![GE HEALTHCARE E-PSMP Technical Manual Download Page 42](http://html1.mh-extra.com/html/ge-healthcare/e-psmp/e-psmp_technical-manual_3949047042.webp)
Module Frames and Modules
2 - 28
2062973-004
3.4.4 Invasive pressure tests
Perform the following test steps to both to the P1 and to the P2 invasive pressure channels.
NOTE: The ‘x’ in the PX refers to the invasive pressure channel being tested.
9.
Zeroing
a.
Ensure that the simulator InvBP output channel is configured to " 0 mmHg static
pressure" or "atmosphere"
b.
Zero the tested invasive pressure channel by pressing the related
Zero Px
module
key.
c.
Check that a message "Zeroing" followed by a message "Zeroed" is shown in the Px
parameter window.
10. Static Pressure
a.
Configure the simulator's InvBP output channel to "200 mmHg static pressure".
b.
Check that the flat pressure line appears on the related waveform field. The reading
in the parameter window shall be 200 ±10 mmHg.
NOTE: Recalibration is needed, if the measured value is not within the specification limits.
Calibrate the invasive pressure channels according to the instructions in
” section.
11. Pressure Waveforms
a.
Configure the simulator's InvBP output channel to "Arterial 120/80".
b.
Check that the pressure waveform for tested invasive pressure channel appears in
the Px waveform field and the Sys/Dia (Mean) pressure values are shown in the
related parameter window.
12. Mark this task as complete on the checkout form.
3.4.5 Temperature tests
Perform the following test step both to the T1 and to the T2 temperature channels.
NOTE: The 'x' in the Tx refers to the temperature channel being tested
13. Temperature test
a.
Check that reading in the Tx temperature channel is 37 ± 0.1 °C and that no error
messages are shown.
NOTE: Recalibration is needed, if the measured value is not within the specification limits.
Calibrate the temperature channels according to the instructions in
” section.
NOTE: You can alternatively perform the above test using a "temperature test set", P/N
884515-HEL. Test each temperature channel with both test plugs. The reading in the Tx
temperature channel shall be within
±
0.1°C of the nominal value.
b.
Mark this task as complete on the checkout form.
3.4.6 SpO2 tests
14. SpO2 test
a.
Connect the SpO2 finger probe to your finger. Wait until a pulse is found.
Summary of Contents for E-PSMP
Page 2: ......
Page 4: ...Module Frames and Modules 2062973 004 For your notes...
Page 6: ...Module Frames and Modules 1 ii 2062973 004...
Page 10: ...Module Frames and Modules 1 4 2062973 004 For your notes...
Page 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual...
Page 12: ......
Page 92: ...Module Frames and Modules 2 78 2062973 004 For your notes...
Page 94: ...Module Frames and Modules 2 80 2062973 004 For your notes...
Page 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual...
Page 96: ......
Page 124: ...Module Frames and Modules 3 26 2062973 004 For your notes...
Page 126: ......
Page 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004...
Page 160: ...Module Frames and Modules 4 32 2062973 004 For your notes...
Page 162: ...Module Frames and Modules 4 34 2062973 004 For your notes...
Page 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual...
Page 164: ......
Page 198: ...Module Frames and Modules 5 32 2062973 004 For your notes...
Page 199: ...6 Masimo Module E MASIMO Technical Manual...
Page 200: ......
Page 202: ...Module Frames and Modules 6 ii 2062973 004...
Page 218: ...Module Frames and Modules 6 16 2062973 004 For your notes...
Page 220: ...Module Frames and Modules 6 18 2062973 004 For your notes...
Page 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual...
Page 222: ......
Page 224: ...Module Frames and Modules 7 ii 2062973 004...
Page 236: ...Module Frames and Modules 7 12 2062973 004...
Page 244: ...Module Frames and Modules 7 20 2062973 004 For your notes...
Page 246: ...Module Frames and Modules 7 22 2062973 004 For your notes...
Page 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual...
Page 248: ......
Page 284: ...Module Frames and Modules 8 34 2062973 004...
Page 316: ...Module Frames and Modules 8 66 2062973 004 For your notes...
Page 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual...
Page 320: ......
Page 336: ...Module Frames and Modules 9 14 2062973 004 For your notes...
Page 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004...
Page 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004...
Page 380: ...Module Frames and Modules 9 58 2062973 004 For your notes...
Page 381: ...10 Single width Airway Module E miniC Technical Manual...
Page 382: ......
Page 410: ...Module Frames and Modules 10 28 2062973 004...
Page 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual...
Page 420: ......
Page 442: ...Module Frames and Modules 11 20 2062973 004 For your notes...
Page 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual...
Page 444: ......
Page 468: ...Module Frames and Modules 12 22 2062973 004 For your notes...
Page 470: ...Module Frames and Modules 12 24 2062973 004 For your notes...
Page 471: ...13 BIS Module E BIS Rev 01 Technical Manual...
Page 472: ......
Page 494: ...Module Frames and Modules 13 20 2062973 004 For your notes...
Page 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual...
Page 496: ......
Page 518: ...Module Frames and Modules 14 20 2062973 004 For your notes...
Page 520: ...Module Frames and Modules 14 22 2062973 004 For your notes...
Page 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual...
Page 522: ......
Page 558: ...Module Frames and Modules 15 34 2062973 004 For your notes...
Page 560: ...Module Frames and Modules 15 36 2062973 004 For your notes...
Page 561: ...16 Patient Data Module v2 Technical Manual...
Page 562: ......
Page 620: ...Module Frames and Modules 16 56 2062973 004...
Page 621: ......