Module Frames and Modules
8 - 22
2062973-004
Check the gas sampling system for possible leakages.
a.
Disconnect the module from the monitor.
b.
Detach the module front cover and casing, see chapters “
” and “
6.2.2. Detaching the Module Casing
c.
Block the OM reference tube with the forceps. Correct positioning of the forceps is
indicated by the figure below.
NOTE: Be careful when attaching the forceps to the tube and avoid stretching the
tube. Short pieces of silicone tubing on the forcep jaws can be used to protect the
tube from breaks that may appear when the tube is compressed between the jaws.
d.
Connect a new D-fend Pro water trap to the module.
e.
Connect a new gas sampling line to the sampling line connector in the water trap.
f.
Connect the other end of the gas sampling line to a pressure manometer and a
pressure pump.
g.
Block the sample gas out (gas exhaust) connector.
h.
Carefully pump 80 mmHg
±
20 mmHg pressure to the gas sampling system. Let the
pressure stabilize for 10 - 20 seconds.
i.
Check that the pressure reading does not drop more than 2 mmHg during
25 seconds.
j.
Release the forceps, and attach the module casing. Make sure that the tubing fits
nicely into the module casing.
2.
Spirometry System Leak Test
NOTE: Perform this test only for E-sCOV, E-sCOVX, E-sCAiOV, and E-sCAiOVX modules.
NOTE: The gas module shall be disconnected from the monitor during the leak test.
NOTE: The spirometry pressure transducers are very sensitive for differential overpressure. A
momentary differential pressure between the two spirometry connectors exceeding 25 cmH
2
O
(18 mmHg) may damage the pressure sensors. To ensure that both pressure channels are
equally pressurized, make sure that the tubing between the manometer and the two
spirometry connectors is connected tightly, the tubes are equally long and thick and not
kinked.
NOTE: Do not overpressure the spirometry sampling system. A static pressure exceeding
300 cmH
2
O (220 mmHg) may damage the pressure sensor.
Check the spirometry sampling system for possible leakages.
Summary of Contents for E-PRESTN
Page 2: ......
Page 4: ...Module Frames and Modules 2062973 004 For your notes ...
Page 6: ...Module Frames and Modules 1 ii 2062973 004 ...
Page 10: ...Module Frames and Modules 1 4 2062973 004 For your notes ...
Page 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual ...
Page 12: ......
Page 92: ...Module Frames and Modules 2 78 2062973 004 For your notes ...
Page 94: ...Module Frames and Modules 2 80 2062973 004 For your notes ...
Page 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual ...
Page 96: ......
Page 124: ...Module Frames and Modules 3 26 2062973 004 For your notes ...
Page 126: ......
Page 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004 ...
Page 160: ...Module Frames and Modules 4 32 2062973 004 For your notes ...
Page 162: ...Module Frames and Modules 4 34 2062973 004 For your notes ...
Page 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual ...
Page 164: ......
Page 198: ...Module Frames and Modules 5 32 2062973 004 For your notes ...
Page 199: ...6 Masimo Module E MASIMO Technical Manual ...
Page 200: ......
Page 202: ...Module Frames and Modules 6 ii 2062973 004 ...
Page 218: ...Module Frames and Modules 6 16 2062973 004 For your notes ...
Page 220: ...Module Frames and Modules 6 18 2062973 004 For your notes ...
Page 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual ...
Page 222: ......
Page 224: ...Module Frames and Modules 7 ii 2062973 004 ...
Page 236: ...Module Frames and Modules 7 12 2062973 004 ...
Page 244: ...Module Frames and Modules 7 20 2062973 004 For your notes ...
Page 246: ...Module Frames and Modules 7 22 2062973 004 For your notes ...
Page 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual ...
Page 248: ......
Page 284: ...Module Frames and Modules 8 34 2062973 004 ...
Page 316: ...Module Frames and Modules 8 66 2062973 004 For your notes ...
Page 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual ...
Page 320: ......
Page 336: ...Module Frames and Modules 9 14 2062973 004 For your notes ...
Page 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004 ...
Page 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004 ...
Page 380: ...Module Frames and Modules 9 58 2062973 004 For your notes ...
Page 381: ...10 Single width Airway Module E miniC Technical Manual ...
Page 382: ......
Page 410: ...Module Frames and Modules 10 28 2062973 004 ...
Page 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual ...
Page 420: ......
Page 442: ...Module Frames and Modules 11 20 2062973 004 For your notes ...
Page 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual ...
Page 444: ......
Page 468: ...Module Frames and Modules 12 22 2062973 004 For your notes ...
Page 470: ...Module Frames and Modules 12 24 2062973 004 For your notes ...
Page 471: ...13 BIS Module E BIS Rev 01 Technical Manual ...
Page 472: ......
Page 494: ...Module Frames and Modules 13 20 2062973 004 For your notes ...
Page 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual ...
Page 496: ......
Page 518: ...Module Frames and Modules 14 20 2062973 004 For your notes ...
Page 520: ...Module Frames and Modules 14 22 2062973 004 For your notes ...
Page 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual ...
Page 522: ......
Page 558: ...Module Frames and Modules 15 34 2062973 004 For your notes ...
Page 560: ...Module Frames and Modules 15 36 2062973 004 For your notes ...
Page 561: ...16 Patient Data Module v2 Technical Manual ...
Page 562: ......
Page 620: ...Module Frames and Modules 16 56 2062973 004 ...
Page 621: ......