2048723-002C
CARESCAPE V100 Vital Signs Monitor
6-41
Parts lists and drawings:
Disassembly/reassembly of FRUs
7. Unplug speaker cable and temperature module cable (if installed) from
circuit board.
Printer
1. Lift the 2 black tabs and remove printer cable.
2. Remove printer assembly.
3. Unplug air hose from valve and filter.
4. Remove remaining 3 cable assemblies.
5. Locate and release retention tab and slide back sub chassis.