CARESCAPE V100 Vital Signs Monitor
5-7
Troubleshooting:
Alarm code interpretation
Exergen-specific error codes
If the scanner is unable to take a temperature determination, or has a low
battery, the monitor will display the ‘
E--
’ error code and generate an audible
alarm. The scanner may also display additional indicators on the LED display.
Exergen LED display
Condition
Description
HI
High Target
> 43
°
C (110
°
F)
LO
Low Target
< 16
°
C (61
°
F)
HI A
High Ambient
> 40
°
C (104
°
F)
LO A
Low Ambient
< 16
°
C (60
°
F)
bAtt
Low Battery
Replace battery soon.
(blank display)
Dead Battery
Replace battery.
Err
Processing Error
Restart scanner. If unit is defective, it will have to be returned to the
manufacturer. Contact GE Technical Support.
Summary of Contents for CareScape V100
Page 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Page 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Page 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Page 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Page 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Page 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Page 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Page 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Page 177: ......