GE H
EALTHCARE
VIVID FIVE
D
IRECTION
5341787-100, R
EVISION
8
B
ASIC
S
ERVICE
M
ANUAL
8 - 6
Section 8-4 - Loading Base Image Software
Section 8-4Loading Base Image Software
NOTE:
While it is believed to be unnecessary, It would not hurt to disconnect the system from the
network and remove all transducers.
NOTE:
Please ensure power cable is connected during system upgrade!
1.) Insert the disk labeled “System & Application Software” into the DVD ROM drive.
2.) Properly turn off the scanner by momentarily pressing the
Power On/Off
Switch. Select “Shut
Down” from the System Exit menu.
3.) If the system will not shutdown normally, hold down the
Power On/Off
Switch until the light
turns from green to amber.
4.) Turn on the system. System will detect the DVD-RW automatically.
5.) Press any key to continue when below message display as shown below.
Figure 8-65 Upgrade message
Summary of Contents for 5346769
Page 1: ...GE Healthcare LOGIQ C3 C5 Premium Basic Service Manual Part Number 5341787 100 Revision 8 ...
Page 2: ......