MAINTENANCE: Checkout Procedures
Revision B
Dash 3000/4000 Patient Monitor
4-33
2000966-105
7.
Test the accuracy of these SpO
2
% settings.
8.
Test the accuracy of these PPR settings.
9.
Return the simulator to these conditions:
•
Set the SpO
2
% to 99.
•
Set the PPR to 70 beats/minute.
10. Set these alarms on the monitor:
•
Set SpO
2
% LO to 90.
•
Set PPR HI to 90.
11. Set PPR on the simulator to 100.
12. Make sure the PPR value on the monitor flashes, and it sounds an
alarm.
13. Return PPR on the simulator to 70.
14. Set SpO
2
% on the simulator to 80.3.
15. Make sure the SpO
2
% value on the monitor flashes, and it sounds an
alarm.
16. Disconnect the simulator cable from the module.
SpO
2
Settings
Simulator Setting
Displayed Value
99%
96 - 100%
90.6%
87 - 94%
80.3%
77 - 83%
PPR Settings
Simulator Setting
Displayed Value
70
67 – 73
100
97 – 103
Summary of Contents for Dash 3000
Page 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Page 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Page 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Page 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Page 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Page 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Page 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Page 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 231: ......
Page 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...