16
Operating Mode
Current Mode :Technician
Text Only
↑↓
Operating Mode
Current Mode :Technician
Basic
↑↓
Operating Mode
Current Mode :Technician
Basic/Peak
↑↓
Operating Mode
Current Mode :Technician
Technician
6. Once the preferred operating mode is highlighted, confirm
the selection by pressing the MODE button.
Operating Mode
Current Mode :Technician
BASIC
Save Changes?
YES
NO CANCEL
7. Press the MODE button with YES highlighted to save the
new operating mode.
Note: Changing modes or otherwise reprogramming the
instrument is reserved for authorized employees.
2.4 Batteries
The Cannonball3 can be equipped with either an alkaline
battery pack or a NiMH (Nickel Metal Hydride) NiMH
rechargeable battery pack.
The Cannonball3 must be located in a
non-hazardous location whenever the alkaline batteries
are removed from the alkaline battery pack. Removing the
alkaline batteries from the alkaline battery pack in a
hazardous location may impair intrinsic safety.
The Cannonball3 must be located in a
non-hazardous location during the charging cycle.
Charging the Cannonball3 in a hazardous location may
impair intrinsic safety.
2.4.1 Alkaline
batteries
The Cannonball3 with alkaline batteries is designed to provide
at least 24 hours of continuous use with each set of 5 fresh D-
cell disposable alkaline batteries.
Use only Energizer E95 or EN95,
Duracell MN1300, or Duracell PC1300, 1.5V D cell Alkaline
batteries in the Cannonball3. Substitution of batteries may
impair intrinsic safety.
2.4.1.1 Replacing alkaline batteries
1. Check that the Cannonball3 is not located in a hazardous
(potentially combustible) area.
2. Make sure that the Cannonball3 is turned off.
3. Loosen the battery retention screws and remove the
battery pack from the instrument.
4. Remove the screw on the top of the battery pack and open
the battery pack. Remove the old batteries and install new
batteries in accordance with the diagram inside the battery
pack.
5. Replace the screw on the top of the battery pack and
reinstall the battery pack into the instrument. Tighten the
battery retention screws.
Note: Always dispose of alkaline and NiMH batteries and
battery packs in accordance with local ordinances.
Note: The Cannonball3 is designed to turn itself on
whenever the battery pack is removed and then replaced.
This ensures that in the event of an interruption in power
the instrument is not accidentally turned off. Any time the
batteries are momentarily removed or replaced it will be
necessary to manually turn the Cannonball3 off if the
instrument is not going to be put into immediate use.
2.4.2
NiMH rechargeable battery
The Cannonball3 equipped with a rechargeable NiMH (nickel
metal hydride) battery is designed to provide at least 16 hours
of continuous use between charging cycles. The NiMH battery
pack in the Cannonball3 is sealed and should not be
disassembled in the field.
The Cannonball3 must be located in a
non-hazardous location during the charging cycle.
Charging the Cannonball3 in a hazardous location may
impair intrinsic safety.
2.4.2.1 Storage guidelines for the NiMH battery.
Never store NiMH-version Cannonball3 instruments at
temperatures above 30
°
Celsius (86
°
Fahrenheit). Nickel
Metal Hydride batteries may suffer deterioration resulting in
damage to the internal components when stored at high
Summary of Contents for Cannonball3