![GE HEALTHCARE E-PRESTN Technical Manual Download Page 344](http://html.mh-extra.com/html/ge-healthcare/e-prestn/e-prestn_technical-manual_571893344.webp)
Module Frames and Modules
9 - 22
2062973-004
c.
Pump 50 mmHg ±10 mmHg (~68 cmH
2
O) pressure to the Spirometry sampling
system. Let the pressure stabilize for approximately 10 seconds.
d.
Verify that the pressure reading does not drop more than 3 mmHg during one
minute.
NOTE: The gas module shall be disconnected from the monitor during the leak test.
NOTE: The spirometry pressure transducers are very sensitive for differential overpressure. A
momentary differential pressure between the two spirometry connectors exceeding 25 cmH
2
O
(18 mmHg) may damage the pressure sensors. To ensure that both pressure channels are
equally pressurized, make sure that the tubing between the manometer and the two
spirometry connectors is connected tightly, the tubes are equally long and thick and not
kinked.
NOTE: Do not overpressure the spirometry sampling system. A static pressure exceeding 300
cmH
2
O (220 mmHg) may damage the pressure sensor.
3.
Fan
a.
Connect the module to the monitor.
b.
Check that the gas module's fan is running.
4.
Module Keys
NOTE: Perform this test only for E-COV, E-COVX, E-CAiOV and E-CAiOVX modules.
a.
Press the
Change Loop
module key.
b.
Check that the spirometry loop is changed from Flow / Vol loop to Paw/Vol loop, or
vice versa.
c.
Leave the Flow / Vol loop to the screen.
5.
Sample Flow Rate Check
Check the sample flow rate.
NOTE: Gas measurement is not available during the first 2 to 5 minutes after the module is
connected due to warming up. A message 'Calibrating Gas Sensor' is shown in the waveform
field. Wait until warm-up is completed before proceeding with the next steps.
NOTE: The ambient temperature and air pressure influence the flow rate measured by the flow
meter. A flow meter, which has been calibrated at 21.1
°
C (70
°
F) and 760 mmHg (1013 mbar),
measures the flow rate correctly under the same conditions, i.e. in room temperature at sea
level. A flow rate correction as instructed by the manufacturer of the flow meter needs to be
performed when measuring flow rate under other conditions, for example in high altitude.
a.
Connect the gas sampling line (2 m / 7 ft with E-COVX and E-CAiOVX, 3 m / 10 ft with
E-CO, E-COV, E-CAiO and E-CAiOV) to the Sampling line connector.
b.
Connect the other end of the gas sampling line to a flowmeter.
c.
Check the sample flow rate reading from the flowmeter. The flow rate shall be within
the specification limit 200 ± 20 ml/min.
NOTE: Readjustment is needed, if the measured value is not within the specification limit. Adjust
the sample gas flow rate according to the instructions in section
.
6.
Reference Gas Flow Rate Check
Check the flow rate in reference gas inlet:
Summary of Contents for E-PRESTN
Page 2: ......
Page 4: ...Module Frames and Modules 2062973 004 For your notes ...
Page 6: ...Module Frames and Modules 1 ii 2062973 004 ...
Page 10: ...Module Frames and Modules 1 4 2062973 004 For your notes ...
Page 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual ...
Page 12: ......
Page 92: ...Module Frames and Modules 2 78 2062973 004 For your notes ...
Page 94: ...Module Frames and Modules 2 80 2062973 004 For your notes ...
Page 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual ...
Page 96: ......
Page 124: ...Module Frames and Modules 3 26 2062973 004 For your notes ...
Page 126: ......
Page 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004 ...
Page 160: ...Module Frames and Modules 4 32 2062973 004 For your notes ...
Page 162: ...Module Frames and Modules 4 34 2062973 004 For your notes ...
Page 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual ...
Page 164: ......
Page 198: ...Module Frames and Modules 5 32 2062973 004 For your notes ...
Page 199: ...6 Masimo Module E MASIMO Technical Manual ...
Page 200: ......
Page 202: ...Module Frames and Modules 6 ii 2062973 004 ...
Page 218: ...Module Frames and Modules 6 16 2062973 004 For your notes ...
Page 220: ...Module Frames and Modules 6 18 2062973 004 For your notes ...
Page 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual ...
Page 222: ......
Page 224: ...Module Frames and Modules 7 ii 2062973 004 ...
Page 236: ...Module Frames and Modules 7 12 2062973 004 ...
Page 244: ...Module Frames and Modules 7 20 2062973 004 For your notes ...
Page 246: ...Module Frames and Modules 7 22 2062973 004 For your notes ...
Page 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual ...
Page 248: ......
Page 284: ...Module Frames and Modules 8 34 2062973 004 ...
Page 316: ...Module Frames and Modules 8 66 2062973 004 For your notes ...
Page 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual ...
Page 320: ......
Page 336: ...Module Frames and Modules 9 14 2062973 004 For your notes ...
Page 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004 ...
Page 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004 ...
Page 380: ...Module Frames and Modules 9 58 2062973 004 For your notes ...
Page 381: ...10 Single width Airway Module E miniC Technical Manual ...
Page 382: ......
Page 410: ...Module Frames and Modules 10 28 2062973 004 ...
Page 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual ...
Page 420: ......
Page 442: ...Module Frames and Modules 11 20 2062973 004 For your notes ...
Page 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual ...
Page 444: ......
Page 468: ...Module Frames and Modules 12 22 2062973 004 For your notes ...
Page 470: ...Module Frames and Modules 12 24 2062973 004 For your notes ...
Page 471: ...13 BIS Module E BIS Rev 01 Technical Manual ...
Page 472: ......
Page 494: ...Module Frames and Modules 13 20 2062973 004 For your notes ...
Page 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual ...
Page 496: ......
Page 518: ...Module Frames and Modules 14 20 2062973 004 For your notes ...
Page 520: ...Module Frames and Modules 14 22 2062973 004 For your notes ...
Page 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual ...
Page 522: ......
Page 558: ...Module Frames and Modules 15 34 2062973 004 For your notes ...
Page 560: ...Module Frames and Modules 15 36 2062973 004 For your notes ...
Page 561: ...16 Patient Data Module v2 Technical Manual ...
Page 562: ......
Page 620: ...Module Frames and Modules 16 56 2062973 004 ...
Page 621: ......