3-16
CARESCAPE V100 Vital Signs Monitor
2048723-002C
Installation:
Host communication connector
DB15 connector pin assignments
Connection details
Host port connector (rear panel)
Communication protocol
GE offers resources to assist customers in the development of software to
exchange data with the monitor, including the Host Communications Reference
Manual and the DINAWIN Application Programming Interface library. Contact
your GE sales representative for further information.
Pin #
Function
1
Common
2
Inverted TTL Transmit Data
3
Inverted TTL Receive Data
4
+5 volts
5
No connection
6
No connection
7
Common
8
Remote Alarm
9
No connection
10
No connection
11
RS232 Transmit Data (TxD)
12
No connection
13
RS232 Receive Data (RxD)
14
No connection
15
No connection
Summary of Contents for CareScape V100
Page 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Page 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Page 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Page 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Page 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Page 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Page 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Page 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Page 177: ......