24
VIEW MENU
ALARM
LEVELS
BAT VOLTAGE
SERVICE
VERSION
EXIT
2. Use the navigation arrows to highlight VERSION and
press the MODE button. The software version screen
then be shown.
VERSIONS
FLASH Version 3.81
2005/06/20 08:47:59
OTP Version 1.11
Serial Number 01661
EXIT
3. Press the MODE button to return to the Main Menu.
3.5 The Alarms Menu
Cannonball3 gas alarms are user-adjustable and may be set
anywhere within the range of the sensor channel. When an
alarm set point is exceeded for a sensor, a loud audible alarm
sounds, and the three bright red LED alarm lights blink.
Cannonball3 alarms are self-resetting unless the alarm latch is
enabled. With the alarm latch disabled, visible and audible
alarms cease when readings drop back below the pre-set
alarm levels. With the alarm latch enabled, visible and audible
alarms continue to sound after the atmospheric hazard has
cleared. The user must then manually reset them by pressing
the MODE button.
Alarm latch procedures are discussed in section 3.5.3.
Factory default settings can be restored at any time during
normal operation by using the procedures discussed in
section 3.5.4.
3.5.1
Entering the Alarms Menu
1. Enter the Main Menu as discussed above in section 3.1.1.
MAIN MENU
SCREEN CALIBRATION
VIEW
ALARMS
TIME OPTIONS
ID INFO
EXIT
2. Use the navigation arrows to highlight ALARMS and press
the MODE button. The Alarms Menu will then be shown.
ALARMS MENU
CUSTOM LATCH
DEFAULT TEMPERATURE
EXIT
3.5.2 Custom
alarm
settings
Cannonball3 gas alarms are user-adjustable and may be set
anywhere within the range of the sensor channel.
To customize alarm settings:
1. Enter the Alarm Menu as described above in section 3.5.1.
ALARMS MENU
CUSTOM
LATCH
DEFAULT TEMPERATURE
WARNINGS EXIT
2. Use the navigation arrows to highlight CUSTOM and press
the MODE button. The current alarm settings screen will
then be shown.
CURRENT ALARM SETTINGS
LOW CEIL
O
2
19.5 23.5
LEL 10
CEIL STEL TWA
CO
100 35 100
H
2
S
10 15 10
EXIT
3. To modify alarm settings, use the up and down navigation
arrows to select the specific sensor channel that requires
adjustment.
CURRENT ALARM SETTINGS
LOW CEIL
O
2
19.5 23.5
LEL 10
CEIL STEL TWA
CO
100 35 100
H
2
S
10 15 10
EXIT
4. Press the MODE button to confirm the selection.