EQUIPMENT OVERVIEW: Technical Specifications
Revision B
Dash 3000/4000 Patient Monitor
2-21
2000966-105
Environmental
Specifications
Physical
Specifications
NOTE: The system may not meet its performance specifications if
stored or used outside the manufaturer’s specified temperature
and humidity range.
Power requirements:
90-132VAC
190-264 VAC
50/60 Hz 2.0A
50/60 Hz 1.0A
Power consumption:
75 watts (fully loaded)
Cooling:
Convection
Heat dissipation:
240 Btu/hr (max)
Battery operation time:
General:
Battery age will affect operating time.
Operating conditions:
Ambient temperature:
While charging batteries:
Capnostat III sensor
Relative humidity:
Vibration:
Altitude:
0 to 40°C (32 to 104°F)
0 to 35°C (32 to 95°F)
10° to 40° C (50° to 104°F)
5 to 95% at 40°C
MIL-STD 810E, Method 514.4, Category 1
-610 to 4, 570 m (-2,000 to 15,000 ft)
Storage conditions (Do not
exceed):
Maximum:
Minimum:
CO2 Sensor:
Batteries:
70°C (158°F) at 95% relative humidity
–40°C (–40°F) at 15% relative humidity
–30 to 65°C (–22 to 149°F)
–20 to 60°C (–4 to 140°F)
Equipment Type:
Portable per IEC 60601-1
Height:
Dash 3000
Dash 4000
26 cm (10.25 inches)
27.38 cm (10.78 inches)
Width:
Dash 3000
Dash 4000
28 cm (11.0 inches)
29.26 cm (11.5 inches)
Depth:
Dash 3000
Dash 4000
20 cm (8 inches)
24.26 cm (9.55 inches)
Weight (without batteries)
Dash 3000
Dash 4000
5.08 kg (11.2 lbs)
5.53 kg (12.2 lbs)
Summary of Contents for Dash 3000
Page 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Page 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Page 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Page 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Page 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Page 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Page 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Page 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 231: ......
Page 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...