MAINTENANCE: Checkout Procedures
4-32
Dash 3000/4000 Patient Monitor
Revision B
2000966-105
Pulse Oximetry Tests
for Masimo SET
SPO2
1.
Set the pulse oximetry (SpO
2
) simulator power switch to the off
position.
2.
Connect the Nellcor-style SpO
2
simulator cable between the SpO
2
connector of the monitor and the SpO
2
simulator.
NOTE: Do not connect the simulator to an AC power supply for these
tests. Operate the simulator on battery power.
3.
Turn the SpO
2
simulator power switch OFF.
4.
Connect the simulator cable to the module.
5.
Set the simulator as follows:
NOTE: Use the white-colored values on the simulator.
•
Set the MODE to NELLCOR.
•
Set the SpO
2
% to 99.
•
Set the PRR to 70 beats/minute.
•
Turn the power ON.
6.
Verify the following are displayed at the monitor: (It might be
necessary to turn the SpO
2
parameter on.)
•
A waveform with an SpO
2
label.
•
An SpO
2
% reading between 97 - 100%.
•
A PRR reading between 67 and 72 beats per minute.
824A
Summary of Contents for Dash 3000
Page 11: ...Revision B Dash 3000 4000 Patient Monitor 1 1 2000966 105 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 19: ...Revision B Dash 3000 4000 Patient Monitor 2 1 2000966 105 2 EQUIPMENT OVERVIEW ...
Page 20: ...EQUIPMENT OVERVIEW 2 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 43: ...Revision B Dash 3000 4000 Patient Monitor 3 1 2000966 105 3 INSTALLATION ...
Page 44: ...INSTALLATION 3 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 51: ...Revision B Dash 3000 4000 Patient Monitor 4 1 2000966 105 4 MAINTENANCE ...
Page 52: ...MAINTENANCE 4 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 93: ...Revision B Dash 3000 4000 Patient Monitor 5 1 2000966 105 5 TROUBLESHOOTING ...
Page 94: ...TROUBLESHOOTING 5 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 123: ...Revision B Dash 3000 4000 Patient Monitor 6 1 2000966 105 6 CONFIGURATION ...
Page 124: ...CONFIGURATION 6 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 144: ...CONFIGURATION 6 22 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 145: ...Revision B Dash 3000 4000 Patient Monitor 7 1 2000966 105 7 CALIBRATION ...
Page 146: ...CALIBRATION 7 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 189: ...Revision B Dash 3000 4000 Patient Monitor 9 1 2000966 105 9 ASSEMBLY DRAWINGS ...
Page 190: ...ASSEMBLY DRAWINGS 9 2 Dash 3000 4000 Patient Monitor Revision B 2000966 105 For your notes ...
Page 231: ......
Page 232: ...GE Medical Systems Information Technologies g gemedicalsystem com ...