Apnea deactivated
PW
Apnea alarm start-up conditions are
not reached.
Apnea alarm detection is
activated after the 3 breaths are
detected.
Sensor INOP
WF
IR Lamp failure.
Ambient pressure is too high or low.
CPU failure.
No response from the gas module,
high temperature inside the module,
or EEPROM checksum failure.
Check lamp wire connection.
Check the ambient pressure
from the Gas Calibrations menu.
Replace CPU.
Return the module to GE
Healthcare for service.
Calibrating gas
sensor
WF
O
2
, CO
2
and N
2
O measurements are
not available during the first 2
minutes after the module is
connected due to warm-up.
Anesthetic agent measurement is not
available during the first 5 minutes
after the module is connected due to
warm-up.
Wait until the warm-up is
completed.
Over Scale
WF
Incorrect waveform scale for the
parameter. The waveform clipped
because measured gas
concentration exceeds the upper
limit of the current scale.
Change to the appropriate
waveform scale.
Low gas sample flow MF
Sample flow deviates to less than
80% of the module specific nominal
flow value.
Gas sampling line, gas output, water
trap, or internal tubing is blocked.
Pump failure.
Error in pneumatic initialization.
Check sample flow rate. Adjust,
if needed.
Check or replace the gas
sampling line, water trap, or
internal tubing.
1.Check that the screw inside
the sampling pump is not loose
(see
).
2.Replace pump, if needed.
Connect sample line and water
trap to module and restart
monitor
Incompatible gas
module
MF
Incompatible gas module detected
by the monitor.
Check the compatibility of the
gas module.
Gas measurements
removed
MF
The module is disconnected.
1. Reconnect the module.
2. Check that the D25 connector
is clean and intact.
Identical gas
modules
MF
The monitor detects gas
measurement from two or more
modules.
Remove excess modules
providing gas measurement.
Message
Location
Possible causes
Possible solutions
Summary of Contents for E-PRESTN
Page 2: ......
Page 4: ...Module Frames and Modules 2062973 004 For your notes ...
Page 6: ...Module Frames and Modules 1 ii 2062973 004 ...
Page 10: ...Module Frames and Modules 1 4 2062973 004 For your notes ...
Page 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual ...
Page 12: ......
Page 92: ...Module Frames and Modules 2 78 2062973 004 For your notes ...
Page 94: ...Module Frames and Modules 2 80 2062973 004 For your notes ...
Page 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual ...
Page 96: ......
Page 124: ...Module Frames and Modules 3 26 2062973 004 For your notes ...
Page 126: ......
Page 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004 ...
Page 160: ...Module Frames and Modules 4 32 2062973 004 For your notes ...
Page 162: ...Module Frames and Modules 4 34 2062973 004 For your notes ...
Page 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual ...
Page 164: ......
Page 198: ...Module Frames and Modules 5 32 2062973 004 For your notes ...
Page 199: ...6 Masimo Module E MASIMO Technical Manual ...
Page 200: ......
Page 202: ...Module Frames and Modules 6 ii 2062973 004 ...
Page 218: ...Module Frames and Modules 6 16 2062973 004 For your notes ...
Page 220: ...Module Frames and Modules 6 18 2062973 004 For your notes ...
Page 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual ...
Page 222: ......
Page 224: ...Module Frames and Modules 7 ii 2062973 004 ...
Page 236: ...Module Frames and Modules 7 12 2062973 004 ...
Page 244: ...Module Frames and Modules 7 20 2062973 004 For your notes ...
Page 246: ...Module Frames and Modules 7 22 2062973 004 For your notes ...
Page 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual ...
Page 248: ......
Page 284: ...Module Frames and Modules 8 34 2062973 004 ...
Page 316: ...Module Frames and Modules 8 66 2062973 004 For your notes ...
Page 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual ...
Page 320: ......
Page 336: ...Module Frames and Modules 9 14 2062973 004 For your notes ...
Page 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004 ...
Page 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004 ...
Page 380: ...Module Frames and Modules 9 58 2062973 004 For your notes ...
Page 381: ...10 Single width Airway Module E miniC Technical Manual ...
Page 382: ......
Page 410: ...Module Frames and Modules 10 28 2062973 004 ...
Page 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual ...
Page 420: ......
Page 442: ...Module Frames and Modules 11 20 2062973 004 For your notes ...
Page 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual ...
Page 444: ......
Page 468: ...Module Frames and Modules 12 22 2062973 004 For your notes ...
Page 470: ...Module Frames and Modules 12 24 2062973 004 For your notes ...
Page 471: ...13 BIS Module E BIS Rev 01 Technical Manual ...
Page 472: ......
Page 494: ...Module Frames and Modules 13 20 2062973 004 For your notes ...
Page 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual ...
Page 496: ......
Page 518: ...Module Frames and Modules 14 20 2062973 004 For your notes ...
Page 520: ...Module Frames and Modules 14 22 2062973 004 For your notes ...
Page 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual ...
Page 522: ......
Page 558: ...Module Frames and Modules 15 34 2062973 004 For your notes ...
Page 560: ...Module Frames and Modules 15 36 2062973 004 For your notes ...
Page 561: ...16 Patient Data Module v2 Technical Manual ...
Page 562: ......
Page 620: ...Module Frames and Modules 16 56 2062973 004 ...
Page 621: ......