ÄKTA start System Cue Card, 29-0240-42 AC

5

Prefill the flow path

Cleaning/prefilling the flow path including all inlets and 
outlets is performed using the system methods listed below. 
For instructions on how to prime the sample inlet tubing 

and sample loop, refer to 

ÄKTA start Operating Instructions

section 

5.7 Sample application

. For instructions on how to 

perform system methods, refer to 

ÄKTA start Maintenance 

Cue Card

.

Pump wash A

/

Pump wash B

 (prime buffer A & B inlets)

Washout fractionation tubing

 (clean fractionation 

tubing)

Column preparation

 (prepare/equilibrate a column)

System cleaning

 (clean the flow path using 1 M NaOH 

and DM water)

The flow path can also be prefilled by using a manual run.

Connect a column

For detailed instructions on how to connect a column, refer 

to 

ÄKTA start Operating Instructions

.

Instruction

Depending on column dimension, choose the 

appropriate location to place the column. The column 
holder rails are indicated in the image below. 

Attach an appropriate column holder to the column 

holder rail on the instrument.

Mount the column on the union connector if the column 

type requires a union.
Fix the column to the column holder, then connect the 
tubing to the top and to the bottom of the column.



1.

2.

3.

4.

Sample application

The sample can be applied to the column in two ways:

via Pump

: The sample can be applied directly from a 

sample container through the Sample valve

 as shown 

in the image below. It allows application of large volume 
of sample (> 5 ml) and facilitates unattended operation.

larger sample volumes (10 - 150 ml), use a Superloop™. 
The loop has to be connected to the 

Injection valve 

(ports 2 and 5

) as depicted below. 

To apply the sample from a loop:

Fill the loop with sample through port 

3 of the Injection 

valve (in Load

 position).

Switch the 

Injection valve to Inject

 position when 

prompted by the message below. Tap 

Continue

. The 

sample is loaded onto the column only when the 

Injection valve is in Inject

 position during the run. 

Switch the 

Injection valve to Load

 position when 

prompted by the message below. Tap 

Continue

.

Note: The system is in hold state while injecting the sample 
from loop. To ensure that the injection mark coincides with 
the injection event, acknowledge the message immediately 
after the action is performed. 
Note: For AC/IEX methods, when loading the sample 
through a loop it is advisable to empty the loop with 3× the 
loop volume to achieve good sample recovery.

Cold room operations

Preparation

Place the ÄKTA start instrument in the cold room.
Allow the instrument to stabilize at the temperature of 
the cold room.

1.
2.

via Loop

: A sample loop allows the injection of small 

sample volumes (25 μl - 5 ml) onto the column. For 

Summary of Contents for akta start

Page 1: ...raphy system intended for preparative purification of proteins at laboratory scale The system can be used for a variety of research purposes to fulfill the needs of the users in the academia and in th...

Page 2: ...w shows the locations of the modules placed on the wet side of the instrument The modules have the following functions To deliver the liquid in the system flow path and divert the flow as required To...

Page 3: ...chromatography runs Allows creation of methods from predefined chromatography templates like Affinity Gel Filtration Ion Exchange and Desalting Gives flexibility to create a customized method by drag...

Page 4: ...n Cond flow cell calibration 1 3 2 System preparation Preparation workflow Instruction Place the buffer bottles A B in the buffer tray and immerse the buffer inlet tubing A B in the respective buffer...

Page 5: ...ple application The sample can be applied to the column in two ways via Pump The sample can be applied directly from a sample container through the Sample valve as shown in the image below It allows a...

Page 6: ...Edit run Method run Prepare system Column preparation Set Run parameters to suit your requirement KTA start Pre defined System preparation Column preparation UNICORN start System control Method run En...

Page 7: ...ument Display or from UNICORN start Operation from the Instrument Display The following options are available from Run view screen View real time values UV absorbance Conductivity Flow rate Pressure T...

Page 8: ...it run in Run view screen when the run is in progress Use appropriate columns as indicated in the template names E g use 1 ml HiTrap column in case of selecting the Quick start AC IEX step 1 ml HiTrap...

Page 9: ...100 B CV NA NA NA NA NA NA 5 0 5 0 NA Re Equilibration Volume CV 5 0 5 0 NA NA 5 0 5 0 5 0 5 0 NA Fractionation Volume ml 1 0 3 0 0 5 3 0 1 0 3 0 1 0 5 0 4 0 the results will be saved only if the USB...

Page 10: ...ration Re equilibration Wash unbound Elution Pump wash Equilibration 100 50 5 CV 5 CV 5 CV 5 CV Sample B Fractionation collection 20 CV G r a d i e n t 0 t o 1 0 0 B GF 16 60 HiPrep Gel Filtration Chr...

Page 11: ...formation visit www gelifesciences com AKTA Selecting protocols columns and buffer Column supported Volume ml Method HisTrap FF 1 or 5 AC HisTrap FF crude 1 or 5 HisTrap HP 1 or 5 GSTrap FF 1 or 5 GST...

Page 12: ...alth Safety Declaration Form for Product Return or Servicing depending on whether the instrument is going to be serviced on site or returned for service respectively Copy the form you need from KTA st...

Reviews: