![GE HEALTHCARE E-PSMP Technical Manual Download Page 395](http://html1.mh-extra.com/html/ge-healthcare/e-psmp/e-psmp_technical-manual_3949047395.webp)
Single-width Airway Module, E-miniC
10 - 13
2062973-004
Monitor configuration
1.
Configure the CO2 waveform field to the monitor screen with adequate priority.
2.
Select
CO2 Setup
tab in
Gases
menu and configure:
FiO2 level
: 21-40%
3.3.2 Procedure
1.
Gas sampling system leak test
Check the gas sampling system for possible leakages.
a.
Disconnect the module from the monitor.
b.
Connect a new Mini D-fend water trap to the module.
c.
Connect a new gas sampling line to the sampling line connector in the water trap.
d.
Connect the other end of the gas sampling line to a pressure manometer and a
pressure pump.
e.
Block the "Sample Gas Out" connector.
f.
Pump 100 mmHg ± 20 mmHg pressure to the gas sampling system. Let the pressure
stabilize for approximately 10 seconds.
g.
Check that the pressure reading does not drop more than 6 mmHg during 1 minute.
NOTE: The gas module shall be disconnected from the monitor during the leak test.
2.
Sample flow check
Check the sample flow rate.
NOTE: Gas measurement is not available during the first 1 minute after the module is
connected due to warm-up. Message 'Calibrating Gas Sensor' is shown in the waveform field.
Wait until warm-up is completed before proceeding with the next steps.
NOTE: The ambient temperature and air pressure influence the flow rate measured by the flow
meter. A flow meter, which has been calibrated at 21.1
°
C (70
°
F) and 760 mmHg (1013 mbar),
measures the flow rate correctly under the same conditions, i.e. in room temperature at sea
level. A flow rate correction as instructed by the manufacturer of the flow meter needs to be
performed when measuring flow rate under other conditions, for example in high altitude.
a.
Connect the module with the gas sampling line to the monitor.
b.
Connect the gas sampling line to the sampling line connector in the water trap.
c.
Connect the other end of the gas sampling line to a flowmeter.
d.
Check the sample flow rate reading from the flowmeter. The flow rate shall be within
the specification limit 150 ± 25 ml/min.
NOTE: Readjustment is needed, if the measured value is not within the specification limits.
Adjust the sample gas flow rate according to the instructions in section
NOTE: If the sampling pump is noisy it indicates possible problems with motor bearing. In this
case it is recommended to replace the noisy sampling pump with a new one.
3.
Zero valve operation
Test the zero valve functionality.
a.
Connect the gas regulator to the calibration gas container.
Summary of Contents for E-PSMP
Page 2: ......
Page 4: ...Module Frames and Modules 2062973 004 For your notes...
Page 6: ...Module Frames and Modules 1 ii 2062973 004...
Page 10: ...Module Frames and Modules 1 4 2062973 004 For your notes...
Page 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual...
Page 12: ......
Page 92: ...Module Frames and Modules 2 78 2062973 004 For your notes...
Page 94: ...Module Frames and Modules 2 80 2062973 004 For your notes...
Page 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual...
Page 96: ......
Page 124: ...Module Frames and Modules 3 26 2062973 004 For your notes...
Page 126: ......
Page 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004...
Page 160: ...Module Frames and Modules 4 32 2062973 004 For your notes...
Page 162: ...Module Frames and Modules 4 34 2062973 004 For your notes...
Page 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual...
Page 164: ......
Page 198: ...Module Frames and Modules 5 32 2062973 004 For your notes...
Page 199: ...6 Masimo Module E MASIMO Technical Manual...
Page 200: ......
Page 202: ...Module Frames and Modules 6 ii 2062973 004...
Page 218: ...Module Frames and Modules 6 16 2062973 004 For your notes...
Page 220: ...Module Frames and Modules 6 18 2062973 004 For your notes...
Page 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual...
Page 222: ......
Page 224: ...Module Frames and Modules 7 ii 2062973 004...
Page 236: ...Module Frames and Modules 7 12 2062973 004...
Page 244: ...Module Frames and Modules 7 20 2062973 004 For your notes...
Page 246: ...Module Frames and Modules 7 22 2062973 004 For your notes...
Page 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual...
Page 248: ......
Page 284: ...Module Frames and Modules 8 34 2062973 004...
Page 316: ...Module Frames and Modules 8 66 2062973 004 For your notes...
Page 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual...
Page 320: ......
Page 336: ...Module Frames and Modules 9 14 2062973 004 For your notes...
Page 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004...
Page 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004...
Page 380: ...Module Frames and Modules 9 58 2062973 004 For your notes...
Page 381: ...10 Single width Airway Module E miniC Technical Manual...
Page 382: ......
Page 410: ...Module Frames and Modules 10 28 2062973 004...
Page 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual...
Page 420: ......
Page 442: ...Module Frames and Modules 11 20 2062973 004 For your notes...
Page 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual...
Page 444: ......
Page 468: ...Module Frames and Modules 12 22 2062973 004 For your notes...
Page 470: ...Module Frames and Modules 12 24 2062973 004 For your notes...
Page 471: ...13 BIS Module E BIS Rev 01 Technical Manual...
Page 472: ......
Page 494: ...Module Frames and Modules 13 20 2062973 004 For your notes...
Page 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual...
Page 496: ......
Page 518: ...Module Frames and Modules 14 20 2062973 004 For your notes...
Page 520: ...Module Frames and Modules 14 22 2062973 004 For your notes...
Page 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual...
Page 522: ......
Page 558: ...Module Frames and Modules 15 34 2062973 004 For your notes...
Page 560: ...Module Frames and Modules 15 36 2062973 004 For your notes...
Page 561: ...16 Patient Data Module v2 Technical Manual...
Page 562: ......
Page 620: ...Module Frames and Modules 16 56 2062973 004...
Page 621: ......