2048723-002C
CARESCAPE V100 Vital Signs Monitor
5-5
Troubleshooting:
Alarm code interpretation
976
System
Power supply voltage has peaked
above 18 Volts (incorrect power
supply)
No
External power brick issue
979
System
Unknown power processor mode
received on power-up
No
Mains PWA issue
980
System
Heap memory exhausted
No
Mains PWA issue
984
System
Unused vector called
No
Return unit for evaluation
985
System
RTK 400hz timer re-entry
No
Return unit for evaluation
986
System
RTK 50hz timer re-entry
No
Mains PWA issue
989
System
RTK overrun
No
Mains PWA issue
994
System
Stack overflow
No
Mains PWA issue
999
System
Background task stalled
No
Mains PWA issue
E10
Printer
Printer no paper
Yes
No paper in printer
Printer problem
E11
Printer
Printer too hot
Yes
Printer problem
E13
Battery
Main battery low
Yes
Battery too low to
operate the unit
Charge battery
External DC source failed
Replace battery
BATTERY
LOW
Battery
Main battery is running low
Yes
Plug monitor in to recharge
E00
Battery
Memory lost
Yes
Usually noted after
changing batteries.
User settings and date/
time revert to default
settings.
All entries in clinical
history are erased.
For units with Ohmeda
TruSignal technology,
verify the Line Frequency
(LF) mode setting. Refer to
---
SpO
2
SpO
2
Sensor off finger
Yes
Reposition SpO
2
sensor
E20
SpO
2
SpO
2
sensor disconnected
Yes
Sensor disconnected
E21
SpO
2
SpO
2
replace sensor
Yes
Replace SpO
2
sensor
E25
SpO
2
SpO
2
lost pulse
Yes
Reposition SpO
2
sensor
Error code or
problem
Source
Definition
Can be
acknowledged
(silenced)?*
Probable source
Summary of Contents for CareScape V100
Page 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Page 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Page 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Page 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Page 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Page 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Page 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Page 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Page 177: ......