (selectable)
Baseline correction
Cubic spline algorithm
Artifact/baseline correction
Finite Residual Filter (FRF) algorithm
Common mode rejection
> 140 dB (123 dB with AC filter disabled)
Input impedance
> 10 M
Ω
@ 10 Hz, defibrillator protected
Patient leakage
< 10 µA
Pace detect
Orthogonal LA, LL, and V6; 750 µV @ 50 µs
Communication/storage
MUSE CV system compatible via diskette or SD card, network (optional)
MUSE CV Web compatible for retrieval, view, printing of MUSE CV
system data
Local storage: minimum 10 GB hard drive storage of complete ECG
record and test results
DVD-RW removable archive (optional)
PDF export of final reports
Microsoft Word export of configured reports
XML or Excel export of special data
Table 3. Display
Item
Description
Type
LCD
Resolution
1680 x 1050 (22'' monitor)
1280 x 1024 (19'' monitor)
Table 2. Performance Specifications – Data Acquisition (Continued)
Item
Description
Summary of Contents for CASE V6.7
Page 1: ...GE Healthcare CASE V6 7 Software Version 6 73 Operator s Manual 2060290 201 ENG Revision C ...
Page 4: ...CE 2 CASE Revision C 2060290 201 CE Marking Information For your notes ...
Page 15: ...Revision C CASE 1 1 2060290 201 1 Introduction ...
Page 16: ...1 2 CASE Revision C 2060290 201 For your notes ...
Page 30: ...1 16 CASE Revision C 2060290 201 Service Information ...
Page 31: ...Revision C CASE 2 1 2060290 201 2 Equipment Overview ...
Page 32: ...2 2 CASE Revision C 2060290 201 For your notes ...
Page 51: ...Revision C CASE 3 1 2060290 201 3 Preparing the Patient ...
Page 52: ...3 2 CASE Revision C 2060290 201 For your notes ...
Page 64: ...3 14 CASE Revision C 2060290 201 Attaching the Acquisition Module For your notes ...
Page 65: ...Revision C CASE 4 1 2060290 201 4 Selecting a Patient Record and Test Type ...
Page 66: ...4 2 CASE Revision C 2060290 201 For your notes ...
Page 73: ...Revision C CASE 5 1 2060290 201 5 Resting ECG ...
Page 74: ...5 2 CASE Revision C 2060290 201 For your notes ...
Page 108: ...5 36 CASE Revision C 2060290 201 Post Test Review For your notes ...
Page 109: ...Revision C CASE 6 1 2060290 201 6 Exercise Test ...
Page 110: ...6 2 CASE Revision C 2060290 201 For your notes ...
Page 171: ...Revision C CASE 7 1 2060290 201 7 Ambulatory Blood Pressure Measurement ...
Page 172: ...7 2 CASE Revision C 2060290 201 For your notes ...
Page 188: ...7 18 CASE Revision C 2060290 201 Post Test Review For your notes ...
Page 189: ...Revision C CASE 8 1 2060290 201 8 Spirometry Test ...
Page 190: ...8 2 CASE Revision C 2060290 201 For your notes ...
Page 208: ...8 20 CASE Revision C 2060290 201 Post Test Review For your notes ...
Page 209: ...Revision C CASE 9 1 2060290 201 9 File Management ...
Page 210: ...9 2 CASE Revision C 2060290 201 For your notes ...
Page 222: ...9 14 CASE Revision C 2060290 201 Working Temporarily with the Local Database For your notes ...
Page 223: ...Revision C CASE 10 1 2060290 201 10 Operation with Administrative Tools ...
Page 224: ...10 2 CASE Revision C 2060290 201 For your notes ...
Page 227: ...Revision C CASE 11 1 2060290 201 11 DICOM Interface ...
Page 228: ...11 2 CASE Revision C 2060290 201 For your notes ...
Page 233: ...Revision C CASE 12 1 2060290 201 12 System Settings ...
Page 234: ...12 2 CASE Revision C 2060290 201 For your notes ...
Page 296: ...12 64 CASE Revision C 2060290 201 System Configuration For your notes ...
Page 297: ...Revision C CASE A 1 2060290 201 Cleaning and Maintenance A Cleaning and Maintenance ...
Page 298: ...A 2 CASE Revision C 2060290 201 Cleaning and Maintenance For your notes ...
Page 301: ...Revision C CASE B 1 2060290 201 Miscellaneous B Miscellaneous ...
Page 302: ...B 2 CASE Revision C 2060290 201 Miscellaneous For your notes ...
Page 325: ...Revision C CASE C 1 2060290 201 Order Information C Order Information ...
Page 326: ...C 2 CASE Revision C 2060290 201 Order Information For your notes ...
Page 330: ...C 6 CASE Revision C 2060290 201 Order Information Order Information ...
Page 331: ...Revision C CASE D 1 2060290 201 Technical Description D Technical Description ...
Page 332: ...D 2 CASE Revision C 2060290 201 Technical Description ...