Calibration (Calibrate sensors)
To access this function you will be required to enter your PIN. This function is used to set new
corrected values for the sensors based on the outcome of the calibration procedures detailed in
section Calibration description. The
Calibration
sub-menu contains the following items:
- Sensor
Opens the
Calibration check
sub-menu showing the current values for Ps and Pt:
To enter new correction values for Ps:
1)
Tap the Ps panel. The Ps sensor correction panel opens.
2)
Follow the on-screen instructions. Tap the “tick” icon. A numbered keypad opens.
3)
Perform the same procedure to enter new correction values for Pt.
- Change CAL PIN
When selected, this function allows you to enter a new PIN code. To change your PIN:
1)
Tap the “Change CAL PIN” panel. A numbered keypad opens.
2)
Enter the new PIN number and tap the “tick” icon. You will be asked to confirm the change.
3)
Enter the new PIN again and tap the “tick” icon again. The numbered keypad closes and the
new PIN is now active.
4)
Tapping the “Cross” icon cancels the action and closes the numbered keypad without
changing the PIN.
Save/Recall ADTS Touch settings
Save ADTS settings
Recall ADTS settings
Delete ADTS settings
Copy all files from USB
Copy all files to USB
Restore last ADTS settings
Default Settings
ADTS Manuals
Safety Manual
User Manual
Customer documents
Test Sequences
Summary of Contents for ADTS542F
Page 19: ...Chapter 1 Introduction ...
Page 30: ...DUAL HAND TERMINAL EN English 19 K0553 Revision D Figure 1 8 ADTS Touch Slave ...
Page 31: ...Chapter 2 Installation ...
Page 34: ...PACKING FOR STORAGE OR TRANSPORTATION EN English 24 K0553 Revision D ...
Page 40: ...Chapter 3 Operation ...
Page 114: ...Chapter 4 Calibration ...
Page 123: ...Chapter 5 Maintenance ...
Page 132: ...Chapter 6 Testing and Fault Finding ...
Page 152: ...Chapter 7 Specification ...