2048723-002C
CARESCAPE V100 Vital Signs Monitor
4-27
Maintenance:
Safety testing
Patient leakage current test (mains voltage on the applied part)
This procedure measures the leakage current from a mains voltage source into
the “Device Under Test” input connector.
The patient connector test body shorts all signals in the connector together.
Refer to the instructions contained with the safety analyzer to perform this test.
Connect the appropriate Test Body to the input connector of the device under
test.
Perform the test with normal and reverse polarity.
NOTE
*The MD represents the network and voltage measuring instrument and its
frequency characteristics per IEC 60601-1.
**Per IEC 60601-1, the resistance to protect the circuitry and the person
performing the test. The resistance must be low enough to accept currents
higher than the allowable values of the leakage current to be measured.
WARNING
Shock hazard. The following step causes high voltage at the
test body. Do not touch the test body.
NOTE
Perform this test once for SpO
2
and once for Temperature.
1. Configure leakage tester as follows:
Polarity – NORMAL
Neutral – CLOSED
GND (Earth) – CLOSED
2. Apply the AC mains voltage to the device under test.
3. Read the current leakage indicated on the tester and record the results on
the
Closed
(Keep cable length as
short as possible.)
**
Patient connector
Test Body
Power Cord
NORM
LINE
LINE
NEUTRAL
NEUTRAL
EARTH
AC MAINS
Device
Under
Test
EARTH
MD*
Leakage Tester
RVS
Summary of Contents for CareScape V100
Page 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Page 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Page 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Page 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Page 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Page 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Page 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Page 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Page 177: ......