![GE AKTAprime plus Operating Instructions Manual Download Page 39](http://html.mh-extra.com/html/ge/aktaprime-plus/aktaprime-plus_operating-instructions-manual_128017039.webp)
1
Connect a sample loop between port 2 and 6 on the injection valve. Make sure that
the sample loop is large enough for your sample.
2
Connect a luer female/1/16" male union to port 3.
3
Fill a syringe with five loop volumes of deionized water or binding buffer.
4
Fit the syringe in the Luer union (1) and carefully inject the buffer.
3
1
5
Remove the syringe and fill it with at least two loop volumes of the sample.
6
Carefully inject the sample into the sample loop. Do NOT remove the syringe after
the injection because the loop might otherwise be emptied due to self-drainage or
air may be introduced in the flow path.
4.4
Performing a run
Selecting template and starting
the run
1
Select Templates in the main menu and press OK.
2
Select Application Template and press OK.
3
Select the appropriate template, for example His Tag Purification HisTrap, and press
OK.
4
Set the sample volume and press OK.
5
Press OK to start the purification run.
ÄKTAprime plus Operating Instructions 28-9597-89 AB
39
4 Operation
4.3 Preparations before start
Summary of Contents for AKTAprime plus
Page 1: ...GE Healthcare Life Sciences ÄKTAprime plus Operating Instructions Original instructions ...