Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8 - 3
8-2-3
Air filter and Knob Screw
8-2-3-1
Tools
•
Common phillips screwdrivers
8-2-3-2
Preparations
•
Shut Down the System and disconnect the power cord.
8-2-3-3
Removal Procedure
1.) Unscrew the knob screw which fix air filter on the edge of power box, refer to
8-2-3 "Air filter and
Knob Screw" on page 8-3
.
2.) Pull rightward the knob screw out the air filter, refer to
8-2-3 "Air filter and Knob Screw" on page 8-3
.
3.) Lift down the air filter and plastic frame, refer to
8-2-3 "Air filter and Knob Screw" on page 8-3
.
4.) Remove the plastic frame , refer to
8-2-3 "Air filter and Knob Screw" on page 8-3
8-2-3-4
Mounting procedure
Install the new parts in the reverse order of removal.
Figure 8-63 Air filter and Knob Screw Disassembly
1)
2)
3)
4)