![GE CareScape V100 Service Manual Download Page 173](http://html.mh-extra.com/html/ge/carescape-v100/carescape-v100_service-manual_128562173.webp)
2048723-002C
CARESCAPE V100 Vital Signs Monitor
C-5
Electromagnetic compatibility (EMC):
Electromagnetic compatibility (EMC): CARESCAPE V100 monitor
Recommended separation distances
The table below provides the recommended separation distances (in meters)
between portable and mobile RF communications equipment and the
CARESCAPE V100 vital signs monitor.
The CARESCAPE V100 vital signs monitor is intended for use in the
electromagnetic environment on which radiated RF disturbances are controlled.
The customer or the user of the CARESCAPE V100 vital signs monitor can help
prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and
the CARESCAPE V100 vital signs monitor as recommended below, according to
the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [
d
] in meters (m) can be estimated using the
equitation applicable to the frequency of the transmitter, where
P
is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE:
These guidelines may not apply in all instances. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Separation distance in meters (m) according to frequency of transmitter
Rated maximum output
power of transmitter in
watts
150 kHz to 80 MHz
a
d
= 1.2
80 MHz to 800 MHz
a
d
= 1.2
800 MHz to 2.5 GHz
a
d
= 2.33
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
a
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
P
P
P
Summary of Contents for CareScape V100
Page 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Page 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Page 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Page 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Page 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Page 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Page 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Page 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Page 177: ......