![GE HEALTHCARE E-PSMP Technical Manual Download Page 345](http://html1.mh-extra.com/html/ge-healthcare/e-psmp/e-psmp_technical-manual_3949047345.webp)
Compact Airway Modules, E-CAiOVX, E-CAiOV, E-CAiO, E-COVX, E-COV
9 - 23
2062973-004
a.
Keep the gas sampling line connected to the Sampling line connector.
b.
Leave the other end of the gas sampling line open to room air.
c.
Connect the flowmeter to the Oxygen reference gas inlet with a piece of tubing.
d.
Check that the Reference Flow is within the following range:
31 - 45 ml/min with E-CO, E-COV, E-CAiO and E-CAiOV modules
27 - 40 ml/min with E-COVX and E-CAiOVX modules
7.
Zero Valve Operation
Test the zero valve functionality:
a.
Connect the gas regulator to the calibration gas container.
b.
Connect the end of the gas sampling line to the regulator on the gas container.
Leave the regulator overflow port open to room air.
c.
Select
Monitor Setup
>
Service Calibrations
.
d.
Enter the User Name and the Password and press
Enter
to get into the Calibrations
menu.
e.
Select
Gases
.
f.
Start feeding the specified calibration gas. Wait until the gas values shown in the
gas calibration menu rise approximately to the level indicated in the labeling of the
calibration gas container.
NOTE: The gas values in the Gas Calibrations menu is in percentages (%).
g.
Open the zero valve to room air by selecting
Zero valve Off
(zero position).
h.
Check that the CO2, N2O and anesthesia agent values drop back near 0% and the
O2 reading near 21% (room air).
i.
Stop feeding the calibration gas.
j.
Turn the zero valve back to the normal measurement position by selecting
Zero
valve On
(measurement position).
8.
Gas Calibration
Perform gas calibration according to the instructions in section
.
9.
Agent Identification
NOTE: Perform this test only for E-CAiO, E-CAiOV and E-CAiOVX modules.
Check agent ID unreliability:
a.
Feed the specified calibration gas for at least 30 seconds.
b.
Check that the anesthesia agent is identified as Desflurane and the
IDu
value
(=agent ID unreliability) shown in the
Monitor Setup / Service calibrations / Gases
menu is lower than 75.
If the value is higher, repeat the gas calibration and check the value again.
If the value does not drop below 75, contact GE Healthcare for factory repair.
10. Ambient Pressure
Use a barometer to check the operation of the absolute pressure sensor.
•
Check that the ambient pressure value shown in the Gas Calibrations menu does not
differ more than ± 10 mmHg from the value shown by the barometer.
NOTE: The ambient pressure value in the Gas Calibrations menu is in mmHg.
Summary of Contents for E-PSMP
Page 2: ......
Page 4: ...Module Frames and Modules 2062973 004 For your notes...
Page 6: ...Module Frames and Modules 1 ii 2062973 004...
Page 10: ...Module Frames and Modules 1 4 2062973 004 For your notes...
Page 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual...
Page 12: ......
Page 92: ...Module Frames and Modules 2 78 2062973 004 For your notes...
Page 94: ...Module Frames and Modules 2 80 2062973 004 For your notes...
Page 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual...
Page 96: ......
Page 124: ...Module Frames and Modules 3 26 2062973 004 For your notes...
Page 126: ......
Page 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004...
Page 160: ...Module Frames and Modules 4 32 2062973 004 For your notes...
Page 162: ...Module Frames and Modules 4 34 2062973 004 For your notes...
Page 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual...
Page 164: ......
Page 198: ...Module Frames and Modules 5 32 2062973 004 For your notes...
Page 199: ...6 Masimo Module E MASIMO Technical Manual...
Page 200: ......
Page 202: ...Module Frames and Modules 6 ii 2062973 004...
Page 218: ...Module Frames and Modules 6 16 2062973 004 For your notes...
Page 220: ...Module Frames and Modules 6 18 2062973 004 For your notes...
Page 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual...
Page 222: ......
Page 224: ...Module Frames and Modules 7 ii 2062973 004...
Page 236: ...Module Frames and Modules 7 12 2062973 004...
Page 244: ...Module Frames and Modules 7 20 2062973 004 For your notes...
Page 246: ...Module Frames and Modules 7 22 2062973 004 For your notes...
Page 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual...
Page 248: ......
Page 284: ...Module Frames and Modules 8 34 2062973 004...
Page 316: ...Module Frames and Modules 8 66 2062973 004 For your notes...
Page 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual...
Page 320: ......
Page 336: ...Module Frames and Modules 9 14 2062973 004 For your notes...
Page 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004...
Page 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004...
Page 380: ...Module Frames and Modules 9 58 2062973 004 For your notes...
Page 381: ...10 Single width Airway Module E miniC Technical Manual...
Page 382: ......
Page 410: ...Module Frames and Modules 10 28 2062973 004...
Page 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual...
Page 420: ......
Page 442: ...Module Frames and Modules 11 20 2062973 004 For your notes...
Page 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual...
Page 444: ......
Page 468: ...Module Frames and Modules 12 22 2062973 004 For your notes...
Page 470: ...Module Frames and Modules 12 24 2062973 004 For your notes...
Page 471: ...13 BIS Module E BIS Rev 01 Technical Manual...
Page 472: ......
Page 494: ...Module Frames and Modules 13 20 2062973 004 For your notes...
Page 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual...
Page 496: ......
Page 518: ...Module Frames and Modules 14 20 2062973 004 For your notes...
Page 520: ...Module Frames and Modules 14 22 2062973 004 For your notes...
Page 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual...
Page 522: ......
Page 558: ...Module Frames and Modules 15 34 2062973 004 For your notes...
Page 560: ...Module Frames and Modules 15 36 2062973 004 For your notes...
Page 561: ...16 Patient Data Module v2 Technical Manual...
Page 562: ......
Page 620: ...Module Frames and Modules 16 56 2062973 004...
Page 621: ......