![GE HEALTHCARE E-PRESTN Technical Manual Download Page 511](http://html.mh-extra.com/html/ge-healthcare/e-prestn/e-prestn_technical-manual_571893511.webp)
NeuroMuscular Transmission Module, E-NMT
14 - 13
2062973-004
5.4 Messages
The messages in the table below appear in the NMT parameter window (PW) or the message
field (MF).
Message
Location
Possible causes
Possible solutions
Cable off
NMT cable removed
PW
MF
NMT sensor cable and sensor
are not connected to the
module.
Connect the NMT sensor and
sensor to the module.
Measurement off
PW
Measurement is stopped.
Start the measurement.
EMG electrodes off
PW
One the recording electrodes
(red or green) in NMT
ElectroSensor is disconnected or
broken.
Attach the electrodes to continue
or start measurement.
Replace the NMT Electrosensor, if
broken.
Check electrodes
PW
One of the stimulating
electrodes (white or brown) is
disconnected or the cable is
broken.
Check the white and brown
stimulating electrodes.
Replace the NMT ElectroSensor or
MechanoSensor cable, if broken.
Supramax search
PW
The search of supramaximal
stimulus pulse is in progress.
Wait until the search of
supramaximal stimulus pulse is
completed.
Setting reference
PW
Reference setting is in progress. Wait until the setting of reference
level is completed.
Regional block
PW
Nerve location with regional
block stimulation is in progress.
Stop regional block stimulation.
TETANIC
PW
Tetanic stimulation is in
progress.
Wait until the tetanic stimulation is
completed.
Response too weak
PW
The maximum gain is insufficient
to increase the response signal
amplitude to a measurable level:
•
stimulus electrodes are loose
•
response electrodes are
attached to a wrong place
Check that the stimulus current is
not too weak.
Check stimulus electrode
placement and connections.
Check that the response electrodes
are not disconnected.
If the electrodes are dry, replace
them.
Check that skin at the electrode
site is properly prepared.
Reference not stable
PW
Reference setting fails because
responses differ more than 10%:
•
patient is relaxated
•
movement artifacts
Start measurement with fixed
current without reference
measurement.
Summary of Contents for E-PRESTN
Page 2: ......
Page 4: ...Module Frames and Modules 2062973 004 For your notes ...
Page 6: ...Module Frames and Modules 1 ii 2062973 004 ...
Page 10: ...Module Frames and Modules 1 4 2062973 004 For your notes ...
Page 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual ...
Page 12: ......
Page 92: ...Module Frames and Modules 2 78 2062973 004 For your notes ...
Page 94: ...Module Frames and Modules 2 80 2062973 004 For your notes ...
Page 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual ...
Page 96: ......
Page 124: ...Module Frames and Modules 3 26 2062973 004 For your notes ...
Page 126: ......
Page 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004 ...
Page 160: ...Module Frames and Modules 4 32 2062973 004 For your notes ...
Page 162: ...Module Frames and Modules 4 34 2062973 004 For your notes ...
Page 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual ...
Page 164: ......
Page 198: ...Module Frames and Modules 5 32 2062973 004 For your notes ...
Page 199: ...6 Masimo Module E MASIMO Technical Manual ...
Page 200: ......
Page 202: ...Module Frames and Modules 6 ii 2062973 004 ...
Page 218: ...Module Frames and Modules 6 16 2062973 004 For your notes ...
Page 220: ...Module Frames and Modules 6 18 2062973 004 For your notes ...
Page 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual ...
Page 222: ......
Page 224: ...Module Frames and Modules 7 ii 2062973 004 ...
Page 236: ...Module Frames and Modules 7 12 2062973 004 ...
Page 244: ...Module Frames and Modules 7 20 2062973 004 For your notes ...
Page 246: ...Module Frames and Modules 7 22 2062973 004 For your notes ...
Page 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual ...
Page 248: ......
Page 284: ...Module Frames and Modules 8 34 2062973 004 ...
Page 316: ...Module Frames and Modules 8 66 2062973 004 For your notes ...
Page 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual ...
Page 320: ......
Page 336: ...Module Frames and Modules 9 14 2062973 004 For your notes ...
Page 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004 ...
Page 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004 ...
Page 380: ...Module Frames and Modules 9 58 2062973 004 For your notes ...
Page 381: ...10 Single width Airway Module E miniC Technical Manual ...
Page 382: ......
Page 410: ...Module Frames and Modules 10 28 2062973 004 ...
Page 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual ...
Page 420: ......
Page 442: ...Module Frames and Modules 11 20 2062973 004 For your notes ...
Page 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual ...
Page 444: ......
Page 468: ...Module Frames and Modules 12 22 2062973 004 For your notes ...
Page 470: ...Module Frames and Modules 12 24 2062973 004 For your notes ...
Page 471: ...13 BIS Module E BIS Rev 01 Technical Manual ...
Page 472: ......
Page 494: ...Module Frames and Modules 13 20 2062973 004 For your notes ...
Page 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual ...
Page 496: ......
Page 518: ...Module Frames and Modules 14 20 2062973 004 For your notes ...
Page 520: ...Module Frames and Modules 14 22 2062973 004 For your notes ...
Page 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual ...
Page 522: ......
Page 558: ...Module Frames and Modules 15 34 2062973 004 For your notes ...
Page 560: ...Module Frames and Modules 15 36 2062973 004 For your notes ...
Page 561: ...16 Patient Data Module v2 Technical Manual ...
Page 562: ......
Page 620: ...Module Frames and Modules 16 56 2062973 004 ...
Page 621: ......