GE H
EALT
HCA
RE
D
IRECTIO
N
514
1177
-100
, R
EVISION
14
B
RIG
H
T
S
PEED
E
LITE
, E
DGE
, E
XC
EL
: P
RE
-I
NST
A
LLA
TION
Page 145
Section 3.0
- Interconn
ect Runs,
W
iring and Cables
Table 9-10 GE Healthcare Supplied Cables for NIO16 Console (Optional, Long Run) (2281840-13)- UL Information
Note: For long cable kits GEHW Cat# is B70992RT, WSO Cat# is B7580JZ
RUN #
LENGTH,
ACTUAL
(USABLE)
UL CABLE INFORMATION
UL S
tyle
Flam.
Rating
Voltage
Rating
Voltage
Actual
Temp
.
Ra
ti
n
g
(C)
Dia.
mm (
inch)
# of
Con
d
Si
ze AWG
PULL SIZE
MM (INCHES)
ft
m
PART # DESCRIPTION
050 63
(55)
19.3
(16.76)
2343529
HVDC, PDU to Gantry
2587 FT4
600
+ & - 350VDC 90
19 (.751)
3
(2) 4
(1) 8
22 (.87) Dia
051 62.5
(55)
19
(16.76)
2343530
HVAC, PDU to Gantry
2587 FT4
600
440Y/254
90
15.3 (.604) 4
14
11.2 (.44) Dia
052 63
(58)
19.3
(17.56)
2343528
LVAC, PDU to Gantry
2587 FT4
600
208Y/120
90
13.8 (.542) 5
8
56.4 (2.22) Dia
053 80
(75)
24.5
(22.86)
2343531
LVAC, PDU to Console
2587 FT4
600
120VAC
90
12.3 (.483) 3
10
56.4 (2.22) Dia
055 63
(55)
19.3
(16.76)
2371450
Ground, PDU to Raceway
1284 VW-1
(FT-1)
600
0
105
15.5 (.608) 1
1/0
15.8 (.62) Dia
056 83
(75)
25.5
(22.86)
2371450-3 Ground, Raceway to Console 1283 VW-1
(FT-1)
600
0
105
11.9 (.467) 1
2
12.2 (.48) Dia
100 70
(62)
21.4
(18.86)
5120646
Signal, Gantry MSUB or
TGPU to PDU
FT-4
300
<30VDC
80
11.2 (.440) 25 22
17 x 58 (.68 x 2.30)
19 x 51 (.75 x 2.01)
101 86
(78)
26.35
(23.71)
5419981
Signal, Gantry MSUB or
TGPU to OC
FT-4
300
<30VDC
80
11.2 (.440) 25 22
17 x 58 (.68 x 2.30)
19 x 51 (.75 x 2.01)
102 86
(81)
26.3
(24.84)
2373436-2 Signal (LAN), Gantry to OC
1900 <30VDC
5.9 (.234)
8
24
15 (.59) Dia
103 80
(75)
24.3
(22.86)
5432019
Fiber Optic, Gantry to OC
N/A
N/A
1
N/A 10 (.39) Dia
Summary of Contents for BrightSpeed Elite
Page 169: ...CT...
Page 170: ...170 www gehealthcare com...