2048723-002C
CARESCAPE V100 Vital Signs Monitor
3-11
Installation:
Configuring the monitor
3. To set the unit of measurement to:
°
F
—
move the switch up
toward
the probe cone
°
C
— move the switch
away
from the probe cone
4. Replace the cover, and tighten the screw.
Temperature display time configuration settings
User selects number of minutes temperature values are displayed.
NOTE
This procedure applies to both the Alaris probe and the Exergen scanner.
1. With the monitor off, press and hold the
Menu
button at the same time as
pressing the
On/Off
button until the display test completes.
2. Press the
Menu
button until
tdt
(temperature display time) appears in the
Pulse Rate
window.
3. Use the
+
/
-
buttons to select the option.
4. To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press the
Menu
button.
Advanced configuration mode
Advanced Configuration mode is entered by holding the
Menu
button and the
-
button simultaneously while powering on with the
On/Off
button.
For a few seconds immediately after power up in this mode, the
Systolic
and
Diastolic
display windows will display the major and minor version codes. The
version codes are numbers that represent the letters of the English alphabet
which are designated to the currently loaded version of the monitor firmware
(e.g., 1 indicates A, 2 indicates B, etc.).
After a moment, this version information is cleared, and the monitor displays the
first page of configuration mode which simply displays
ACF
in the
Systolic
display window indicating that the monitor is in advanced configuration mode.
F
/
C
switch
°F
°C
Summary of Contents for CareScape V100
Page 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Page 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Page 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Page 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Page 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Page 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Page 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Page 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Page 177: ......