Preparing for ECG Recording: Electrode Application
Applying Plate (Limb) Electrodes
Plate electrodes are applied by means of a rubber strap, and electrode
paper is the recommended contact medium.
Moisten the electrode paper with tap water and place it between skin
and electrode.
Secure the electrode with the rubber strap, but do not hinder blood
circulation.
Applying Limb-Lead Electrodes
&$87,21
Use only silver-silver chloride electrodes, if the patient
may have to be defibrillated. (See
on page 8-4 for details.)
Applying Suction Electrodes (Chest)
Shave application points, if necessary.
Moisten the electrode paper with tap water and place it between skin
and electrode. Electrode cream or gel can be used instead of paper. On
hairy chests, the cream or gel improves adhesion of the electrodes.
Lead
Description
RA (white)
electrode on right arm
LA (black)
electrode on left arm
LL (red)
electrode on left leg
RL (green)
electrode on right leg
RA
RL
LL
LA
V
103A
Summary of Contents for MAC 1200
Page 1: ...MAC 1200 Operator s Manual Software Version 6 2012250 022 Revision B ...
Page 12: ...viii MAC 1200 Revision B 2012250 022 For your notes ...
Page 13: ...Revision B MAC 1200 1 1 2012250 022 1 The Basics ...
Page 14: ...1 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 27: ...Revision B MAC 1200 2 1 2012250 022 2 Controls and Indicators ...
Page 28: ...2 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 33: ...Revision B MAC 1200 3 1 2012250 022 3 Operating and Performance Tests ...
Page 34: ...3 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 45: ...Revision B MAC 1200 4 1 2012250 022 4 Preparing for ECG Recording ...
Page 46: ...4 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 61: ...Revision B MAC 1200 5 1 2012250 022 5 Recording in 12 Lead Mode ...
Page 62: ...5 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 87: ...Revision B MAC 1200 6 1 2012250 022 6 Recording in 6 Lead Mode ...
Page 88: ...6 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 93: ...Revision B MAC 1200 7 1 2012250 022 7 Arrhythmia Mode Recording ...
Page 94: ...7 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 102: ...7 10 MAC 1200 Revision B 2012250 022 Arrhythmia Mode Recording Recording For your notes ...
Page 103: ...Revision B MAC 1200 8 1 2012250 022 8 Pacemaker Patients Recording During Defibrillation ...
Page 104: ...8 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 107: ...Revision B MAC 1200 9 1 2012250 022 9 System Setup ...
Page 108: ...9 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 138: ...9 32 MAC 1200 Revision B 2012250 022 System Setup Direct ECG Transmission For your notes ...
Page 139: ...Revision B MAC 1200 10 1 2012250 022 10 Loading Chart Paper ...
Page 140: ...10 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 145: ...Revision B MAC 1200 11 1 2012250 022 11 Cleaning Disinfection and Maintenance ...
Page 146: ...11 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 149: ...Revision B MAC 1200 12 1 2012250 022 12 Troubleshooting ...
Page 150: ...12 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 153: ...Revision B MAC 1200 13 1 2012250 022 13 Technical Specifications ...
Page 154: ...13 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 165: ...Revision B MAC 1200 14 1 2012250 022 14 Order Information ...
Page 166: ...14 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 171: ...Revision B MAC 1200 A 1 2012250 022 A Entering Special Characters ...
Page 172: ...A 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 175: ...Revision B MAC 1200 Index 1 2012250 022 Index ...
Page 176: ...Index 2 MAC 1200 Revision B 2012250 022 ...
Page 179: ......