Module Frames and Modules
16 - 18
2062973-004
3.3.3 Impedance respiration tests
6. Respiration Rate
a. Check that the RESP waveform is shown and the RR value is 20 (±5).
b. Configure the simulator’s Apnea Simulation to “32 sec”.
7. Apnea Detection
a. Check that the monitor activates the APNEA alarm.
b. Configure the simulator’s Apnea Simulation to “OFF”
8. Mark this task as complete on the checkout form.
3.3.4 Invasive pressure tests
9. Zeroing
a. Ensure that the simulator InvBP output channels are configured to
“0 mmHg static
pressure” or “atmosphere”.
b. Zero the invasive pressure of the PDM by pressing the
IBP Zero All
button on the PDM
or
Zero the invasive pressure of the patient monitor by pressing the
Zero All Pressures
button on the patient monitor.
10. Static Pressure
a. Configure the simulator’s InvBP output channel to “240 mmHg static pressure”.
b. Check that the pressure waveform for the P1, P2, P3 and P4 waveform fields displays
the flat pressure line in the related waveform field.
c. Check that a reading of 240 ± 4 mmHg displays on the patient monitor.
11. Pressure Waveforms
a. Configure the simulator’s InvBP output channels to generate a waveform.
b. Check that the pressure waveform for P1, P2, P3 and P4 waveform fields displays an
appropriate waveform.
12. Mark this task as complete on the checkout form.
3.3.5 Temperature tests
Perform the following test step both to the T1 and T2 temperature channels.
NOTE: The ‘x’ in Tx refers to the temperature channel being tested.
13. Temperature check
a. Check that the reading in the Tx temperature channel is 37°C/98.6°F ± 0.1°C/0.18°F.
NOTE: You can alternatively perform this test using a temperature test set (P/N 884515-HEL).
Test each temperature channel with both test plugs. The reading in the Tx temperature
channel shall be 37°C/98.6°F ± 0.1°C/0.18°F.
b. Mark this task as complete on the checkout form.
3.3.6 SpO2 tests
14a. SpO
2
functions (Masimo)
a. Make sure that the SpO
2
parameter is turned on.
b. Verify the following are displayed at the patient monitor:
•
A waveform with an SpO
2
label.
•
An SpO
2
% reading between 78-84%.
Summary of Contents for E-PRESTN
Page 2: ......
Page 4: ...Module Frames and Modules 2062973 004 For your notes ...
Page 6: ...Module Frames and Modules 1 ii 2062973 004 ...
Page 10: ...Module Frames and Modules 1 4 2062973 004 For your notes ...
Page 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual ...
Page 12: ......
Page 92: ...Module Frames and Modules 2 78 2062973 004 For your notes ...
Page 94: ...Module Frames and Modules 2 80 2062973 004 For your notes ...
Page 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual ...
Page 96: ......
Page 124: ...Module Frames and Modules 3 26 2062973 004 For your notes ...
Page 126: ......
Page 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004 ...
Page 160: ...Module Frames and Modules 4 32 2062973 004 For your notes ...
Page 162: ...Module Frames and Modules 4 34 2062973 004 For your notes ...
Page 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual ...
Page 164: ......
Page 198: ...Module Frames and Modules 5 32 2062973 004 For your notes ...
Page 199: ...6 Masimo Module E MASIMO Technical Manual ...
Page 200: ......
Page 202: ...Module Frames and Modules 6 ii 2062973 004 ...
Page 218: ...Module Frames and Modules 6 16 2062973 004 For your notes ...
Page 220: ...Module Frames and Modules 6 18 2062973 004 For your notes ...
Page 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual ...
Page 222: ......
Page 224: ...Module Frames and Modules 7 ii 2062973 004 ...
Page 236: ...Module Frames and Modules 7 12 2062973 004 ...
Page 244: ...Module Frames and Modules 7 20 2062973 004 For your notes ...
Page 246: ...Module Frames and Modules 7 22 2062973 004 For your notes ...
Page 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual ...
Page 248: ......
Page 284: ...Module Frames and Modules 8 34 2062973 004 ...
Page 316: ...Module Frames and Modules 8 66 2062973 004 For your notes ...
Page 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual ...
Page 320: ......
Page 336: ...Module Frames and Modules 9 14 2062973 004 For your notes ...
Page 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004 ...
Page 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004 ...
Page 380: ...Module Frames and Modules 9 58 2062973 004 For your notes ...
Page 381: ...10 Single width Airway Module E miniC Technical Manual ...
Page 382: ......
Page 410: ...Module Frames and Modules 10 28 2062973 004 ...
Page 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual ...
Page 420: ......
Page 442: ...Module Frames and Modules 11 20 2062973 004 For your notes ...
Page 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual ...
Page 444: ......
Page 468: ...Module Frames and Modules 12 22 2062973 004 For your notes ...
Page 470: ...Module Frames and Modules 12 24 2062973 004 For your notes ...
Page 471: ...13 BIS Module E BIS Rev 01 Technical Manual ...
Page 472: ......
Page 494: ...Module Frames and Modules 13 20 2062973 004 For your notes ...
Page 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual ...
Page 496: ......
Page 518: ...Module Frames and Modules 14 20 2062973 004 For your notes ...
Page 520: ...Module Frames and Modules 14 22 2062973 004 For your notes ...
Page 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual ...
Page 522: ......
Page 558: ...Module Frames and Modules 15 34 2062973 004 For your notes ...
Page 560: ...Module Frames and Modules 15 36 2062973 004 For your notes ...
Page 561: ...16 Patient Data Module v2 Technical Manual ...
Page 562: ......
Page 620: ...Module Frames and Modules 16 56 2062973 004 ...
Page 621: ......