2048723-002C
CARESCAPE V100 Vital Signs Monitor
4-9
Maintenance:
Long-term storage
6. Insert the battery into the compartment.
7. Replace the cover, help card tray, and screws. Insert the external DC power
converter plug into the external DC power socket and plug into an AC outlet.
NOTES
Error code ‘
E00
’
appears (MEMORY LOST) alerting you that the user
settings (including alarm limits and inflation pressure) and date/time will
go back to default values and all entries in clinical history are erased.
For units with Ohmeda TruSignal technology, it is important to verify the
Line Frequency mode setting.
Configured settings and time/date will
not
be lost when reset due to
pressing and holding the
On/Off
button regardless of whether the DC
charger is attached.
8. Reset the date/time and applicable user settings.
9. Verify the configuration settings prior to returning the monitor to clinical use.
Exergen temporal scanner battery
NOTE
If the Exergen scanner is not used regularly, remove the battery to prevent
possible damage due to chemical leakage.
Replacing the Exergen temporal scanner battery
1. Disconnect the scanner cable from the monitor.
a. Loosen the two thumbscrews from the scanner’s modular plug.
b. Unplug the scanner cable from the monitor’s host communication port.
Summary of Contents for CareScape V100
Page 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Page 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Page 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Page 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Page 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Page 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Page 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Page 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Page 177: ......