![GE HEALTHCARE E-PRESTN Technical Manual Download Page 61](http://html.mh-extra.com/html/ge-healthcare/e-prestn/e-prestn_technical-manual_571893061.webp)
Hemodynamic Modules E-PRESTN, E-RESTN, E-PRETN, E-PSMP, E-PSM
2 - 47
2062973-004
Weak pulsation
PW
All invisible reasons, weak
oscillation, or poor performance
like algorithm etc. technically
related reasons.
Weak or unstable oscillation
signal due to:
improper cuff position or
attachment
weak or abnormal blood
circulation
slow heart rate associated with
artifacts
problems with measurement of
diastolic pressure
small air leakage
insufficient data due to artifacts.
Requires immediate correction
i.e. changed cuff position etc.
Cuff overpressure
PW
Cuff squeezed during
measurement and exceeding
pressure safety limits.
Check cuff and tubes. Restart the
measurement.
Unable to measure Sys
PW
Systolic pressure probably higher
than maximum inflation pressure
or artifacts in systolic area.
Automatic retrial with increased
pressure.
Unable to measure Dia
Unable to meas.
PW
Inflation pressure incorrect or
artifact preventing measurement.
Message indicates failed
measurement (
Sys
,
Dia
).
Non-Technical. Check the patient
status and cuff placement.
Perform a new measurement
Unstable zero pressure
PW
NIBP: Unstable zero pressure
when cuff inflated.
Retry the measurement. If the
problem persists, check the
internal tubing and recalibrate
NIBP.
Incorrect infl. limits
PW
Adult or child cuff is used, but the
selected infant mode restricts the
inflation pressure too low to be
able to measure the blood
pressure.
Adjust NIBP setup.
Long measurement time
PW
Measurement time over 2 min
(adult and child mode) and 1 min
15 s – 1 min 20 s (infant mode).
New NIBP measurement is
started automatically.
Control measurement
PW
Pressure alarm limit has been
violated. New measurement is
performed automatically to
check the blood pressure.
New NIBP measurement is
started automatically.
Message
Location
Possible causes
Possible solutions
Summary of Contents for E-PRESTN
Page 2: ......
Page 4: ...Module Frames and Modules 2062973 004 For your notes ...
Page 6: ...Module Frames and Modules 1 ii 2062973 004 ...
Page 10: ...Module Frames and Modules 1 4 2062973 004 For your notes ...
Page 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual ...
Page 12: ......
Page 92: ...Module Frames and Modules 2 78 2062973 004 For your notes ...
Page 94: ...Module Frames and Modules 2 80 2062973 004 For your notes ...
Page 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual ...
Page 96: ......
Page 124: ...Module Frames and Modules 3 26 2062973 004 For your notes ...
Page 126: ......
Page 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004 ...
Page 160: ...Module Frames and Modules 4 32 2062973 004 For your notes ...
Page 162: ...Module Frames and Modules 4 34 2062973 004 For your notes ...
Page 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual ...
Page 164: ......
Page 198: ...Module Frames and Modules 5 32 2062973 004 For your notes ...
Page 199: ...6 Masimo Module E MASIMO Technical Manual ...
Page 200: ......
Page 202: ...Module Frames and Modules 6 ii 2062973 004 ...
Page 218: ...Module Frames and Modules 6 16 2062973 004 For your notes ...
Page 220: ...Module Frames and Modules 6 18 2062973 004 For your notes ...
Page 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual ...
Page 222: ......
Page 224: ...Module Frames and Modules 7 ii 2062973 004 ...
Page 236: ...Module Frames and Modules 7 12 2062973 004 ...
Page 244: ...Module Frames and Modules 7 20 2062973 004 For your notes ...
Page 246: ...Module Frames and Modules 7 22 2062973 004 For your notes ...
Page 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual ...
Page 248: ......
Page 284: ...Module Frames and Modules 8 34 2062973 004 ...
Page 316: ...Module Frames and Modules 8 66 2062973 004 For your notes ...
Page 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual ...
Page 320: ......
Page 336: ...Module Frames and Modules 9 14 2062973 004 For your notes ...
Page 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004 ...
Page 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004 ...
Page 380: ...Module Frames and Modules 9 58 2062973 004 For your notes ...
Page 381: ...10 Single width Airway Module E miniC Technical Manual ...
Page 382: ......
Page 410: ...Module Frames and Modules 10 28 2062973 004 ...
Page 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual ...
Page 420: ......
Page 442: ...Module Frames and Modules 11 20 2062973 004 For your notes ...
Page 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual ...
Page 444: ......
Page 468: ...Module Frames and Modules 12 22 2062973 004 For your notes ...
Page 470: ...Module Frames and Modules 12 24 2062973 004 For your notes ...
Page 471: ...13 BIS Module E BIS Rev 01 Technical Manual ...
Page 472: ......
Page 494: ...Module Frames and Modules 13 20 2062973 004 For your notes ...
Page 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual ...
Page 496: ......
Page 518: ...Module Frames and Modules 14 20 2062973 004 For your notes ...
Page 520: ...Module Frames and Modules 14 22 2062973 004 For your notes ...
Page 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual ...
Page 522: ......
Page 558: ...Module Frames and Modules 15 34 2062973 004 For your notes ...
Page 560: ...Module Frames and Modules 15 36 2062973 004 For your notes ...
Page 561: ...16 Patient Data Module v2 Technical Manual ...
Page 562: ......
Page 620: ...Module Frames and Modules 16 56 2062973 004 ...
Page 621: ......