WARNING
Hazardous biological agents during run. When using hazardous
biological agents, run System CIP and Column CIP to flush the
entire system tubing with bacteriostatic solution (e.g. NaOH) fol-
lowed by a neutral buffer and finally distilled water, before service
and maintenance.
WARNING
There must always be a sample loop connected to ports 2 and 6
of the injection valve. This is to prevent liquid spraying out of the
ports when switching the valve. This is especially dangerous if
hazardous chemicals are used.
CAUTION
Hazardous chemicals in UV flow cell. Make sure that the entire
flow cell has been flushed thoroughly with bacteriostatic solution,
for example NaOH, and distilled water, before service and mainte-
nance.
NOTICE
If the ÄKTAprime plus is kept in a cold room, cold cabinet or similar,
keep the system switched on in order to minimize the risk of con-
densation. (The UV lamp can be turned off to save lamp life time
when the system is not in use.)
NOTICE
When switching off the cold cabinet, make sure that you also switch
off the ÄKTAprime plus system and leave the door to the cold
cabinet open to avoid overheating.
ÄKTAprime plus Operating Instructions 28-9597-89 AB
21
2 Safety instructions
2.1 Safety precautions