Section 10-3 - Maintenance Task Schedule
10 - 3
Section 10-3 Maintenance Task Schedule
(cont’d)
NOTE:
PMs are not mandatory, the table above is for reference only.
NOTE:
May require specialized equipment to complete.
Table 10-2 Customer Care Schedule
Service at Indicated Time
Daily
Weekly Monthly
Per Facilities
QA Program
Notes
Clean Probe Holders
•
Clean Air Filter
•
more frequently depending on
your environment
Inspect AC Mains Cable
•
Mobile Unit Check Weekly
Inspect Cables and Connectors
•
Clean Console
•
Inspect Wheels, Casters, brakes and
Swivel Locks
•
Mobile Unit Check Daily
Check Control Panel Movement
•
Mobile Unit Check Daily
Console Leakage Current Checks
•
also after corrective
maintenance
Peripheral Leakage Current Checks
•
also after corrective
maintenance
Surface Probe Leakage Current Checks
•
also after corrective
maintenance
Endocavity Probe Leakage
Current Checks
•
also after corrective
maintenance
Transesphongeal Probe Leakage
Current Checks
•
also after corrective
maintenance
Surgical Probe Leakage
Current Checks
•
also after corrective
maintenance
Measurement Accuracy Checks
•
also after corrective
maintenance
Functional Checks
•
also after corrective
maintenance