Fluorescence Custom
The Custom option is designed to allow for various customized combinations of wave-
lengths of the illuminating light and filters.
Please contact your GE representative for more details as the customized methods require
service.
Fluorescence Trans-UV
Capturing with the Fluorescence image capturing method option Trans-UV is performed
using the UV Trans Tray (transparent tray) and a gel sheet.
Fluorescence Trans-UV sample preparation
Follow the instructions below to prepare the sample.
Action
Step
Cut out a gel sheet slightly larger than the sample size.
1
Place the gel sheet on the UV Trans Tray (transparent tray).
2
Place the sample on the gel sheet.
3
Proceed with the capturing.
Note:
After use, wash the gel sheet with mild detergent, rinse with water then dry
well. A gel sheet can be reused about 20 times.
4
Follow the instructions below to perform fluorescence capture using Fluorescence UV.