FIELD REPLACEABLE UNITS AND UPGRADES: Display Assembly Components
Revision B
Dash 3000/4000 Patient Monitor
8-13
2000966-105
Replacing the
Backlight Inverter
PCB
Complete the following steps for the monitor you are servicing.
Dash 3000
with display shield flex circuit
Dash 4000
1.
Remove the handle assembly. See page 8-7.
2.
Remove the display assembly from the main unit. See page 8-9.
3.
Remove and replace the backlight inverter. Follow the steps for
the monitor you are servicing.
a.
Flip over the display shield
flex circuit. Remove the male
end of the two snap rivets.
b.
Flip over the display shield
flex circuit. Disconnect the
display shield cable from the
backlight inverter PCB.
c.
Use a screw driver to remove
the female end of the two
snap rivets. Do not re-use
these rivets.
d.
Replace the defective
backlight inverter PCB.
a.
Remove the LCD color
display. See page 8-15.
b.
Disconnect the backlight
cable. Remove the two screws
anchoring the backlight
inverter PCB to the front
panel PCB.
c.
Replace the defective
backlight inverter PCB.
4.
Reassemble the display assembly.
plastic snap rivets (male end)
display shield cable
plastic snap rivets (female end)
two screws
backlight cable
Summary of Contents for Dash 3000
Page 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Page 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Page 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Page 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Page 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Page 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Page 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Page 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 231: ......
Page 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...