FIELD REPLACEABLE UNITS AND UPGRADES: Main Unit Components
Revision B
Dash 3000/4000 Patient Monitor
8-35
2000966-105
Verify Wireless LAN is
Enabled
Confirm the configuration of the optional Wireless LAN.
Activate the Boot Code:
1.
Hold down NBP GO/STOP and ZERO ALL.
2.
Press and release the Trim Knob control.
3.
Keep holding NBP GO/STOPNBP GO/STOP and ZERO ALL until
the Boot Code information appears on the display.
4.
Select SERVICE MENU -> SET CONFIGURATION.
5.
In the Configuration Menu, select Configure Wireless LAN.
6.
Select Set SSID to factory default.
7.
Verify Wireless LAN is Enable.
8.
Select Exit -> Exit. Reboot the monitor (press and release NBP GO/
STOP, ZERO ALL and the Trim Knob control.
Verify the Wireless LAN
ID Number
1.
Select MORE MENU -> MONITOR SETUP -> REVISION AND ID.
2.
Select Next -> Next.
3.
Verify that WIRELESS LAN ID is showing US 01 V4.65 xxxxxx. If no
ID is displayed,
•
recheck the setup in the Boat Loader menu, or
•
check if the Symbol PC card is installed correctly
Verify Wireless LAN
Communications
1.
Connect a Symbol Access Point with AC power and a Unity MC
Network node. Refer to the Wireless LAN (Symbol Access Point)
Installation and Service Manual. Power up the access point
2.
Disconnect the Ethernet line from the back of the monitor then
connect a patient simulator to display ECG waveforms on the
monitor’s display. Verify the unit name matches the unit name at the
central station. Admit the patient at the monitor.
3.
Verify waveforms are displayed at the central station for the
monitor.
4.
Connect the Ethernet jack to the back of the monitor. Verify that no
more than three seconds of drop out occurs at the central station
during the network switch over.
5.
With the monitor now operating on the Ethernet connection,
disconnect the Ethernet jack from the back of the monitor and verify
that no more than three seconds of drop out occurs at the central
station during the network switch over.
6.
Place the simulator into a Crisis alarm and verify the audio on the
speaker is functional and that the Alarm Light (if installed) is
flashing RED when in alarm.
7.
Discharge test patient and remove all cables. End of test.
Summary of Contents for Dash 3000
Page 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Page 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Page 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Page 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Page 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Page 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Page 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Page 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 231: ......
Page 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...