EQUIPMENT OVERVIEW: Technical Specifications
Revision B
Dash 3000/4000 Patient Monitor
2-15
2000966-105
GE Pulse Oximetry
(SPO2)
Masimo SET SPO2
Module
Parameters monitored:
Arterial oxygen saturation (SpO2) and peripheral pulse
rate (PPR)
Measurement Range
SPO2:
PPR:
50 - 100%
30 - 300 beats per minute
Accuracy
SpO2:
PPR:
Actual accuracy depends on probe. Please reference
manufacturer’s specifications.
± 2% (70 - 100% SpO2) ±1 standard deviation
± 3% (50 - 69% SpO2) ±1 standard deviation
± 3 beats per minute
Alarms:
User-selectable upper and lower limits for SpO2 and
PPR.
Display Messages
LOW QUALITY, PROBE IS OFF THE PATIENT, PROBE OR
MODULE MALFUNCTION, PULSE SEARCH
Measurement Range
Saturation:
Pulse Rate:
Perfusion:
30 to 100% SpO
2
25 to 240 beats per minute
0.02 to 20%
Accuracy
Displayed saturation data
Saturation, no motion:
Saturation, motion:
Pulse Rate, no motion:
Pulse Rate, motion:
Functional saturation
SpO
2
over the range 70 to 100%, below 69% is unspecified
±2 digits for adults and pediatrics, ±3 digits for neonates
SpO
2
over the range 70 to 100%, below 69% is unspecified
±3 digits for adults, pediatrics, and neonates
25 to 240 bpm, ±3 bpm
25 to 240 bpm, ±5 bpm
Alarms:
User-selectable upper and lower limits for SpO2 and PPR.
Patents
5482036, 5490505, 5632272, 5685299, 5769785,
6002952, 6036642, 6067462, 5,758,644; 6206830;
6157850
.
Summary of Contents for Dash 3000
Page 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Page 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Page 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Page 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Page 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Page 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Page 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Page 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 231: ......
Page 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...