Revision B
MAC 1200
9-13
2012250-022
System Setup: Arrhythmia Mode
Muscle Filter Frequency
Cut-off frequency of the muscle filter: [40 Hz], 20 Hz.
The frequency range is indicated in the lower margin of the recording
strip.
0.08 – 40 Hz (40-Hz muscle filter enabled)
0.08 – 20 Hz (20-Hz muscle filter enabled)
0.08 – 100 Hz (muscle filter off).
Trend Recording
The slow trend recording of 5 mm/s automatically begins at program
start ([no]/yes).
Arrhythmia Data
The recorder will document arrhythmias in the following situations (all
[unequal], no).
All – the recorder documents all arrhythmias that occur.
Unequal – the recorder documents every arrhythmia that occurs that
is different from the preceding event.
No – arrhythmias are not documented.
Episodes
Final report includes episode report, with episodes listed by one of the
following criteria ([chron.], prio., ventr., no.).
In chronological order
According to priorities (see Table 3,
Ventricular beats only
No episode report
Summary of Contents for MAC 1200
Page 1: ...MAC 1200 Operator s Manual Software Version 6 2012250 022 Revision B ...
Page 12: ...viii MAC 1200 Revision B 2012250 022 For your notes ...
Page 13: ...Revision B MAC 1200 1 1 2012250 022 1 The Basics ...
Page 14: ...1 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 27: ...Revision B MAC 1200 2 1 2012250 022 2 Controls and Indicators ...
Page 28: ...2 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 33: ...Revision B MAC 1200 3 1 2012250 022 3 Operating and Performance Tests ...
Page 34: ...3 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 45: ...Revision B MAC 1200 4 1 2012250 022 4 Preparing for ECG Recording ...
Page 46: ...4 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 61: ...Revision B MAC 1200 5 1 2012250 022 5 Recording in 12 Lead Mode ...
Page 62: ...5 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 87: ...Revision B MAC 1200 6 1 2012250 022 6 Recording in 6 Lead Mode ...
Page 88: ...6 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 93: ...Revision B MAC 1200 7 1 2012250 022 7 Arrhythmia Mode Recording ...
Page 94: ...7 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 102: ...7 10 MAC 1200 Revision B 2012250 022 Arrhythmia Mode Recording Recording For your notes ...
Page 103: ...Revision B MAC 1200 8 1 2012250 022 8 Pacemaker Patients Recording During Defibrillation ...
Page 104: ...8 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 107: ...Revision B MAC 1200 9 1 2012250 022 9 System Setup ...
Page 108: ...9 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 138: ...9 32 MAC 1200 Revision B 2012250 022 System Setup Direct ECG Transmission For your notes ...
Page 139: ...Revision B MAC 1200 10 1 2012250 022 10 Loading Chart Paper ...
Page 140: ...10 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 145: ...Revision B MAC 1200 11 1 2012250 022 11 Cleaning Disinfection and Maintenance ...
Page 146: ...11 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 149: ...Revision B MAC 1200 12 1 2012250 022 12 Troubleshooting ...
Page 150: ...12 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 153: ...Revision B MAC 1200 13 1 2012250 022 13 Technical Specifications ...
Page 154: ...13 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 165: ...Revision B MAC 1200 14 1 2012250 022 14 Order Information ...
Page 166: ...14 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 171: ...Revision B MAC 1200 A 1 2012250 022 A Entering Special Characters ...
Page 172: ...A 2 MAC 1200 Revision B 2012250 022 For your notes ...
Page 175: ...Revision B MAC 1200 Index 1 2012250 022 Index ...
Page 176: ...Index 2 MAC 1200 Revision B 2012250 022 ...
Page 179: ......