Page 130
Section 3.0 - Recommended Power Distribution System
Section 3.0
Recommended Power Distribution System
A dedicated feeder run from the facility main isolation transformer is recommended to power the
BrightSpeed CT scanner. If the scanner must be powered from an existing distribution transformer
and secondary feeder, such as the equipment distribution panel of an X-ray department, installation
with other X-Ray equipment that use rapid film changers should be avoided. These changers use
a large number of high powered, closely spaced exposures, which may coincide with the CT scan
and produce image artifacts.
WARNING
IF THE POWER FEED FOR THE A1/PDB PANEL IS NOT ON A DEDICATED
POWER TRANSFORMER ANY DEVICE THAT SHARES POWER FROM THAT
TRANSFORMER MAY BE IMPACTED BY INADVERTENT POWER
INTERUPTION CAUSED BY AN A1/PDB POWER PANEL FAULT.
CONVERSLY, THE OPERATION OF OTHER DEVICES SHARING THE POWER
TRANSFORMER MAY ALSO IMPACT THE OPERATION OF THE CT/PET
SCANNER.
If a dedicated distribution transformer is provided for the scanner, the minimum recommended
transformer size is 112.5 kVA, rated 2.4% regulation at unity power factor. For this configuration,
the minimum recommended feeder size and overcurrent protection device based on line voltage is
shown in
Minimum Feeder Wire Size.
In all cases, qualified personnel must verify that the transformer and feeder, at point of take-off, plus
the run to the BrightSpeed CT scanner meet all the requirements stated in this document.
SYSTEM CHARACTERISTICS:
Note:
(xxx):The value in the brackets is for 75kVA.
•
Maximum power demand = 90kVA
(75kVA)
@ 0.85 PF: at a Selected Technique of 140 kV,
380 mA
(300mA)
.
•
Continuous (average) power demand at maximum duty cycle = 20kVA
(16.7kVA)
.
•
Maximum allowable total source regulation is 6%.
•
Minimum recommended transformer size: 112.5 kVA
(93.75kVA)
, with 2.4% rated regulation
at unity power factor. Resultant maximum allowable feeder regulation is 3.4%.
The nominal line voltage must fall within one of the ranges listed below
Nominal Line Voltage
200
220
240
380
400
420
440
460
480
Hi-Line Limit, +10%
220
242
264
418
440
462
484
506
528
Lo-Line Limit, -10%
180
198
216
342
360
378
396
414
432
Continuous Line Current
58
52
48
30
29
27
26
25
24
Momentary Line Current
260
236
217
137
130
124
118
113
108
Maximum Line Current
289
262
241
152
144
137
131
126
120
Minimum Recommended
Circuit Protection Rating
150
150
150
110
110
100
100
90
90
Table 8-1 Nominal Line Voltage