Action
Step
In the Capture tab, select Fluorescence. The options available for Fluores-
cence capturing are displayed to the right on the screen.
1
Choose the UV.
2
Select an exposure option in the drop down menu:
3
•
Auto or
•
Semi-Auto or
•
Manual, enter the exposure time and press Set.
To use the full dynamic range of the image file, check High dynamic range.
4
To minimize exposure time, leave High dynamic range unchecked.
Note:
This function is only effective with Auto and Semi-Auto exposure setting.
If Auto or Manual exposure was selected, select Start.
5
Result: A progress bar appears on the screen indicating the time remaining
of the imaging process and the exposure time. Proceed to the final step of
this instruction.
If Semi-Auto exposure was selected, select Next.
6
Result: A low resolution preview appears.
Amersham Imager 600 Operating Instructions 29-0645-17 AD
141
5 Operation
5.5 Capture an image
5.5.4 Fluorescence capture